Dial bore gauge: how to set it, measure a bore and calibrate it

12 min read Published Updated

A dial bore gauge is a comparator for holes. You set it to zero on a known size, usually a setting ring, then rock it in the bore and read the turning point of the pointer, which gives the true diameter at that spot as a plus or minus difference from the setting size.

On this page

What is a dial bore gauge?

A dial bore gauge has a moving contact on one side of its head and a fixed anvil on the other. A spring-loaded centring guide keeps the two on a diameter of the bore. The stem passes the movement of the contact up to a dial indicator or a digital indicator at the top. Interchangeable anvils, washers or extensions set the range.

It is a comparator, not an absolute instrument. Mahr's instructions call its bore gauge a comparison measuring instrument that needs a reference of the nominal size. The indicator only shows how far the bore differs from the size the gauge was set to, so a measurement is only as good as that setting. A height gauge with a test indicator works the same way when it is set to zero on gauge blocks.

Two-point gauges with a centring guide are the common kind. Some two-point gauges sit in a ground guide cylinder made for one bore size and need no rocking. Three-point gauges have three contacts 120° apart that centre the head in the bore, as on Mahr's Micromar 844 A inside measuring pistols. The two designs see the shape of a bore differently.

How two-point and three-point gauges see the shape of a bore
GaugeContactsWhat it cannot see
Two-pointA moving contact opposite a fixed anvilAn odd number of lobes, such as a three-lobed bore. Every diameter of that shape reads the same.
Three-pointThree contacts 120° apartA plain oval. Its effect on the three contacts cancels out, so the reading does not change as the gauge turns.
Tip: When the form of a bore matters, measure it on a roundness instrument. A bore gauge reads size, and what it shows of the form depends on its design.

How to set a dial bore gauge to size

Fit the anvil, washers or extension that bring the nominal size inside the travel of the indicator. Use only the parts the maker supplies for that range. Mitutoyo, for example, does not guarantee accuracy with extra sub-anvils and limits how many extension rods may be joined (its catalogue says two at most, its user's manuals one), so check the manual for your model.

Then set the zero on a reference of the nominal size. Put the gauge in, rock it slowly to find the turning point of the pointer, and turn the bezel until zero sits on it. Rock it again to confirm.

Set and measure at the same temperature. Insize asks for the gauge, the setting ring and the part to reach the same temperature first, and Dorsey Metrology advises setting the zero close to the working temperature and checking it often against the master during use. Hold the gauge by its insulated grip so your hand does not warm the stem.

Three ways to set a dial bore gauge
Set onStrengthWatch for
A setting ringIt has the same shape as the bore, so the gauge sits the same way in both and centring errors cancel out. Makers call it the most accurate method.One ring for each size, and each ring needs its own calibration.
Gauge blocks in a holder with jawsAny size from one set of blocks.The contacts must be free of wear, and finding the true axis between flat jaws takes skill.
An outside micrometer locked at the sizeQuick, with no extra equipment.The micrometer's own error and the feel of the person setting it go into the zero.

How to measure a bore: rock for the turning point

Tilt the gauge and put the centring guide and the moving contact into the bore first, then bring the fixed anvil to the wall. Mitutoyo and Insize both describe this order.

Rock the gauge slowly in the plane of the bore axis. The pointer moves, stops and turns back. That turning point is the smallest size the gauge sees, and Mahr and Mitutoyo both give it as the true diameter at that spot. A tilted gauge spans a longer line than the diameter, so every other position reads larger.

The spring-loaded centring guide keeps the contacts close to a diameter, so the rocking that matters is in the plane of the bore axis. Follow your gauge's manual: Mitutoyo's tells you to swing the gauge vertically and horizontally and read the point where the contact is pushed in deepest. In a horizontal bore, Mahr asks for the fixed contact to point down so the gauge is supported correctly.

Reading plus and minus from zero

The reading at the turning point is the difference between the bore and the setting size. Add it to the setting size with its sign. On a gauge zeroed on a 50.000 mm ring, a turning point 0.012 mm on the larger side of zero means a bore of 50.012 mm, and 0.012 mm on the smaller side means 49.988 mm.

Which side is which depends on the indicator and how its dial is numbered, so work it out on your own gauge. When the moving contact is pushed in, the gauge is reading a smaller bore. Press the contact by hand and watch which way the pointer goes. Dorsey Metrology's own check, for example, expects the pointer to move to the minus side when the gauge is pushed off centre in a ring, where the contacts span a chord shorter than the diameter.

Tip: Read the reversal itself. A reading taken before the pointer turns back is larger than the bore.

How to check a bore for taper and out-of-round

One reading gives the diameter at one depth in one direction. To see the shape, measure near the top, in the middle and near the bottom of the bore, and at each depth in two directions at right angles. That is the same pattern of three planes and two directions that DKD-R 4-3 Blatt 4.1 uses for ring gauges.

Taper is the change in diameter from one depth to another in the same direction. Out-of-round, as a two-point gauge sees it, is the difference between the two directions at the same depth. The table shows both.

Example readings in a bore, from a gauge zeroed on a 50.000 mm ring (plus means larger than the ring)
DepthDirection ADirection B, at 90°Out-of-round
Near the top+0.008 mm+0.002 mm0.006 mm
Middle+0.006 mm+0.001 mm0.005 mm
Near the bottom+0.003 mm−0.001 mm0.004 mm
Tip: In this example the bore tapers by 0.005 mm in direction A and 0.003 mm in direction B, and it is larger at the top. Its largest diameter is 50.008 mm and its smallest 49.999 mm.

Common errors with dial bore gauges

Maker instructions warn about these errors.

  • Reading before the turning point, or after it while the gauge is still rocking.
  • Setting the zero at one temperature and measuring at another, or warming the gauge in your hand.
  • Setting between flat jaws with worn contacts. Dorsey Metrology notes that a flat worn on a contact puts an error into a setting made on flat surfaces.
  • Centralizer plungers of unequal length. The gauge then measures a chord rather than the diameter. A setting ring cancels most of this error and a setting on flat surfaces does not.
  • More extension rods or sub-anvils than the maker allows.
  • An indicator clamped with no free gap above its plunger. Dorsey asks for a small gap so that the pointer returns to the same rest position, and for a fresh check after the indicator is moved or the gauge is dropped.
  • A sticky plunger. Push each plunger and the centralizer by hand and feel for free movement.
  • Not checking the zero against the setting ring during a long run of measurements.

How to calibrate a dial bore gauge

Calibrate the gauge in two parts: first the indicator on its own, then the whole gauge. The transfer from the contact to the indicator, the centring and the contacts all add errors of their own.

The indicator is calibrated like any dial indicator, against a dial indicator calibrator at points across its range in both directions. ISO 463:2006 covers mechanical dial indicators, and ASME B89.1.10M-2001, reaffirmed in 2021, is the US standard. A draft revision, ISO/DIS 463, will replace both ISO 463:2006 and ISO 13102:2012 for digital indicators.

For the whole gauge, makers state their own figures. Mahr gives limits for error span, repeatability and hysteresis for its 844 N bore gauges and supplies a test certificate over 10 measuring points. Mitutoyo states wide-range accuracy, adjacent error and repeatability as JIS values, and JIS B 7515:1982 is the Japanese standard for cylinder gauges. In Germany the test instruction is VDI/VDE/DGQ 2618 Blatt 13.2 (2005), for bore gauges with two-point contact.

A check against setting rings covers the same ground. Set the gauge on one calibrated ring and read a second ring of known size within the travel. The reading should equal the difference between the two certified sizes, and rings at several sizes cover the working range. The rings are reference standards and need their own calibration, which EURAMET Calibration Guide No. 6 describes.

Checks on a dial bore gauge and what each one finds
CheckHowWhat it finds
IndicatorCalibrate it on its own at points across its range, in both directionsThe error of indication, hysteresis and repeatability of the indicator
Free movementPush each plunger and the centralizer by handSticking or a damaged mechanism
RepeatabilityZero in a ring and take repeated readings, rocking through the turning point from both sidesA worn plunger bearing, which Dorsey says shows as readings that do not repeat
CentringZero in a ring, then push the gauge off centre to one side and then to the otherCentralizer plungers of unequal length. The pointer should move to the smaller side both times.
Error over the travelZero on one calibrated ring and read others of known size within the travelThe error of the transfer and the contacts across the working range
Tip: Calibration checks the gauge against rings. A gage R&R study on your own parts shows how much variation the gauge and the people using it add.

How often to calibrate a dial bore gauge

No standard sets one interval. ISO 9001 asks for calibration or verification at specified intervals, or before use. ILAC-G24 and OIML D 10 (2022 edition) list the factors to weigh, such as how much the gauge is used, its environment and its calibration history.

Zeroing on the setting ring before each use checks the setting, not the error over the travel. It does not replace the periodic calibration of the indicator and of the whole gauge, or the calibration of the rings. Check the gauge again after a drop or after the indicator has been moved. ISO/IEC 17025:2017 clause 6.4.10 calls checks like these intermediate checks.

As-found results and the record to keep

Record the results before anything is cleaned, adjusted or repaired. When an as-found error is outside the limit, ISO 9001 clause 7.1.5.2 asks you to determine whether earlier results were adversely affected and to act where needed. Compare the size of the error with the tolerances of the bores the gauge measured since its last good calibration.

  • The gauge ID and range, and the ID and scale interval of its indicator.
  • The anvils, washers or extensions fitted during the calibration.
  • The setting rings or gauge blocks used, with their certificate numbers and due dates.
  • The results of each check, as found and as left.
  • The temperature, the uncertainty and the decision rule behind any pass or fail statement.
  • The date, the person or lab, and the next due date.

How to calibrate a dial bore gauge, step by step

  1. Check that the setting rings and the dial indicator calibrator have current certificates and that the room is close to 20 °C.
  2. Clean the gauge, check that each plunger and the centralizer move freely, and look for damage to the contacts.
  3. Calibrate the indicator on its own, or confirm that it has a current calibration.
  4. Refit the indicator with the gap or preload the maker specifies, and fit the anvils for the range you will check.
  5. Let the gauge and the rings settle to the same temperature.
  6. Zero the gauge on a calibrated ring at the turning point. Take repeated readings for the repeatability, and push the gauge off centre each way to check the centring.
  7. Read further calibrated rings of known size across the working range and work out the error at each.
  8. Compare the results with the maker's specification or your own limits, using your documented decision rule.
  9. If anything needs repair or adjustment, have it done, then repeat the checks and record the as-left results.
  10. Record the result, the rings used, the uncertainty and the next due date, and label the gauge with its status.

How Axiospec helps with dial bore gauges

Axiospec is calibration management software. It does not calibrate a bore gauge or decide whether it passes. The limits, the decision rule and the interval stay with you or your lab. The bore gauge and its indicator can each be an instrument on the register, with calibration history, certificates and due dates.

Setting rings can sit on the register as reference standards with their own certificates and due dates. With a standard such as ISO 9001 or ISO/IEC 17025 selected for the workspace, you pick the ring, or a set of rings tracked as one instrument, from the register as the reference standard when you log a calibration of the bore gauge. Its Reverse Traceability card then lists the instruments it was used to calibrate. One record holds up to 100 named test points, each with a nominal value, a tolerance and the as-found and as-left readings, so each ring size can be its own point.

An accepted Fail or Damaged result puts the gauge in quarantine. With a standard selected, a Fail, a Damaged, or a Pass with an as-found reading out of tolerance asks you to write down the impact on the bores it measured, and any not yet written show in Needs attention. The live demo has a bore gauge that failed as found, with its impact written. The free plan covers up to 50 active instruments. A new workspace starts with a 45-day trial of Professional, with room for 1,500 active instruments and no credit card. We load your existing records for you at no charge on every plan.

Common questions

Is a dial bore gauge an absolute or a comparative instrument?
It is a comparator. It shows only the difference between the bore and the size it was set to, so it has to be zeroed on a known reference, such as a setting ring, before it measures anything.
How do you zero a dial bore gauge?
Put it in a setting ring of the nominal size, rock it slowly to find the turning point of the pointer and turn the bezel so zero sits on that point. Rock it again to confirm, with the gauge and the ring at the same temperature.
Why do you rock a dial bore gauge?
A tilted gauge spans a longer line than the diameter, so its reading is too large. Rocking it along the bore axis passes through the square position, where the pointer turns back, and that turning point gives the true diameter.
How do you check a bore for taper and out-of-round?
Measure near the top, in the middle and near the bottom of the bore, in two directions at right angles at each depth. Taper is the change between depths in one direction, and out-of-round is the difference between the two directions at one depth.
How do you calibrate a dial bore gauge?
Calibrate the indicator on its own against a dial indicator calibrator, then check the whole gauge. Zero it on a calibrated setting ring, check its repeatability and centring, and read other rings of known size across its travel to find the error over its range.
What is the difference between a two-point and a three-point bore gauge?
A two-point gauge measures between two opposite contacts and shows ovality, but it cannot see a bore with an odd number of lobes. A three-point gauge has three contacts 120° apart and centres itself, but it cannot see a plain oval.