Height gauge: how to use it on a surface plate and how to calibrate it
A height gauge measures and marks heights above a surface plate: a slider carrying a scriber or probe moves along a vertical column, and its scale, dial or display shows the height above the plate or above a reference you set. It is calibrated by measuring gauge blocks, a step gauge or a height master standing on the plate and comparing each reading with the known height.
On this page
What is a height gauge?
ISO 13225 covers height gauges that measure lengths at right angles to the surface plate they stand on. A slider carrying a measuring stylus moves along a vertical beam against a scale, and the plate is the reference for all its readings.
One instrument does two jobs. With a scriber it marks lines on a part at a set height, which is called marking out. With a probe or a test indicator it measures heights, steps and the distances between features.
Height gauges differ in how they show the reading and how they touch the part. The table shows the common kinds and the height master used to check them.
| Kind | How it reads | Notes |
|---|---|---|
| Vernier height gauge | A main scale on the beam and a vernier scale on the slider | Mitutoyo's example reads to 0.02 mm. Read it straight on to avoid parallax. |
| Dial height gauge | A counter for whole millimetres and a dial for the fraction | Mitutoyo calls its version a mechanical digit height gauge. |
| Digital height gauge | A digital display with zero and preset | Many can hold a reading and allow for the diameter of the probe ball. |
| High-precision height gauge | A probe driven by a handwheel, with a control panel that works out the results | Measures up and down with a probe constant. TESA's TESA-HITE floats on an air cushion over the plate. |
| Height master | Measuring faces at known heights | A reference standard for setting and checking height gauges, with faces that point up and down. |
Start with the surface plate
The surface plate is the reference plane of a height gauge, and the gauge's own calibration assumes a good plate. Mitutoyo asks for a plate of JIS class 1 or better under its QM-Height gauges when their air-float function is used.
Granite plates are graded by flatness. In the US, ASME B89.3.7-2013 defines grades AA, A and B, each with two limits: the overall flatness of the whole plate and the local flatness over a small area, also called the repeat reading. The table gives one plate size as Mitutoyo tabulates it.
Plates need calibration too. Mitutoyo describes measuring the overall flatness along lines across the plate with an autocollimator, electronic levels, a laser interferometer or a straightedge and indicator, and the local flatness with a repeat reading gauge swept over the surface. A plate that is out can be re-lapped on site.
Keep the plate and the gauge base clean and free of burrs. NPL's good practice guide on dimensional metrology suggests repeating a measurement on different parts of the plate, so that a result does not rest on one spot.
| Grade | Local flatness (repeat reading) | Overall flatness |
|---|---|---|
| AA | 45 µin (about 1.1 µm) | 100 µin (about 2.5 µm) |
| A | 70 µin (about 1.8 µm) | 200 µin (about 5.1 µm) |
| B | 120 µin (about 3.0 µm) | 400 µin (about 10.2 µm) |
How to zero a height gauge
Most height gauges read from the surface plate. Clean the plate and the base, lower the scriber or probe slowly until it just touches the plate and set zero there. Mitutoyo advises a light touch because a heavy hand on the slider can lift the base off the plate.
On a vernier height gauge whose main scale can move, set zero by moving the scale and then tighten its fixing nuts. A digital height gauge zeros at the press of a button or takes a preset value. Mitutoyo's QM-Height manual, for example, sets the plate as 0 mm, or a 25 mm gauge block as +25.000 mm.
Measuring both up and down takes one more step. The probe ball touches a surface that faces down with its top and a surface that faces up with its bottom, so the two readings differ by the ball diameter. Mitutoyo has you measure the ball on a calibration block. TESA finds a probe constant on the master piece supplied with the gauge, which also covers the bending of the probe under the measuring force and the hysteresis of the measuring system. TESA asks for the probe constant to be found again after any change in the measuring conditions, such as switching the gauge off or changing or moving the probe.
Some high-precision gauges have a fixed zero, as TESA lists for its TESA-HITE. TESA's manual strongly recommends not taking the reference directly on the granite plate. Instead, probe a gauge block standing on it and enter the block's size as a preset, so heights still count from the plate.
How to mark out with a height gauge
Fit the scriber, set the height and lock the slider. Mitutoyo advises confirming the setting after you clamp it, because on some height gauges the clamp moves the setting slightly. ISO 13225 treats the effect of slider locking as a separate item, in clause 5.3.
Keep the scriber short and close to the column. Mitutoyo shows that any bend in the column is multiplied by the distance from the column to the tip, so at 150 mm the error is 1.5 times what it is at 100 mm. It also asks for the scriber's measuring face to be parallel to the base within 0.01 mm.
Draw the line by sliding the base across the plate with the scriber point on the part. The tip is very sharp, so handle it with care.
How to measure heights and steps
For a direct height, zero on the plate, bring the probe down onto the surface and read. Use light, even force. Mitutoyo lists excessive measuring force among the causes of error, and its QM-Height beeps when the probe touches with a consistent force.
For the best accuracy, measure by comparison. Fit a test indicator, set it to zero on a stack of gauge blocks of the nominal height, then move it to the part and read the difference. The error of the indicator, or of the height gauge if you move the slider to bring the indicator back to zero, then acts only over that small difference, and the gauge blocks carry the size. Mitutoyo's QM-Height manual describes setting a gauge block as the reference for a dial test indicator in this way.
A step between surfaces that face opposite ways, such as the width of a slot or the distance from the top of a rib to the underside of a flange, needs probing up and down, so the probe diameter has to be allowed for as described above.
Common errors with height gauges
Mitutoyo points out that a height gauge does not follow Abbe's principle: its scale runs beside the line of measurement, not in line with it. Its notes and the other makers' advice come down to these errors.
- Parallax, from reading a vernier or dial at an angle.
- Too much measuring force, which can lift the base off the plate.
- A temperature difference between the gauge and the part. Let them settle together.
- A long scriber or indicator arm, which multiplies any bend in the column.
- Dust or burrs on the plate, under the base or on the scriber mounting.
- A setting that moved when the slider was clamped.
- A probe ball diameter not allowed for when measuring both up and down.
- Moving the gauge by its column. Mitutoyo's QM-Height manual says to move it only by the base or the carrier grip.
How to calibrate a height gauge
Calibrating a height gauge means measuring known heights standing on a surface plate and comparing each reading with the known value. The usual standards are gauge blocks to ISO 3650, a step gauge or a height master, each with its own calibration certificate.
ISO 13225:2012 is the international standard for the design and metrological characteristics of height gauges with analogue or digital indication, and ISO confirmed it in 2023. Its characteristics are limited by maximum permissible errors (MPE), and informative annexes describe the error tests and the calibration. JIS B 7517:2018, the Japanese standard, is a modified adoption of it. As Mitutoyo explains, its 2018 revision renamed the instrumental error as the indication error.
The main characteristic in JIS B 7517 is the height measurement error. Measure a gauge block on a precision surface plate with the contact moving down, subtract the block size and compare the error with the maximum permissible error for that length. The limit covers the straightness and flatness of the measuring face and its parallelism with the base. The table shows some of the limits as Mitutoyo publishes them.
Makers of high-precision gauges also measure each gauge's errors in the factory. For its earlier TESA-hite 400/700, TESA described measuring each assembled gauge step by step against stepped gauge blocks and storing the corrections in the gauge, which then corrected each reading before it was displayed. TESA's manual states the TESA-HITE's performance against the parameters of ISO 13225. For the TESA-HITE it states a maximum permissible error that grows with the measured length and a repeatability limit given as 2σ.
In Germany the test instructions are VDI/VDE/DGQ 2618 Blatt 9.3 (2006), for height calipers with a vernier, dial or digital display to DIN 862 up to 1000 mm, and VDI/VDE/DGQ/DKD 2618 Blatt 16.1 (2025), for height measuring instruments. Its 2023 draft covered vertical length measuring instruments that capture the measurement digitally, including models that also measure straightness and perpendicularity. The 2025 edition adds an example uncertainty budget.
| Measuring length | Scale interval or resolution 0.05 mm | Scale interval or resolution 0.02 or 0.01 mm |
|---|---|---|
| 50 mm or less | ±0.05 mm | ±0.02 mm |
| Over 100 mm, up to 200 mm | ±0.07 mm | ±0.03 mm |
| Over 200 mm, up to 300 mm | ±0.08 mm | ±0.04 mm |
| Over 500 mm, up to 600 mm | ±0.11 mm | ±0.05 mm |
| Over 900 mm, up to 1000 mm | ±0.15 mm | ±0.07 mm |
Perpendicularity, repeatability and the other checks
The height measurement error is the main result, but a full calibration looks at more. Which checks apply depends on the gauge and on what you use it for. The table draws on the standards and the makers' documents, and its methods for perpendicularity and condition are common practice.
Perpendicularity matters most on a gauge used to check squareness. TESA adjusts some of its height gauges, the TESA-HITE among them, for perpendicularity in the frontal plane so that they can check squareness with a lever-type test indicator. It states a frontal perpendicularity limit for the TESA-HITE and gives none for its TESA-HITE MAGNA models. A common check runs an indicator on the slider up a calibrated square standing on the plate.
| Check | How | What it finds |
|---|---|---|
| Height error, downward | Gauge blocks, a step gauge or a height master at points across the range, contact moving down | The error of indication over the range |
| Height error, both directions | Faces of a step gauge or height master probed up and down | The error when surfaces facing opposite ways are measured, including the probe compensation |
| Repeatability | Probe the same face several times | The scatter of the probing |
| Slider locking | Read a setting before and after clamping the slider | A shift caused by the clamp |
| Scriber face | Check its parallelism to the base | A face that marks or measures at the wrong height |
| Perpendicularity | Run an indicator on the slider up a calibrated square | A column that leans or bends |
| Condition | Run the slider over its whole travel and look at the base, the column and the display | Burrs, wear, a slider that sticks or a display that does not hold zero |
How often to calibrate a height gauge
No standard sets one interval. ISO 9001 asks for calibration or verification at specified intervals, or before use. ILAC-G24 and OIML D 10 (2022 edition) list the factors to weigh, such as how much the gauge is used and its environment (5.1), and describe how to review the interval from its calibration history (6).
Between calibrations, measure a gauge block or a height master before an important job and after a knock or a move. ISO/IEC 17025:2017 clause 6.4.10 calls checks like these intermediate checks. Give the surface plate its own interval too, since the gauge cannot be better than the plate.
As-found results and the record to keep
Record the readings before anything is adjusted or repaired. When an as-found error is outside the limit, ISO 9001 clause 7.1.5.2 asks you to determine whether earlier results were adversely affected and to act where needed. Compare the size and sign of the error with the tolerances of the heights the gauge measured or marked since its last good calibration.
- The gauge ID, type, range and resolution, and the scriber or probe fitted.
- The surface plate used and its calibration status.
- The gauge blocks, step gauge or height master used, with their certificate numbers and due dates.
- The error at each point as found and as left, with the direction of probing.
- The temperature, the uncertainty and the decision rule behind any pass or fail statement.
- The date, the person or lab, and the next due date.
How to calibrate a height gauge, step by step
- Check that the gauge blocks, step gauge or height master and the surface plate have current certificates, and that the room is close to 20 °C.
- Clean the plate, the base, the scriber or probe and the standards, and let everything settle to the room temperature.
- Inspect the gauge. Look for burrs on the base, run the slider over its travel and check the scale or display and the clamp.
- Zero the gauge on the plate. If it measures in both directions, set the probe diameter or probe constant as the maker describes.
- Measure the standards at points across the whole range with the contact moving down, and record each error as found.
- If the gauge measures in both directions, probe faces up and down and record those errors too.
- Probe one face several times for the repeatability, and read a setting before and after clamping the slider.
- Compare the errors with the maker's specification or the limits of JIS B 7517, using your documented decision rule.
- If the gauge needs adjustment or repair, have it done, repeat the checks and record the as-left results.
- Record the result, the standards and plate used, the uncertainty and the next due date, and label the gauge with its status.
How Axiospec helps with height gauges
Axiospec is calibration management software. It does not calibrate a height gauge or decide whether it passes. The limits, the decision rule and the interval stay with you or your lab. The height gauge, its test indicator and the surface plate can each be an instrument on the register, with calibration history, certificates and due dates.
Gauge blocks, a step gauge or a height master can sit on the register as reference standards. With a standard such as ISO 9001 or ISO/IEC 17025 selected for the workspace, you pick the one you used from the register when you log a calibration of the height gauge, and its Reverse Traceability card then lists the instruments it was used to calibrate. One record holds up to 100 named test points, each with a nominal value, a tolerance and the as-found and as-left readings, so each height you check can be its own point.
An accepted Fail or Damaged result puts the gauge in quarantine. With a standard selected, a Fail, a Damaged, or a Pass with an as-found reading out of tolerance asks you to write down the impact on the parts it measured, and any not yet written show in Needs attention. In the live demo, a height gauge was calibrated against a gauge block set, and the set's Reverse Traceability card lists it. The free plan covers up to 50 active instruments. A new workspace starts with a 45-day trial of Professional, with room for 1,500 active instruments and no credit card. We load your existing records for you at no charge on every plan.
Common questions
- What is a height gauge used for?
- A height gauge measures and marks heights above a surface plate. With a scriber it marks lines on a part at a set height, and with a probe or a test indicator it measures heights, steps and the distances between features.
- How do you zero a height gauge?
- Clean the surface plate and the base, lower the scriber or probe slowly until it just touches the plate and set zero there, or set a known value on a gauge block. If you measure surfaces that face both up and down, set the probe diameter or probe constant as well.
- How accurate is a height gauge?
- It depends on the type and the length measured. As Mitutoyo publishes the JIS B 7517:2018 limits, a conventional height gauge with a 0.01 mm or 0.02 mm resolution may be off by ±0.02 mm up to 50 mm and by ±0.04 mm over 200 mm up to 400 mm. High-precision gauges are much closer: TESA specifies (2.5 + 4L/1000) µm for its TESA-HITE, with L in mm.
- How do you calibrate a height gauge?
- Stand gauge blocks, a step gauge or a height master on a calibrated surface plate, measure them at points across the range with the contact moving down and compare each error with the maker's limit or JIS B 7517. Gauges that measure up and down are also checked in both directions, for repeatability and for the effect of clamping the slider.
- What is the difference between a height gauge and a height master?
- A height gauge measures parts. A height master is a reference standard with measuring faces at known heights, used to set and check height gauges and other instruments, and it needs a calibration of its own.
- Which standard covers height gauges?
- ISO 13225:2012 covers the design and metrological characteristics of height gauges, and JIS B 7517:2018 is the Japanese standard based on it. In Germany the test instructions are VDI/VDE/DGQ 2618 Blatt 9.3 and VDI/VDE/DGQ/DKD 2618 Blatt 16.1.
Put it into practice
Got a gage list? Send it over and we load it for you, usually in a couple of business days. Free on every plan. Then log every calibration to a tamper-evident audit trail and produce records on demand.