Hygrometer calibration: humidity sensors, salt solutions and dew point references
To calibrate a hygrometer, put its sensor in a stable humidity at a known temperature next to a calibrated reference, such as a chilled-mirror dew-point hygrometer, or in a humidity generator, and compare the readings at three or more humidity points across the range you use once both have settled. Control and measure the temperature as closely as the humidity: at 50 %rh and 20 °C, a 1 °C rise lowers relative humidity by about 3 %rh (NPL).
On this page
Relative humidity or dew point: what are you calibrating?
Relative humidity compares the water vapor in the air with the most it could hold at the same temperature and pressure, as a percentage. That is the WMO definition NIST uses for its calibrations (NIST SP 250-83r1, section 3). Dew point is the temperature at which that vapor would start to condense at the same pressure. Below 0 °C the condensate is ice and the term is frost point.
The two behave differently when air warms or cools. Dew point stays the same, because the amount of water has not changed. Relative humidity falls as the air warms and rises as it cools. A table in NPL's beginner's guide to humidity measurement shows air with a dew point of 10 °C at about 72 %rh at 15 °C, 53 %rh at 20 °C and 39 %rh at 25 °C.
So calibrate in the quantity the instrument reports, as NPL advises, and always record the temperature. DKD-R 5-8:2019, the German Calibration Service guideline published by PTB for hygrometers that measure relative humidity directly, defines the aim as the deviation of the indicated relative humidity from the reference value, stated with its uncertainty, the conditions and the procedure (clause 4).
- Capacitive polymer and resistive sensors are the electronic probes inside most thermo-hygrometers and humidity data loggers. DKD-R 5-8 covers these and resistive-electrolytic sensors (7.1.1).
- Mechanical hygrometers, such as hair hygrometers, are covered too (clause 8), but DKD-R 5-8 does not recommend them as reference standards (5.6).
- Dew-point hygrometers and psychrometers work out relative humidity from other measurements, so DKD-R 5-8's procedure for direct-reading instruments does not apply to them (7.1).
Which reference should you use?
DKD-R 5-8 requires every reference or working standard to be traceable to a national standard and compares the hygrometer directly with it (clause 5). The references differ mainly in their uncertainty and in where you can use them.
NIST's Hybrid Humidity Generator shows what a primary generator covers. It produces dew and frost points from −90 °C to +85 °C and calibrates relative humidity sensors from 2 %rh to 98 %rh at temperatures from −34 °C to 85 °C (NIST SP 250-83r1, 2021). Above a frost point of −10.3 °C it works as a two-pressure generator.
| Reference | How it works | Where it fits |
|---|---|---|
| Two-pressure or two-temperature humidity generator | Saturates air at a measured temperature and pressure, then expands it to the chamber pressure or brings it to the chamber temperature. The humidity follows from those temperatures and pressures. | The lowest uncertainties: national institutes and calibration laboratories |
| Chilled-mirror dew-point hygrometer | Cools a mirror until dew or frost forms and measures the mirror temperature. Relative humidity is calculated from that dew point and the air temperature and pressure. | A reference inside a climatic chamber or a humidity calibrator |
| Calibrated relative humidity probe | Read side by side with the hygrometer under test | A working standard in the lab or on site, with frequent checks |
| Saturated salt solution | Holds a fixed relative humidity in a small sealed space above the solution at a given temperature | Checks between calibrations and simple field use |
What does each reference need?
- A relative humidity probe used as a standard needs a partner. DKD-R 5-8 asks for at least two, preferably of different types, recalibration at least every 12 months and intermediate checks no more than a month apart (5.5.2).
- A chilled-mirror hygrometer used as a standard is recalibrated at least every 24 months, and its mirror has to be checked and cleaned regularly (5.3.2). NIST cleans customer mirrors with ethanol and then distilled water before calibrating them (SP 250-83r1, 9.1).
- In a climatic chamber, use standards that are independent of the chamber's own control sensor, placed in the calibration volume with the item (6.1.2).
Saturated salt solutions: which humidity at which temperature?
Above a saturated solution of a salt in water, a sealed space settles at a relative humidity fixed by the salt and the temperature. The solution stays saturated as long as undissolved salt remains, even if a little water is gained or lost, which is what makes the method simple (Greenspan, 1977). ASTM E104-20a is the standard practice. It covers small closed containers, humidities from dry to near saturation and temperatures from 0 °C to 50 °C, for conditioning and for calibrating hygrometers, and it does not recommend the method for flowing air unless precautions keep the source stable.
The values below come from Lewis Greenspan's evaluated compilation at the US National Bureau of Standards, now NIST (J. Res. NBS 81A, 1977, Table 2). The table gives seven common salts at 20 °C and 25 °C. The ± figure is Greenspan's estimated uncertainty: three times the standard deviation of the fitted value, or the estimated uncertainty of the original data where that was larger. It does not include the uncertainty of the vapor pressure equations behind the values.
| Salt | At 20 °C | At 25 °C |
|---|---|---|
| Lithium chloride (LiCl) | 11.31 ± 0.31 | 11.30 ± 0.27 |
| Magnesium chloride (MgCl₂) | 33.07 ± 0.18 | 32.78 ± 0.16 |
| Potassium carbonate (K₂CO₃) | 43.16 ± 0.33 | 43.16 ± 0.39 |
| Magnesium nitrate (Mg(NO₃)₂) | 54.38 ± 0.23 | 52.89 ± 0.22 |
| Sodium chloride (NaCl) | 75.47 ± 0.14 | 75.29 ± 0.12 |
| Potassium chloride (KCl) | 85.11 ± 0.29 | 84.34 ± 0.26 |
| Potassium sulfate (K₂SO₄) | 97.59 ± 0.53 | 97.30 ± 0.45 |
How do you use a salt solution as a check?
- Keep the temperature stable and the set-up sealed and clean. Make sure equilibrium has been reached above the solution and that no salt aerosol reaches the sensor. With no air circulation, expect a long settling time (DKD-R 5-8, 6.1.4).
- ASTM E104 warns that saturated salt solutions are highly corrosive and that the air above them may carry corrosive vapor.
- Under DKD-R 5-8, a salt's table value is not a reference. Clause 5.7 does not accept salt solutions as reference material for traceability but allows them, carefully applied, for intermediate checks. Clause 6.1.4 does allow a salt cell as the humidity source in a calibration when a separate calibrated reference measures the humidity. NPL's beginner's guide treats salt solutions as a field calibration method that can be made traceable. Decide which applies in your procedure.
Why temperature matters as much as humidity
Relative humidity is relative to temperature. NPL's beginner's guide gives the scale: at 50 %rh, warming the air from 20 °C to 21 °C lowers relative humidity by about 3 %rh. A sensor a fraction of a degree warmer than the reference therefore reads low, even if both are perfect. DKD-R 5-8 requires all thermal influences to be considered, not only the hygrometric ones (clause 5).
- Thermal coupling. Put the whole sensor, with at least 1 m of its cable where it has one, inside the calibration volume. DKD-R 5-8 notes that heat conducted along the probe can reach about 10% of the difference between the calibration temperature and the room (7.4.4).
- Self-heating. A probe's own electronics can warm it. DKD-R 5-8 estimates this from the change between a reading just after switching on and one after at least 30 minutes, or by comparing very different air speeds (7.4.3).
- Radiation. Warm or cold walls, lamps and heaters change the temperature a sensor settles at, and DKD-R 5-8 counts the whole radiation effect in the uncertainty (6.2.3).
- Condensation. Keep the reference and the item warmer than the dew point in the device at every point, including when you take them out (9.3).
How long should a hygrometer settle before you read it?
Longer than most people expect. DKD-R 5-8 says a stabilization time of at least 3 hours may be needed, depending on the air flow and the design of the instrument, and longer at low flow (9.4). Before a calibration it asks for the hygrometer to be unpacked and kept at room conditions for at least 24 hours (7.2.2).
It proposes a check you can repeat at each point. Start the clock when the reference is within 2 %rh and 0.2 K of the target. Wait 30 minutes, then record for at least 10 minutes and at least 10 values. Wait another 10 minutes and record again. The hygrometer has settled when its error changed by less than 20% of the uncertainty you aim for between the two records and the temperatures agree as well. The second record is the result (9.4.1).
NIST works the same way. It records readings every 90 seconds for at least 30 minutes after everything reaches a steady state and averages the last 30 minutes (SP 250-83r1, 10.2).
Which test points, and in what order?
DKD-R 5-8 recommends at least three humidity points at each temperature, covering the range the hygrometer is used over, for example 20 %rh, 50 %rh and 80 %rh (9.3). With more than one temperature it starts at the highest and works down, finishing all the humidity points at one temperature before moving on. Above 40 °C it starts with a point near room temperature, 23 °C ± 3 K, to show whether the high temperature changed the sensor.
The order of the humidity points matters because many sensors show hysteresis: they read differently at the same humidity when it is rising than when it is falling. DKD-R 5-8 defines seven sequences. It names B2 as the usual one and A1 or A2 for high demands (9.3). Only A1 and A2 measure hysteresis (7.4.5).
| Sequence | Order of the humidity points | Shows hysteresis? |
|---|---|---|
| A1 | Rising from a lower humidity, then falling back through the same points | Yes |
| A2 | First point approached from room humidity, then rising, then falling | Yes, plus repeatability at the lowest point |
| B1 | Rising, with the first point approached from below | No |
| B2 | Rising, with the first point approached from room humidity. The usual choice | No |
| C1 and C2 | Falling, with the first point approached from above or from room humidity | No |
| D | An order agreed with the customer | As agreed |
How much do humidity sensors drift?
Enough to calibrate them regularly. NPL's beginner's guide says hygrometers can suffer from bias and long-term drift, from their conditions of use or simply from aging, and that older sensors may also respond more slowly. Using a sensor outside its specified temperature or humidity range can shift its calibration, and NPL advises keeping sensors away from condensation unless they are made to tolerate it.
DKD-R 5-8 gives working figures for probes used as reference standards. Without intermediate checks it counts a drift of at least 2 %rh over a 12-month interval, and if hysteresis is not measured, at least 3 %rh (5.5.2 and 5.5.3). To limit drift from contamination, it asks for clean equipment, careful inspection of each item, suitable storage and packaging that gives off nothing harmful to the sensor (5.5.2 and 7.3).
For scale, NPL suggests that relative humidity measured in ideal conditions might reach an uncertainty of ±2 %rh to ±3 %rh, and ±5 %rh to ±10 %rh or worse in difficult conditions, all at 95% confidence.
Field checks or laboratory calibration?
A check is not a calibration, as NPL puts it, but checks between calibrations are very useful. They show whether a hygrometer still works as expected and whether it needs repair or recalibration.
| Question | Field check | Laboratory calibration |
|---|---|---|
| Reference | A calibrated reference probe at the same spot, or a salt cell at a stable temperature | A humidity generator, or a chamber with a traceable reference |
| Points | Usually one or two near the working conditions | Three or more per temperature across the range |
| Result | Still within your check limit, or not | The error at each point with its uncertainty, on a certificate |
| Due date | Unchanged | Set again from the calibration date |
What changes for warehouses, printing, electronics and test chambers?
- Warehouses and storage. Humidity data loggers are often spread across a large site. Calibrate them at the temperatures they actually see, not only at room temperature, and keep a calibrated reference probe for spot checks between calibrations.
- Printing and electronics assembly. Humidity is a process condition there. Calibrate around the humidity and temperature the process runs at, and record which instrument monitors each area.
- Environmental and climatic test chambers. The chamber's own sensor controls it, so check it against an independent reference in the working volume and map the volume before you rely on it for calibration (DKD-R 5-8, 6.1.2 and 6.2.1).
- Metrology and calibration rooms. The room's thermo-hygrometer supplies the conditions written on calibrations made there, so its own calibration supports all of those records.
How often should a hygrometer be calibrated?
No standard fixes one interval for every hygrometer. NPL suggests that relative humidity instruments might be calibrated every 6 to 12 months, more often if treated harshly, and a reference hygrometer in a laboratory every 1 to 2 years. How often depends on use, the risk of drift and how critical the measurement is.
For instruments used as reference standards, DKD-R 5-8 sets upper limits: 12 months for relative humidity probes, with intermediate checks no more than a month apart, and 24 months for chilled-mirror hygrometers and psychrometers (5.5.2, 5.3.2 and 5.4.2). ILAC-G24:2022 / OIML D 10:2022 lists the factors for choosing an interval (5.1), and its clause 6 describes how to review it from your own results.
What should the certificate and record show?
DKD-R 5-8 (clause 11) adds most of the items below to the general content required by ISO/IEC 17025:2017 clause 7.8. The first and last items are good practice for your own records rather than DKD requirements, and a lab certificate shows a due date only if you agreed it (ISO/IEC 17025 clause 7.8.4.3).
- The instrument, with the probe and the display unit identified separately and, on a multi-channel unit, which probe was on which channel.
- Any pre-treatment, the method and sequence used, and the order of the points.
- How far into the calibration volume the instrument was installed.
- The air temperature at each point, the reference humidity, the indicated humidity, the error and the uncertainty.
- Whether hysteresis or repeatability is included in the uncertainty, and whether the humidity is relative to water or to ice.
- The room temperature, and the room humidity where it matters.
- For an analog output such as 4 mA to 20 mA, the signal measured with a calibrated multimeter and the scaling used to convert it (7.4.7).
- The readings before any adjustment and after it, the date, who did the work and the next due date.
What if the as-found reading is out of tolerance?
Record the readings before anyone adjusts the hygrometer. If an as-found error is outside its limit, ISO 9001 clause 7.1.5.2 asks you to determine whether earlier results were adversely affected and to act as needed. For a hygrometer, look at what its readings decided since its last good calibration or check, such as goods accepted from storage, a chamber test run, a print or coating batch, or the conditions written on calibrations made in a metrology room.
How to calibrate a hygrometer, step by step
- Identify the hygrometer, its probe and display unit, the range it is used over and the limit that applies, from the maker's specification or your own requirement.
- Inspect the probe and its filter for damage and dirt, check that it works and keep it unpacked at room conditions for at least 24 hours.
- Choose at least three humidity points across the range in use, at the temperature or temperatures of use, and decide whether you need rising and falling points to see hysteresis.
- Check the reference's certificate and due date, and that its uncertainty suits the limit. For a salt check, use a solution with undissolved salt at a stable temperature.
- Mount the probe close to the reference, with the whole probe and some of its cable inside the calibration volume, and keep everything above the dew point.
- At each point, wait until the reading has settled, then record the reference humidity, the air temperature and the hygrometer reading.
- Work out the error at each point and compare it with the limit. Record these as-found results before any adjustment.
- If the hygrometer needs adjusting, adjust it as the maker describes and repeat the points as left.
- If an as-found result was out of tolerance, assess what the hygrometer measured since its last good check. Then set the next due date.
Where Axiospec fits
Axiospec is calibration management software. It keeps each hygrometer, thermo-hygrometer and humidity data logger with its calibration history, certificates and next due date, and the interval can be set in days, months or years. One calibration record holds up to 100 test points, each with its nominal value, its own tolerance, a name such as the air temperature, whether it was read up-scale or down-scale, and the as-found and as-left readings. With a standard such as ISO 9001 or ISO/IEC 17025 selected for the workspace, each calibration also keeps the room temperature and humidity and names the reference standard used, as free text or as a link to a master instrument.
On an in-house calibration, a final reading outside its tolerance sets the result to Fail, and recording a pass anyway takes a written reason. An accepted Fail or Damaged puts the instrument in Quarantined. With a standard selected, any approved Fail, Damaged or Pass with as-found out of tolerance needs an impact written, and unrecorded ones show in Needs attention. Axiospec does not run the calibration or pick the points, the limit or the interval. Those stay with you or your lab.
The free plan covers up to 50 active instruments. A new workspace starts with a 45-day trial of Professional, with room for 1,500 active instruments and no credit card. We load your existing records for you at no charge on every plan.
Common questions
- How do you calibrate a hygrometer?
- Place it next to a calibrated reference, such as a chilled-mirror dew-point hygrometer or a reference probe, in a stable humidity at a known temperature, or in a humidity generator. Let both settle, then compare the readings at three or more humidity points across the range you use. Record the error at each point against the hygrometer's limit.
- Can you calibrate a hygrometer with salt?
- A saturated salt solution in a sealed container holds a known relative humidity, for example 75.3 %rh over sodium chloride at 25 °C (Greenspan, 1977), so it makes a good check. DKD-R 5-8 accepts salts for intermediate checks but not to establish traceability, so under that guideline a salt test that relies on the table value is a check rather than a full calibration.
- How often should a hygrometer be calibrated?
- NPL suggests every 6 to 12 months for relative humidity instruments, more often if they are treated harshly, and every 1 to 2 years for a laboratory reference hygrometer. Adjust the interval from each instrument's as-found history and how critical its readings are.
- Why does temperature matter when calibrating a hygrometer?
- Relative humidity depends on temperature. At 50 %rh, a 1 °C rise from 20 °C lowers it by about 3 %rh, so a small temperature difference between the hygrometer and the reference shows up as a humidity error. The temperature has to be stable, measured and recorded at each point.
- What is the difference between relative humidity and dew point?
- Relative humidity is the water vapor in the air as a percentage of the most it could hold at that temperature, so it changes when the air warms or cools. Dew point is the temperature at which the vapor would condense, and it stays the same when the temperature changes.
- How long does a humidity sensor take to stabilize?
- Often longer than expected. DKD-R 5-8 says at least 3 hours may be needed at a calibration point, depending on the air flow and the instrument. It treats a reading as settled when it changes by less than 20% of the target uncertainty between two records separated by a 10-minute wait.
Put it into practice
Got a gage list? Send it over and we load it for you, usually in a couple of business days. Free on every plan. Then log every calibration to a tamper-evident audit trail and produce records on demand.
Axiospec is a documentation and workflow tool. It helps you keep clean, traceable, audit-ready records. Certification depends on your own processes, scope and assessor.
Take the audit-prep checklist with you
First the checklist, with each item phrased the way an auditor asks it. Then four short emails with practical tips. You can unsubscribe in one click.
First we email you a link to confirm your address. Nothing else comes until you confirm, and you can unsubscribe from any email after that. See our Privacy Policy.