Clamp meter: how it works, how to use it and how to calibrate it

13 min read Published

A clamp meter measures the current in a wire without breaking the circuit: its jaws close around one conductor and sense the magnetic field that the current creates. It is calibrated by passing a known current through the jaws, usually from a calibrator through a multi-turn coil, and comparing its readings with that reference at points across each range.

On this page

What is a clamp meter and how does it measure current?

A clamp meter is a current meter with hinged jaws. You open the jaws, close them around a conductor and read the current, so the circuit never has to be opened. Most clamp meters also measure voltage and resistance through test leads, much like a digital multimeter.

The jaws act as a sensor for the magnetic field around the conductor. CEM, the Spanish national metrology institute, describes two basic designs in its clamp meter calibration procedure EL-007 (clause 4.1). Colombia's national metrology institute, INM, adds the flexible probe as a third type in its 2022 calibration guide.

The three kinds of clamp meter (CEM EL-007 clause 4.1, INM guide clause 5)
TypeHow it senses currentWhat it measures
Current transformerThe conductor acts as a primary winding of one turn. A coil wound on the iron core in the jaws is the secondary, so it carries a small current in proportion to the measured one.AC only
Hall effectThe iron core guides the magnetic field to a Hall sensor set in a gap in the core. The sensor gives a voltage in proportion to the field.AC and DC. CEM says the upper frequency depends on the design and usually does not exceed 1 kHz, or 10 kHz at most.
Flexible probe (Rogowski coil)A coil with no iron core is wrapped around the conductor. It responds to the rate of change of the magnetic field.AC only
Tip: A Hall effect clamp has to be zeroed before a DC reading. Fluke explains that the earth's magnetic field and other fields near the measuring point would otherwise leave an offset in the reading.

How do you measure current with a clamp meter correctly?

The points below come from makers' instructions, mainly Fluke's guide to measuring current with a clamp meter, and from the checks CEM and INM make before a calibration. Your own meter's manual comes first.

  • Clamp one conductor only. The meter reads the sum of the currents inside its jaws, and currents flowing in opposite directions cancel. Around a cable that holds both the live and the neutral conductor, the reading is close to zero.
  • Take the test leads out of the meter before measuring through the jaws, and keep your fingers behind the tactile barrier (Fluke).
  • Close the jaws fully and keep the faces where they meet clean. Before a calibration, CEM checks that the jaws close properly and that those faces are in good condition, and INM cleans oxide and dirt off them.
  • Center the conductor using the alignment marks on the jaws, with the jaws at right angles to it. That is the position calibrations are made in (CEM EL-007 5.2 r, INM guide 6.4.4).
  • For DC current, zero the meter first, with the jaws closed and no conductor inside them (Fluke).
  • Before you clamp or connect anything, check that the meter's measurement category and voltage rating suit the circuit.
Tip: Leakage current clamps are a separate kind, built to read very small currents. IEC 61557-13:2023 sets their performance requirements for low-voltage distribution systems up to 1,000 V AC and 1,500 V DC, including their behavior in strong external magnetic fields.

True RMS or average responding?

The RMS value of an alternating current is its equivalent DC value, the one with the same heating effect. A true RMS meter works it out for any wave shape within its specified limits. An average-responding meter is only right on a pure sine wave.

Many loads draw distorted current. Fluke names variable-speed drives and computers among them, and says an averaging meter can read up to 40% low or 10% high on distorted waveforms. On those circuits, use a true RMS clamp meter.

Look at the crest factor in the specification as well. It is the ratio of a waveform's peak to its RMS value, and a true RMS specification holds only up to a stated crest factor. The Fluke 376 FC, for example, allows a crest factor of 3 at 500 A and adds 2% to its accuracy above a crest factor of 2.

How accurate is a clamp meter?

Makers state accuracy as ±(percentage of reading + digits), as for a multimeter, with separate figures for each function and frequency band. The fixed digits weigh more near the bottom of a range.

Take the Fluke 376 FC. For AC current through the jaws from 10 Hz to 100 Hz it is specified at ±(2% + 5 digits), on a 999.9 A range that reads to 0.1 A. At 100.0 A the limit is 2.0 A plus 5 digits of 0.1 A, so ±2.5 A. At 10.0 A it is 0.2 A plus 0.5 A, so ±0.7 A, which is 7% of the reading.

Conditions apply too. The same specification adds 0.1 times the stated accuracy for each degree Celsius above 28 °C or below 18 °C. For calibration, CEM EL-007 asks for a calibration uncertainty no larger than a third of the error the clamp is allowed in use (clause 2).

Which safety rating does a clamp meter need?

IEC 61010-2-032 sets the safety requirements for hand-held and hand-manipulated current sensors, clamp meters included. The current edition, the fifth, dates from 2023. It sorts current sensors into types by the protection they give during clamping and states that a clamp multimeter is type A or type B. Read the maker's markings and manual before you put a clamp around a bare live conductor.

Measuring circuits are also rated by measurement category, defined in IEC 61010-2-030 (Annex AA). Hioki's manuals put them in practical terms. CAT II covers circuits connected directly to outlets, with household appliances and portable tools as examples. CAT III covers the distribution part of a building's installation, such as distribution boards and circuit breakers. CAT IV covers the source of the installation, before the main fuse or breaker. Hioki warns never to use an instrument on mains above its rated category.

A rating such as CAT III 1000 V / CAT IV 600 V, the rating of the Fluke 376 FC, gives each category with the voltage the meter is rated for in it.

Tip: Flexible probes need their own care. Fluke's safety sheet for its iFlex flexible current probes asks you to de-energize the circuit, or wear protective equipment that meets local requirements, before you put the probe on or take it off.

How is a clamp meter calibrated?

A calibration passes a known current through the jaws and compares the clamp's reading with it. CEM EL-007 (clause 4.2) describes two ways to supply that current.

Two ways to apply a known current (CEM EL-007 clause 4.2)
MethodHow it worksReference
Direct, single conductorA current equal to the test point flows through one conductor in the jaws and is measured at the same time by a reference system.For DC, a shunt and a digital voltmeter. For AC, a reference current transformer and an ammeter.
Indirect, multi-turn coilA calibrator supplies a small current, at most 10 A or 20 A, through a coil with a known number of turns, typically 10 to 50. The jaws close around the coil, so the clamp should read the calibrator current times the number of turns.The calibrator, with its own calibration
Tip: A coil has no parts that drift. Fluke does not recommend recalibrating its coils but suggests intermediate checks: a visual inspection and a confidence check with a clamp meter. INM's guide likewise has coils verified and characterized rather than calibrated.

Coil or single conductor: which is used?

CEM's procedure uses the direct method because it reaches further. It covers 10 A to 2,000 A, DC and AC at 50 Hz. INM's guide uses the coil method and calls it the most widely used. Fluke gives the arithmetic: a 50-turn coil with 20 A from the calibrator presents 1,000 A to the clamp.

The number of turns is exact, but a clamp does not read a coil quite as it reads one straight conductor. Coil makers state that interaction as an uncertainty. Fluke specifies its 50-turn 5500A/COIL at ±(0.25% of effective output + 0.5 A) with toroidal-wound clamps and ±(0.50% + 0.5 A) with other clamps. A 2020 paper at the IMEKO TC4 symposium notes that Hall effect clamps are more sensitive than current transformer clamps to an uneven magnetic field inside the jaws.

What limits a coil calibration?

  • Inductance. The coil and the clamp together load the calibrator. Their impedance, mostly inductance, rises with frequency, and the voltage needed to drive the current can exceed the calibrator's compliance voltage, the most voltage it can supply while holding the current (INM guide clause 5). Coils are therefore specified for a limited frequency range: Fluke's 5500A/COIL from DC and from 45 Hz to 440 Hz.
  • Heat. High currents warm the coil. Fluke rates the 5500A/COIL at 11 A continuous and, above that, up to 20 A for 2 minutes on and 8 minutes off.
  • Jaw size. That coil needs clamp jaws with an inside diameter of at least 2.54 cm (1 in.).
  • Load on the calibrator. INM warns that some clamps load a calibrator enough to trip its protection or distort its output. If that happens, measure the applied current with a reference shunt and voltmeter as part of the setup.

Which test points should you use?

CEM EL-007 sets the points as a percentage of each range's full scale (clause 5.3.2, Table 1), and INM's guide uses the same table. Take five readings at each point once the reading has settled (CEM 5.3.3).

Test points for AC and DC current (CEM EL-007 Table 1)
RangesPointsValues, % of full scale
All ranges except one intermediate range210% and 90%
One intermediate range, for linearity510%, 30%, 50%, 70% and 90%
Tip: EL-007 covers the current functions only. For the voltage and resistance functions, CEM and INM both point to their multimeter procedures, so test those as you would a digital multimeter.

Where do the test points fall on a real clamp?

The 10% point can move up to 25% or 50%, so that it is not a value the next range down could measure. The 90% point can sit anywhere from 80% to 95%. On the range used for linearity, 10% and 90% are meant exactly. CEM calibrates AC at 50 Hz, the mains frequency in Spain, while INM covers 40 Hz to 400 Hz and suggests 60 Hz, the mains frequency in Colombia. Include the frequency your clamp is used at.

In CEM's worked example for a 3¾-digit clamp, the 400 A DC range is tested at 40.0 A, 120.0 A, 200.0 A, 280.0 A and 360.0 A, and the 2,000 A range at 500 A and 1,800 A. CEM leaves out negative DC currents: most clamps do not show them, and turning the clamp round gives a positive reading.

What affects the reading during a calibration?

  • Position. CEM places the conductor in the center of the jaws and at right angles to them (5.2 r). Fluke aligns the clamp's axis with the coil's axis and keeps the jaw plane square to the coil's conductors. INM records readings in the central position and counts the effect of other positions and angles in the uncertainty (6.4.4 and 6.5.1.3).
  • Stray fields. CEM keeps other conductors and metal objects at least 40 cm from the clamp (5.2 r). INM recommends keeping the return conductors of a single-turn loop at a distance of at least four times the size of the meter (6.4.3.1).
  • Remanence in DC clamps. Before each DC range, CEM demagnetizes the core: after a current near full scale, the clamp comes off the conductor and its jaws are opened and closed sharply three to five times. It then zeroes the clamp with no current flowing (5.2 s and t).
  • Hysteresis and core heating. INM adds uncertainty terms for the hysteresis of Hall effect clamps on DC and for the jaws warming up (6.5.1.2 and 6.5.1.4).
  • Conditions. CEM works at 18 °C to 28 °C and no more than 70 %rh, and warms mains-powered instruments up for at least 30 minutes, or as long as their maker says (5.2 i to k).
Tip: How much position matters depends on the clamp. In the IMEKO paper's example, a toroidal-wound clamp at 200 A and 60 Hz on a 50-turn coil, taking readings at different positions raised the relative standard uncertainty from 1.7 × 10⁻³ to 2.2 × 10⁻³.

As-found readings, adjustment and as-left readings

CEM EL-007 describes three sequences (5.3.1). The normal one is a first calibration, an adjustment if it is needed and a final calibration, with records of both calibrations kept. The first calibration gives the as-found readings, which show how the clamp behaved since its last calibration, and the final one gives the as-left readings. Calibration without adjustment is the variant for errors inside the set limits or for an owner who has not authorized an adjustment.

Most clamp meters cannot be adjusted at all. CEM notes that, as industrial rather than laboratory instruments, the great majority have no means of adjustment, so a clamp outside its tolerance is repaired or replaced. It returns a clamp to its planned use when its final errors are inside the tolerance less the calibration uncertainty. Otherwise the clamp is repaired or moved to a job that allows a wider tolerance (6.2).

If an as-found reading was out of tolerance, ISO 9001 clause 7.1.5.2 expects you to check whether results obtained with the clamp can still be trusted and to act on what you find. Look at what its readings decided since its last good calibration, such as a motor load check or a protective device setting.

How often should a clamp meter be calibrated?

No standard fixes one interval for every clamp meter. CEM EL-007 calls 9 to 12 months a reasonable period and leaves the choice to the user, who sets it from the clamp's proven stability, its use and its tolerance and reviews it against earlier results (6.2). Fluke says clamp meters typically need calibrating once a year, and more often after a drop or before critical measurements.

ILAC-G24:2022 / OIML D 10:2022 describes how to review an interval from your own results. Between calibrations, check the clamp against a known current, such as a calibrator and coil, and check it again after any drop or overload.

What should the certificate and record show?

CEM EL-007 lists the data to record at each point (clause 5.4) and adds the uncertainty (clause 6.1). The coil details in the first item and the last two items below are good practice for your own records rather than CEM requirements.

  • The equipment used, such as the current source with the shunt or current transformer and its certified value, or the calibrator with the coil and its number of turns.
  • The ambient conditions and the person who did the calibration.
  • The range, and the settings of the clamp, the source and the standards, such as the resolution or measuring speed chosen.
  • At each point, the clamp and reference readings for each repetition, their mean and standard deviation, and the error.
  • The tolerance assigned to the clamp at each point, where there is one, and the uncertainty.
  • Whether the clamp was adjusted, with the readings before and after.
  • The date and the next due date.

How to calibrate a clamp meter, step by step

  1. Identify the clamp meter, its functions and ranges, and the tolerance it has to meet, from the maker's specification or from how it is used.
  2. Check that the jaws close properly and are clean and undamaged, check the battery and run the meter's own start-up checks.
  3. Check that the calibrator or reference system has a valid calibration, that the coil has passed its last check, and that the calibration uncertainty is no larger than a third of the clamp's tolerance.
  4. Let mains-powered equipment warm up, and record the temperature and humidity.
  5. Center the coil or conductor in the jaws at right angles, away from other conductors. For DC, demagnetize the core and zero the clamp with no current flowing.
  6. At each test point, let the reading settle, then record five readings from the clamp and the reference before any adjustment.
  7. Work out the error at each point and compare it with the tolerance. If the clamp can be adjusted and needs it, adjust it by the maker's method and repeat the points.
  8. Test the voltage and resistance functions as for a digital multimeter.
  9. If an as-found reading was out of tolerance, assess what the clamp measured since its last good calibration. Then record the results and set the next due date.

Where Axiospec fits

Axiospec is calibration management software. It keeps each clamp meter with its calibration history, certificates and next due date. One calibration record holds up to 100 test points, each with its nominal value, its own tolerance, a name such as the range or frequency, and the as-found and as-left readings. A tolerance can be an absolute value or a percentage of the nominal or of the span, so a limit in percent of reading plus digits goes in as the absolute value worked out for that point. With a standard such as ISO 9001 or ISO/IEC 17025 selected for the workspace, each calibration also names the reference standard used, as free text or as a link to a master instrument such as your calibrator.

On an in-house calibration, a final reading outside its tolerance sets the result to Fail, and recording a pass anyway takes a written reason. An accepted Fail or Damaged puts the instrument in Quarantined. With a standard selected, any approved Fail, Damaged or Pass with as-found out of tolerance needs an impact written, and unrecorded ones show in Needs attention. Axiospec does not calibrate the clamp or choose its points, tolerance or interval. Those stay with you or your lab.

The free plan covers up to 50 active instruments. A new workspace starts with a 45-day trial of Professional, with room for 1,500 active instruments and no credit card. We load your existing records for you at no charge on every plan.

Common questions

How does a clamp meter measure current?
Its jaws close around a conductor and sense the magnetic field that the current creates, so the circuit never has to be opened. A current transformer clamp measures AC through a winding on the jaws, and a Hall effect clamp uses a sensor in a gap in the core to measure both AC and DC.
Why does my clamp meter read zero around a whole cable?
A clamp meter reads the sum of the currents inside its jaws. In a cable, the current in the live and the neutral conductor flows in opposite directions and cancels, so clamp one conductor at a time.
Can a clamp meter measure DC current?
Only an AC/DC clamp meter can, and these usually work on the Hall effect. Current transformer clamps and flexible Rogowski probes measure AC only. Zero a DC clamp with its jaws closed and empty before each DC measurement.
How do you calibrate a clamp meter?
Pass a known current through the jaws, usually from a calibrator through a coil of 10 to 50 turns, or from a high-current source through one conductor measured with a reference shunt or current transformer. CEM EL-007 compares the readings at 10% and 90% of each range and at five points on one range. Calibrate the voltage and resistance functions as for a multimeter.
How often should a clamp meter be calibrated?
Commonly once a year. CEM EL-007 calls 9 to 12 months reasonable and Fluke says clamp meters typically need calibrating yearly. Set the interval from the clamp's own as-found results, its use and its tolerance, and check it after any drop.
What is the difference between a clamp meter and a multimeter?
A multimeter measures current by being connected into the circuit, which has to be opened. A clamp meter measures current through its jaws without breaking the circuit, and most also measure voltage and resistance through test leads like a multimeter.