Compliance and audits
What is calibration? Meaning, purpose and types
Calibration means comparing what an instrument reads with a reference standard of known value and uncertainty, then recording the difference so the instrument's results can be trusted or corrected. Calibration on its own does not adjust the instrument or pass or fail it. Adjustment and verification are separate steps.
On this page
What does calibration mean?
The International Vocabulary of Metrology (VIM, JCGM 200:2012, entry 2.39) describes calibration as two steps carried out under stated conditions. First you compare the values provided by measurement standards, each with its uncertainty, with what the instrument indicates. Then you use that comparison to work out how a reading on the instrument turns into a measurement result.
The result can take several forms. The VIM lists a statement, a calibration function, a diagram, a curve or a table. In some cases it is a simple correction that you add to the reading or multiply it by, always with its uncertainty. A note to the entry adds that people often treat the first step alone as the calibration, and that first step is what a typical certificate reports: the error at each test point and the uncertainty of that error.
Take a digital caliper checked against gauge blocks. At 25 mm it reads 25.02 mm, so the calibration records an error of +0.02 mm at that point, with the uncertainty of the check. Whether +0.02 mm is acceptable is a separate question, answered by comparing it with the caliper's tolerance.
| Reference value | Caliper reading | Error |
|---|---|---|
| 10.00 mm | 10.01 mm | +0.01 mm |
| 25.00 mm | 25.02 mm | +0.02 mm |
| 100.00 mm | 100.02 mm | +0.02 mm |
Calibration vs adjustment and verification
Calibration measures. It does not change the instrument. Adjustment (VIM 3.11) is the work done on an instrument so that it reads correctly, such as setting its zero, offset or span. The VIM calls calibration a prerequisite for adjustment and notes that an adjusted instrument usually has to be calibrated again.
Verification (VIM 2.44) means showing, with evidence, that something meets the requirements set for it. For an instrument that usually means comparing the calibration result with the tolerance it has to hold and deciding pass or fail.
The three words get mixed up in other languages too, and some languages add a trap of their own. The table gives the usual terms in five languages.
| Language | Calibration | Adjustment | Verification | Watch out for |
|---|---|---|---|---|
| English | calibration | adjustment | verification | In everyday English, calibrate often means adjust. The VIM notes that adjustment is often wrongly called self-calibration. |
| German | Kalibrierung | Justierung | Verifizierung | Eichung is not a calibration. Under the Mess- und Eichgesetz (§ 3) it is an official check, assessment and marking that allows a legally controlled instrument to be used for a further verification period. |
| French | étalonnage | ajustage | vérification | Calibrage is a false friend. The OQLF lists it with ajustage as a French term for adjustment, not for calibration. |
| Spanish | calibración | ajuste | verificación | The Spanish VIM published by CEM uses these three terms and warns against calling adjustment autocalibración. |
| Portuguese | calibração | ajuste | verificação | Aferição is common in everyday speech but does not appear in the Portuguese VIM published by INMETRO and IPQ in 2012. Use calibração or verificação for what was actually done. |
Why calibrate?
Instruments drift. Wear, shock, temperature and age move their readings away from the values they gave when new, and nothing on the display tells you. Calibration measures that drift against something better known, so the readings you base decisions on can be trusted.
Those decisions have consequences. A micrometer that reads small can pass shafts that are oversize, and a thermometer that reads high can make a heat treatment look complete when it is not.
Quality standards make calibration a requirement. ISO 9001 clause 7.1.5.2 applies when measurement traceability is required, or when you consider it essential to confidence in your results. It then expects the instruments concerned to be calibrated or verified on a set schedule or before use, against references that trace back to national or international standards. Each one has to be identified so that its calibration status can be determined, for example with a label or a code, and protected from adjustment, damage or deterioration that would spoil its readings. If an instrument turns out to be unfit, you check whether earlier results were affected and act on what you find.
Calibration laboratories have their own standard, ISO/IEC 17025:2017, which covers the competence, impartiality and consistent operation of testing and calibration laboratories.
Traceability: why the reference matters
A calibration is only as good as the reference behind it. Metrological traceability (VIM 2.41) means you can link a result back to a reference, normally the SI unit, through calibrations that are documented with no gaps, each one adding its share of uncertainty. So the reference standard you calibrate against needs a valid calibration of its own, and so on up to a national metrology institute.
The reference also has to be good enough for the job. The test uncertainty ratio (TUR) compares the tolerance of the instrument with the uncertainty of its calibration. The higher it is, the less the calibration's own doubt can turn a pass into a fail or a fail into a pass.
As-found, as-left and what a certificate shows
As-found readings are taken when the instrument arrives, before anyone adjusts it. They show whether it was measuring correctly during the interval that just ended. If the instrument is adjusted, as-left readings are taken afterwards and become the baseline for the next interval. If nothing was adjusted, the two are the same.
A calibration certificate from an ISO/IEC 17025 laboratory reports the results with their measurement uncertainty and a statement of how they are traceable. Under clause 7.8.4.1 it also gives the results before and after any adjustment or repair, where they are available.
A pass or fail appears only as a statement of conformity, and clause 7.8.6 then requires the laboratory to document the decision rule behind it. Clause 7.8.4.3 says the certificate must not recommend a calibration interval unless that was agreed with the customer.
How often should an instrument be calibrated?
No standard sets one interval for all instruments. ISO 9001 asks for specified intervals and leaves their length to you. ILAC-G24:2022 / OIML D 10:2022 lists the factors to weigh, among them the tendency to wear and drift, the extent and severity of use, the environment and the calibration history, and describes methods for reviewing intervals from your own results.
In practice, start from the maker's recommendation or your customer's requirement. Shorten the interval for an instrument found out of tolerance and lengthen it, with evidence, for one that stays stable. Calibrate again after a drop, an overload or a repair, whatever the due date says.
In-house or outsourced calibration?
ISO 9001 does not require an outside laboratory. You can calibrate in-house if your own reference standards are calibrated traceably, your method is written down and the people doing the work are competent.
A common split is to check simple hand tools in-house and send reference standards and complex instruments to an accredited laboratory. Check that its scope of accreditation covers the quantity and the range you need.
Types of calibration by quantity
Calibration is grouped by the quantity measured, because each quantity has its own references, methods and standards. The table lists the common types in industry with the reference usually used. Each guide below covers the procedure, the standards and the interval for one kind of instrument, and the guide to measuring instruments describes what each instrument is used for.
| Type | Typical instruments | Usual reference |
|---|---|---|
| Dimensional | Calipers, micrometers, dial indicators, dial bore gauges, go/no-go gauges, height gauges, feeler gauges, coordinate measuring machines (CMMs) | Gauge blocks, setting rings, length bars |
| Pressure | Pressure gauges, transmitters, pressure switches | Deadweight tester or a more accurate reference gauge |
| Temperature | Thermometers, temperature probes, thermocouples, infrared thermometers | Reference thermometer in a stirred bath or dry block, blackbody source for infrared |
| Humidity | Hygrometers, humidity sensors, thermo-hygrometers | Humidity generator or chilled-mirror dew-point hygrometer |
| Electrical | Multimeters, clamp meters, power supplies | Multifunction calibrator, with a current coil for clamp meters |
| Mass | Scales and balances | Test weights of a known OIML R 111 class |
| Force | Load cells, force gauges, testing machines | Deadweight force machine or reference load cell |
| Torque | Torque wrenches and torque screwdrivers | Torque transducer or torque calibration system |
| Hardness | Rockwell, Brinell and Vickers hardness testers | Certified reference blocks, with direct checks of force, indenter and measuring system |
| Flow | Liquid and gas flow meters, anemometers | Weighing tank, prover or master meter, and a wind tunnel for anemometers |
| Chemical | pH meters, refractometers | Buffer solutions of known pH, or distilled water and reference solutions for refractometers |
| Acoustic | Sound level meters | Sound calibrator |
- Caliper calibration
- Micrometer calibration
- Dial indicator calibration
- Gauge block calibration
- Dial bore gauge calibration
- Go/no-go gauge calibration
- Height gauge calibration
- Feeler gauge calibration
- CMM calibration
- Pressure gauge calibration
- Thermometer calibration
- Infrared thermometer calibration
- Hygrometer calibration
- Multimeter calibration
- Clamp meter calibration
- Scale calibration
- Load cell calibration
- Torque wrench calibration
- Hardness tester calibration
- Flow meter calibration
- Anemometer calibration
- pH meter calibration
- Refractometer calibration
- Sound level meter calibration
- Measuring instruments: types and uses
How a calibration is done, step by step
- Identify the instrument, its range and the tolerance it has to meet, from the maker's specification, your drawing or a customer requirement.
- Choose a reference standard with a current calibration and an uncertainty small enough for that tolerance.
- Let the instrument and the reference settle in the same stable conditions, such as 20 °C for precise length work.
- Take as-found readings at several test points across the range the instrument is used over, before any adjustment.
- Work out the error at each point and the uncertainty of the calibration, then compare the errors with the tolerance.
- If the instrument needs adjusting, adjust it and repeat the readings as left.
- If an as-found reading was out of tolerance, check what the instrument measured since its last good calibration.
- Record the results, issue the certificate or label and set the next due date.
Where Axiospec fits
Axiospec is calibration management software. It records each calibration with its as-found and as-left readings and the certificate, works out the next due date and puts an instrument into quarantine when a failed calibration is accepted. With a standard such as ISO 9001 or ISO/IEC 17025 selected for the workspace, each calibration also names the reference standard used. The free plan covers up to 50 active instruments.
Common questions
- What is calibration in simple words?
- Calibration is checking an instrument against a reference whose value is known more accurately, and recording how far off the instrument reads and how sure that result is. It tells you the error. It does not fix the instrument or decide whether it passes.
- What is the difference between calibration and adjustment?
- Calibration compares the instrument with a reference and records the error without changing anything. Adjustment changes the instrument, for example its zero or span, so it reads closer to the reference. The VIM treats calibration as a prerequisite for adjustment, and an adjusted instrument usually has to be calibrated again.
- Does a calibrated instrument mean it passed?
- No. Calibration gives the error and its uncertainty at each test point. A pass or fail comes from comparing those results with a tolerance, which is verification, and a certificate can report results with no statement of conformity at all.
- How often should instruments be calibrated?
- No standard sets one interval for all instruments. Start from the maker's recommendation or your customer's requirement, then adjust the interval from the instrument's own as-found history, its use and its environment.
- What is a calibration certificate?
- It is the document that reports a calibration: the instrument, the reference standards used, the results at each test point with their measurement uncertainty and a statement of how the results are traceable. Under ISO/IEC 17025:2017 it also gives the results before and after any adjustment, where they are available.
- What are the main types of calibration?
- Calibration is usually grouped by quantity: dimensional, pressure, temperature, electrical, mass, force, torque, flow, chemical and acoustic. Each uses its own references, such as gauge blocks for length, test weights for mass and a multifunction calibrator for electrical instruments.
Put it into practice
Got a gage list? Send it over and we load it for you, usually in a couple of business days. Free on every plan. Then log every calibration to a tamper-evident audit trail and produce records on demand.
Axiospec is a documentation and workflow tool. It helps you keep clean, traceable, audit-ready records. Certification depends on your own processes, scope and assessor.
Take the audit-prep checklist with you
First the checklist, with each item phrased the way an auditor asks it. Then four short emails with practical tips. You can unsubscribe in one click.
First we email you a link to confirm your address. Nothing else comes until you confirm, and you can unsubscribe from any email after that. See our Privacy Policy.