In-house vs outsourced calibration: how to decide what to calibrate yourself

11 min read Published Updated

In-house calibration means your own people calibrate your gages against reference standards you own and keep calibrated. ISO 9001 allows it when those standards are traceable to national or international measurement standards, and many small and mid-size shops use a mix: simple hand gages calibrated in-house, and reference standards and complex instruments sent to an outside lab.

On this page

What in-house and outsourced calibration mean

In-house calibration is done by your own staff, with your own reference standards, under your own procedures. A technician checks a caliper against a set of gauge blocks, or a thermocouple against a reference thermometer in a dry block, and records the results. You own the method, the evidence and the decision.

Outsourced calibration is done by a calibration provider, either at its lab or on-site at your premises. The provider may be accredited to ISO/IEC 17025 for that calibration, or not. On-site work by an outside lab is still outsourced, because the method, the reference standards and the certificate are the provider's.

Many in-house checks are verifications rather than full calibrations. The gage is compared with a reference and judged against a tolerance. That is legitimate, as long as the procedure says which one you do.

What the standards allow

ISO 9001 does not require an outside laboratory, and it does not require accreditation. Clause 7.1.5.2 applies where measurement traceability is a requirement or is essential to confidence in the results. Measuring equipment is then calibrated or verified at specified intervals, or before use, against measurement standards traceable to international or national measurement standards. Where no such standard exists, you retain the basis you used. Clause 7.1.5.1 asks you to keep evidence that the equipment is fit for its purpose, and clause 7.2 asks you to ensure the competence of the people doing the work.

ISO/IEC 17025:2017 is the competence standard for laboratories. It applies to any organization performing laboratory activities, whatever its size, so an internal lab can work to it or be accredited to it. ISO 9001 does not ask for either. If your lab does work to it, clause 7.6.2 requires an uncertainty evaluation for all calibrations, including calibrations of its own equipment.

IATF 16949 is more specific. Clause 7.1.5.3.1 says an internal laboratory needs a defined scope, included in the quality management system documentation, that covers the inspections, tests or calibrations it performs. The lab has to specify and implement at least these requirements.

  • Adequate technical procedures.
  • Competent laboratory personnel.
  • Testing of the product.
  • The capability to perform the services correctly, traceable to the relevant process standard. Where no national or international standard exists, you define and implement a method to verify the capability of the measurement system.
  • Customer requirements, if any.
  • Review of the related records.
Tip: A note to IATF 16949 clause 7.1.5.3.1 says ISO/IEC 17025 accreditation may be used to demonstrate that an internal lab conforms. It is one way to show conformity, not a requirement.

External labs, aerospace and customer requirements

For outside labs, IATF 16949 clause 7.1.5.3.2 asks for a defined laboratory scope that covers the service you buy. The lab must also be accredited to ISO/IEC 17025 or a national equivalent with that service in its scope, with the certificate carrying the mark of a national accreditation body. Sanctioned interpretation 10 allows one alternative. You may use a lab without that accreditation, such as the equipment manufacturer, if you hold evidence that it was evaluated and meets the internal laboratory requirements of clause 7.1.5.3.1.

AS9100 keeps the ISO 9001 rules and adds, among other things, a register of the equipment that needs calibration or verification. The register records the method, frequency and acceptance criteria for each item, and you need a process to recall equipment for calibration. AS9100 does not itself require an accredited lab. Aerospace customers can require one in their purchase order terms, so read your flowdowns.

Customer requirements come on top of all of these. A purchase order or quality agreement that asks for accredited calibration overrides the freedom the standard gives you.

What in-house calibration needs

Before you take a gage type in-house, make sure you can show each of these to an auditor.

  • Reference standards with their own traceable calibration. Gauge blocks, master rings, reference thermometers and test weights go out to a competent lab, usually an accredited one, on their own interval. Their uncertainty has to be small enough for the tolerances you will check with them.
  • A documented procedure for each instrument type. It names the reference, the test points, the method, the environment and the acceptance criteria.
  • A suitable environment. Dimensional measurements are referenced to 20 °C (ISO 1). A temperature difference between the gage and the reference adds uncertainty, so let them settle together before you measure.
  • Competent people, with training records and a way to show they can do the procedure, such as a supervised first calibration or a comparison with an outside result.
  • An uncertainty estimate or a test uncertainty ratio for each procedure, so you know the method is good enough for the tolerance.
  • A record of each calibration: the instrument, the reference standard used, the as-found and as-left readings, the result, the person, the date and the next due date.
  • Status control and a process for gages found out of tolerance, so you can assess what was measured with them.

Which instruments to keep in-house and which to send out

The split usually follows two questions. Can you buy and maintain a reference standard that is clearly better than the instrument? And is the method simple enough to write down and repeat? The table shows how that often works out. Your tolerances and your customers can move any row.

Typical choices by instrument type
Instrument typeUsual choiceWhat decides it
Calipers, micrometers, dial and test indicators, height gagesIn-houseGauge blocks and a written procedure cover most of the range. The gauge blocks themselves go out.
Plain plug, pin and snap gagesEitherIn-house needs a measuring instrument with a good ratio to the gage tolerance. Gage tolerances are small, so this is harder than it looks.
Thread gages and master setting ringsUsually outPitch diameters and ring diameters need specialized measuring machines and methods.
Reference standards such as gauge blocks, reference thermometers and test weightsOut, to an accredited labThey carry the traceability for everything you calibrate in-house.
Torque wrenches and screwdriversEitherIn-house needs a torque tester whose own calibration and uncertainty suit your tolerances. ISO 6789-2 sets the calibration requirements for hand torque tools.
Pressure gaugesEitherIn-house needs a calibrated reference gauge or deadweight tester and a pressure source for the range.
Load cells and force transducersUsually outThey need a force standard machine or a calibrated reference transducer in a suitable frame. A cell used as your own reference goes to an accredited lab.
Flow metersUsually out, or on site with a prover or master meterA weighing tank or prover that is clearly better than the meter is a large investment. On site you need suitable connections and a stable flow.
Multimeters, clamp meters and process calibratorsUsually outThey need a multifunction calibrator, specialized procedures and an electrical uncertainty evaluation.
Temperature sensorsEitherA dry block and a calibrated reference probe cover many sensors. Very low uncertainties and high temperatures usually go out.
CMMs, vision systems, hardness testers and balancesOutside service, often on-site, with in-house checks between visitsThey need calibrated artifacts and methods that the maker or a specialist service brings. Between visits, run interim checks such as a check standard on the CMM or test weights on a balance.

The hybrid model

A common arrangement has three tiers. Reference standards go to an accredited lab on a fixed interval. Working gages are calibrated in-house against those standards. Complex instruments are calibrated by an outside provider, at its lab or on-site, with in-house checks between visits.

The tiers are connected. Each in-house calibration rests on a reference standard whose certificate came from outside, so a lapsed reference standard affects every gage calibrated against it. Track the due dates of your reference standards as closely as those of your gages.

Revisit the split when something changes. A customer may ask for accredited certificates, a gage type may grow in number, a tolerance may tighten or the person who ran a procedure may leave.

Cost factors to weigh

No price list settles this, because the answer depends on your volume, your tolerances and your people. These are the costs to put side by side.

  • Reference standards for in-house work, and their own outside calibration on their interval.
  • Technician time, training and competence records for the people doing in-house work.
  • A written procedure and an uncertainty evaluation for each in-house method.
  • A temperature-stable space, and the audit time your own lab will draw.
  • For outsourced work, the provider's fee per instrument, shipping, and the admin of purchase orders, supplier approval and certificate review.
  • Downtime while gages are away, which may mean buying spares, and the chance of damage in transit.
Tip: In-house tends to pay where you have many gages of the same simple type and the reference standards are easy to maintain. Outsourcing tends to pay for low-volume, specialized or high-accuracy work.

Risks on each side

The main in-house risk is an error nobody outside sees. A reference standard used past its due date, a procedure that misses a test point or an untrained technician can pass bad gages without anyone noticing. Sending the occasional gage out after an in-house calibration and comparing the two results is a practical check. Expect the auditor to read your procedures, competence records and reference standard certificates, because there is no provider's certificate to rely on.

The main outsourced risk is buying the wrong service. A lab can be accredited, but not for your parameter or range. A certificate can arrive without as-found data or without an uncertainty. Gages are also away for the turnaround time, and an adjustment made without your approval can erase the as-found evidence you need if the gage was out of tolerance.

Records you need either way

Who does the work changes the records you hold, not whether you hold them.

  • An equipment list or register with each instrument's identification, location, interval, status and next due date.
  • For in-house work, the procedure used, the reference standard and its certificate, the as-found and as-left readings, the result, and who did the work and when.
  • For outsourced work, the provider's calibration certificate, your review of it and the supplier approval behind the order.
  • For any instrument found out of tolerance, your assessment of the measurements and product it affected, and the action taken. ISO 9001 clause 7.1.5.2 asks you to determine whether earlier results were adversely affected.
  • Under IATF 16949, the internal laboratory scope and, for outside labs, evidence of their accreditation scope or evidence that they were evaluated against the internal laboratory requirements.

How to decide, step by step

  1. List your instruments by type, with the number of each and the tightest tolerance each type checks.
  2. Check what your standard, your customers and your contracts require, including any demand for accredited certificates.
  3. For each type, decide whether you can hold a reference standard with a good ratio to that tolerance and keep it calibrated.
  4. Write the procedure and estimate the uncertainty or TUR of the in-house method before the first calibration.
  5. Train the people and keep evidence of their competence.
  6. Compare the in-house costs with the outsourced costs for that type, including downtime.
  7. Send out what fails these checks, and approve those providers as external suppliers.
  8. Write the split down, as an internal laboratory scope if you work to IATF 16949, and review it when volumes, tolerances or customers change.

Where Axiospec fits

Axiospec is calibration management software, and it records both kinds of work. An in-house calibration records the reference standard used, the as-found and as-left readings and the certificate. An instrument sent out carries the vendor and an out-for-calibration status until it comes back, and due dates cover both. It does not make an internal lab conform or a provider qualified. Those decisions stay with you.

Common questions

Can I calibrate my own gages under ISO 9001?
Yes. ISO 9001 does not require an outside or accredited lab. It requires calibration or verification at specified intervals, or before use, against measurement standards traceable to international or national measurement standards, with the status identified and evidence kept. In-house work meets that when your reference standards have their own traceable calibration and competent people follow a written procedure.
Does an in-house calibration lab need ISO/IEC 17025 accreditation?
Not under ISO 9001. IATF 16949 does not require it for an internal lab either. Clause 7.1.5.3.1 asks for a defined scope and set controls, and a note says accreditation may be used to demonstrate conformity. Customers can still ask for accredited certificates in their contracts, and your reference standards should be calibrated by a competent lab, usually an accredited one.
What equipment do I need to calibrate gages in-house?
Reference standards for each quantity, such as gauge blocks for calipers and micrometers, with current calibration certificates. They need to be clearly better than the gages they check. You also need a stable environment, which for dimensional gages means close to 20 °C, a written procedure for each instrument type and a way to record the results.
Is in-house calibration cheaper than outsourcing?
It depends on volume and complexity. In-house work carries fixed costs: reference standards and their own calibration, a suitable room, procedures, training and technician time. Outsourcing costs a fee per instrument plus shipping, admin and downtime. In-house tends to pay for many simple gages of one type, and outsourcing for specialized or high-accuracy instruments.
What does IATF 16949 require for an internal laboratory?
Clause 7.1.5.3.1 requires a defined laboratory scope in your quality management system documentation, covering the inspections, tests or calibrations the lab performs. It sets minimum requirements for technical procedures, personnel competence, product testing, traceable capability, customer requirements and the review of records. ISO/IEC 17025 accreditation may be used to demonstrate conformity, but it is not required.
If I calibrate in-house, do I still need an outside lab?
Yes, for your reference standards. Their calibration is what links your in-house results to national or international measurement standards. A common setup sends the reference standards to an accredited lab on a fixed interval and calibrates the working gages against them.