Coordinate measuring machine (CMM): how it measures and how it is verified
A coordinate measuring machine (CMM) measures a part by probing points on its surface, recording the X, Y and Z coordinates of each point and fitting features such as planes, circles and cylinders to them. It is verified with the ISO 10360 tests, which measure calibrated test lengths and a test sphere to show the machine still meets its maximum permissible errors, and it is checked between those tests on an artifact such as a ball bar or a check part.
On this page
What is a coordinate measuring machine?
A CMM works out the geometry of a part from points on its surface. ISO 10360-1 describes it, in NPL's summary, as a measuring system that can move a probe and find the coordinates of points on a workpiece. Scales on the axes report where the probe is, and the probe signals the moment it meets the surface.
Software then fits features such as planes, circles, cylinders and cones to the points, aligns them to the part's datums and reports size, position and form against the drawing or the CAD model. The touch-trigger probe, invented in 1972 by David McMurtry at Rolls-Royce, was a major factor in making fast automated inspection on these machines possible (NPL Good Practice Guide No. 43).
Most shop-floor CMMs are Cartesian, with three linear axes at right angles. ISO 10360-2 is written for these machines with contacting probes that take discrete points. Articulated arm CMMs have their own part, ISO 10360-12:2016, and other parts of the series cover further sensors and machine types.
| Probe | How it takes points | Strengths and limits |
|---|---|---|
| Touch-trigger probe | Records one point each time the stylus touches the surface and the probe triggers. The machine then backs off and moves to the next point. | Simple and robust. The stylus bends slightly before it triggers, by an amount that depends on direction, and qualification on the reference sphere accounts for this. |
| Scanning (analogue) probe | Stays in contact with the surface and measures how far the stylus is deflected, so it sends a continuous stream of points. | Many points per feature describe form far better than a few touches, and the measurement is more repeatable for size and position. |
| Non-contact optical sensor | A laser or camera measures the surface without touching it. | Fast, with many points and no contact force. NPL notes that traceability is harder to demonstrate and that surface finish and material affect the result. |
How many points does a feature need?
Enough to show its form, not just its size. Three points define a circle, but they say nothing about how round it is. NPL's Good Practice Guide No. 41 quotes BS 7172, which recommends at least 7 points on a circle, so that up to six lobes can be detected, and 9 on a plane.
More points take longer, so a measurement plan is a trade between time and confidence. Write the number and spread of points into the part program so that every run measures the feature the same way.
Why is a CMM result only as good as the machine?
Each moving axis of a CMM can go wrong in six ways: one positioning error along its travel, two straightness errors across it and three angular errors, called roll, pitch and yaw. With three axes and the three squareness errors between them, NPL counts 21 kinematic error sources. Makers measure these errors and correct them in software, which is often called error mapping.
The result also depends on things outside the machine's geometry: the temperature and its gradients, vibration, the probing strategy and force, the workpiece itself and the software that fits features to the points (NPL Good Practice Guide No. 42). The ISO 10360 tests sample the machine's length measuring performance. NPL stresses that they cannot give the uncertainty of every task, because the errors combine in complicated ways.
Three words get mixed up around CMMs. NPL's guide separates them as the table shows.
| Term | What it means | How often |
|---|---|---|
| Probe qualification | Measuring the reference sphere to find the effective size of each stylus tip and the offsets between tips | Day to day |
| Verification | Testing whether the CMM still meets its specification, usually with the ISO 10360 tests | At periodic intervals, often once a year |
| Calibration (error mapping) | Measuring all 21 kinematic error sources so they can be corrected | At installation, then as needed |
How to qualify the probe and keep it reliable
Qualification corrects each probed point for the size of the stylus tip, so a wrong qualification spoils every measurement until the next one. Qualify on the reference sphere supplied with the machine, with its latest calibrated diameter entered in the software. NPL notes that entering the wrong sphere value is a common error, and that ISO 10360 results only hold for measurements made with the reference sphere used in the test.
Approach the sphere the way you approach the part, with the same probing force and speed. NPL recommends 25 to 50 touch points per stylus tip for a reliable qualification.
- Keep the stylus simple, short and stiff, with the largest tip the features allow. A single straight stylus generally performs better than one with bends and joints.
- Check that each tip is tight on its stem, especially after a collision. A loose tip gives variable results and poor qualification data.
- Clean the tips and the reference sphere before qualifying. A stylus left on the machine overnight can pick up dust and fibres.
- Record the effective tip diameter at each qualification and compare it with earlier values. If it has changed, find out why before measuring.
- After qualifying, measure a known artifact such as a calibrated gauge block or setting ring. If the error is outside the machine's specification at that length, requalify and check again.
Why temperature matters most
NPL calls temperature variation the most significant factor affecting CMM accuracy, because it changes the size of the part and of the machine. Dimensions are defined at 20 °C, the standard reference temperature in ISO 1:2022, and steel grows by about 11.5 µm per metre for each degree.
NPL's worked example: a steel bar measured as 300.015 mm at 23.4 °C is 300.003 mm at 20 °C. CMM software can correct for part temperature, but NPL points out that the correction still leaves an uncertainty from the temperature reading and from the expansion coefficient behind it.
- Let a part that is warm from machining sit until it has reached the room's temperature before you probe it.
- Keep temperature gradients and changes over time small, and run the machine close to 20 °C.
- Without a controlled room, keep doors and windows shut during a measurement, leave the heating off and shield the machine from sunlight and other heat sources.
- Let styli reach room temperature before qualification. A 100 mm steel stylus shaft changes length by about 3.6 µm for a 3 °C change (NPL Good Practice Guide No. 43). Gloves shorten the wait, and a room that fluctuates needs more frequent requalification.
What do E0, E150, R0 and the probing error mean?
A CMM's accuracy is stated as maximum permissible errors (MPEs) for the ISO 10360 tests. ISO 10360-2:2009 covers length measurement with contacting probes, and ISO 10360-5:2020 covers probing systems with one or more styli. In the US, ASME B89.4.10360.2-2008 is a technical report that takes in the whole of ISO 10360-2, adds US requirements in text boxes and replaces B89.4.1-1997. In Germany, VDI/VDE 2617 Blatt 2.1 (2014) is a guide to applying DIN EN ISO 10360-2.
| Symbol | What it is | How it is found |
|---|---|---|
| E0 | Length measurement error with no stylus tip offset from the ram axis, in practice as small as possible | Measuring calibrated test lengths in several positions and directions in the measuring volume |
| E150 | Length measurement error with the stylus tip offset 150 mm from the ram axis | The same kind of measurement, repeated with a 150 mm stylus offset |
| R0 | Repeatability range of the length measurement error | The spread of the three repeated measurements of each test length |
| P (probing error) | How far points on a calibrated test sphere scatter about the fitted sphere | Probing a test sphere with one stylus and taking the range of the radial distances |
How the ISO 10360 tests are run
The length test measures calibrated test lengths traceable to the metre. Gauge blocks, a step gauge, a ball bar or a laser interferometer with a gauge block can provide them. They are normally probed from opposite directions at the two ends, and measurements probed in one direction must be backed by some probed in both. NPL's guide to the 2009 edition describes the core of the test: five test lengths, each measured three times, in seven positions. Four positions are the space diagonals of the measuring volume, and the user chooses the other three, with defaults parallel to the machine axes. That makes 105 measurements.
NPL adds that the longest test length should be at least 66% of the longest space diagonal of the machine's operating volume and the shortest under 30 mm. Each of the 105 errors is compared with E0,MPE, taking the test uncertainty into account under the ISO 14253-1 decision rules, and R0 is compared with its own limit. The standard allows a limited number of test lengths with one failing value to be measured again.
NPL's 2011 guide describes the probing test as it stood in ISO 10360-5:2010, the part it moved to from ISO 10360-2 with the 2009 edition. It uses a calibrated test sphere between 10 mm and 50 mm in diameter, not the reference sphere used for qualification. The 25 points are spread over at least a hemisphere and fitted with a least-squares sphere, and the range of their radial distances is the probing error. ISO 10360-5 is now in its 2020 edition, so check that edition for the current test parameters.
- The acceptance test is run at installation and after an overhaul or upgrade, against the maker's MPEs and in the environment the maker specifies.
- The reverification test repeats the procedure periodically. The user may state the limits, for example to suit the age and condition of the machine, and under the 2009 edition these cover E0, R0 and E150. The user may also specify the environment for the test.
- If the reference sphere is damaged, replace it with one of similar material and specification and repeat the ISO 10360 test.
Interim checks between verifications
Annex A of ISO 10360-2:2009 strongly recommends checking the CMM regularly between reverifications, at an interval the user sets from the performance required, the environment and how much the machine is used. Measure the check artifact right after a verification to get its baseline. Check again after any event that could affect the machine, such as a collision, a heavy impact load or a large temperature change (NPL Good Practice Guide No. 42).
Use the probing strategy and speeds you use on parts, and an artifact whose thermal expansion is similar to your typical workpieces. To compare an interim check with the maker's E0,MPE, use a calibrated test length and the ISO 10360 procedure. A short version, such as one long test length along each space diagonal, generally reveals more than five lengths along one axis.
| Artifact | Strengths | Limits |
|---|---|---|
| Purpose-made test piece or calibrated replica of a typical part | Tests the machine and its software on the kinds of features you measure | Can be hard and costly to make and calibrate, and a new task may need a new piece |
| Ball bar | Low cost, portable, stable and robust | Its length comes from computed sphere centres, and it repeats a task rarely met on real parts |
| Bar located kinematically between a fixed sphere and the stylus sphere | Quick to use, and arms of different lengths cover a large volume | Needs a machine that can move in a circle under program control, and tests one kind of task |
| Plain setting ring | A simple check of size and of the probe qualification on a familiar feature | Needs a ring of a grade suitable for the CMM under test |
| Ball plate or hole plate | Stable, and can be measured in several orientations | A ball plate is heavy, and it will not reveal a qualification problem unless the balls are calibrated for size |
What ISO 10360 does not tell you: task-specific uncertainty
Passing ISO 10360 shows that the machine meets its specification for length measurement. It does not give the uncertainty of a particular feature on a particular part, and NPL stresses that the tests do not guarantee traceability for every measurement task.
Two documents cover that step. ISO 15530-3:2011 uses calibrated workpieces: you measure a calibrated part of similar size, form and material with the same strategy and compare the result with its certificate. ISO/TS 15530-4:2008 covers simulation software, often called a virtual CMM, that estimates the uncertainty of a given task.
For a measurement that decides acceptance, a study on your own parts adds what the tests cannot. A Type 1 study on a calibrated reference part shows bias and repeatability on one feature, and a gage R&R study shows how much of the observed variation comes from the measurement system.
How often should a CMM be verified?
No standard fixes one interval. ISO 10360-2 leaves reverification to the user's requirements and the use of the machine, and NPL notes that verification is often done once a year. ISO 10360-12 says outright that it does not specify how often, or whether, arm CMMs are tested.
Set the interval from the machine's history and how critical its results are. ILAC-G24:2022 / OIML D 10:2022 lists the factors to weigh. Interim checks between visits let you hold an interval with confidence or show you when to shorten it.
- Check or reverify after a collision, a move, a repair or a change to the error map.
- Shorten the interval when interim checks drift toward the limit or the room's temperature control changes.
- Lengthen it only on a record of as-found results well inside the limits.
What the verification record should show
A good record lets anyone see what was tested, with what, and against which limits. The list below is good practice built around the ISO 10360 tests, not a clause-by-clause requirement.
- The CMM's ID, its software version and the probe, stylus and reference sphere used.
- The artifacts used, with their certificate numbers, calibrated values and due dates.
- The temperatures of the machine and the artifacts during the test.
- E0, E150, R0 and the probing error, each against its limit, and whether the limits are the maker's or your own.
- The test uncertainty and the decision rule behind each pass or fail.
- The results as found before any adjustment or new error map, and as left after it.
- The date, the person or company, and the next due date.
What if the CMM fails as found?
Record the as-found results before anyone adjusts or remaps the machine. If an error is outside its limit, ISO 9001 clause 7.1.5.2 asks you to determine whether earlier results were adversely affected and to act as needed. Look at the parts accepted on the CMM since its last good interim check, starting with the features whose tolerance is closest to the size of the error. Regular interim checks keep that window short.
How to run a CMM interim check, step by step
- Choose a calibrated check artifact that suits your work, such as a ball bar, a setting ring or a check part, and confirm that its certificate is current.
- Place it in a fixed, repeatable position and let it reach the temperature of the machine. Record both temperatures.
- Clean the stylus, the reference sphere and the artifact, then qualify the probe as you do for production.
- Run a saved program with the probing strategy and speeds you use on parts.
- Compare the results with the artifact's certified values and with your check limit, and add them to the control chart.
- If a result is outside the limit, stop using the CMM to accept parts. Clean, requalify and repeat the check once, and call for reverification if it fails again.
- Record the result. If the CMM was out, assess the parts measured since the last good check.
How Axiospec helps with CMMs
Axiospec is calibration management software. It does not test a CMM or decide whether it passes. The limits, the decision rule and the interval stay with you or your service provider. The CMM can be an instrument on the register with its verification history, certificates and next due date, and the interval can be set in days, months or years.
Step gauges, ball bars, test spheres and setting rings can sit on the register as reference standards with their own certificates and due dates. With a standard such as ISO 9001 or ISO/IEC 17025 selected for the workspace, you pick the artifact, or a set tracked as one instrument, from the register as the reference standard when you log a verification, and the record keeps the room temperature and humidity. The artifact's Reverse Traceability card then lists the instruments it was used on. One record holds up to 100 named test points, each with a nominal value, a tolerance and the as-found and as-left readings, so each test length or the probing error can be its own point.
An accepted Fail or Damaged result puts the CMM in quarantine. With a standard selected, any approved Fail, Damaged or Pass with an as-found reading out of tolerance needs an impact written, and any not yet written show in Needs attention. A manager can set a CMM that is away for repair to Maintenance, a separate status that pauses its reminders and does not quarantine it. The free plan covers up to 50 active instruments. A new workspace starts with a 45-day trial of Professional, with room for 1,500 active instruments and no credit card. We load your existing records for you at no charge on every plan.
Common questions
- What is a CMM used for?
- A coordinate measuring machine measures the size, position and form of features on a part, such as holes, faces and profiles. It probes points on the surface, fits geometry to them and compares the result with the drawing or the CAD model.
- How is a CMM calibrated?
- Strictly, a CMM is verified rather than calibrated. The ISO 10360 tests measure calibrated test lengths and a test sphere to show the machine still meets its maximum permissible errors. Calibration in NPL's sense means measuring all 21 kinematic error sources for the error map, which is done at installation and as needed.
- What does E0,MPE mean on a CMM specification?
- It is the maximum permissible length measurement error with the stylus tip close to the ram axis, usually written ±(A + L/K) µm with L in millimetres. For example, ±(4 + L/200) µm allows ±4.5 µm on a 100 mm length.
- How often should a CMM be verified?
- No standard fixes the interval. NPL notes that verification is often annual, with interim checks on an artifact between visits and after any collision or large temperature change. Adjust the interval from the machine's as-found history.
- What is a CMM interim check?
- A short check between reverifications on a calibrated artifact, such as a ball bar, setting ring, ball plate or check part, using your normal probing strategy. ISO 10360-2:2009 Annex A strongly recommends regular interim checks.
- Why does temperature matter so much on a CMM?
- The part and the machine both change size with temperature, and dimensions are defined at 20 °C. Steel grows by about 11.5 µm per metre per degree, so in NPL's example a 300 mm steel bar at 23.4 °C measures 12 µm longer than at 20 °C.
Put it into practice
Got a gage list? Send it over and we load it for you, usually in a couple of business days. Free on every plan. Then log every calibration to a tamper-evident audit trail and produce records on demand.
Axiospec is a documentation and workflow tool. It helps you keep clean, traceable, audit-ready records. Certification depends on your own processes, scope and assessor.
Take the audit-prep checklist with you
First the checklist, with each item phrased the way an auditor asks it. Then four short emails with practical tips. You can unsubscribe in one click.
First we email you a link to confirm your address. Nothing else comes until you confirm, and you can unsubscribe from any email after that. See our Privacy Policy.