Anemometer: types, how to use one and how to calibrate it
An anemometer measures air speed, from a vane or cups that the air turns, a heated sensor that it cools, or a pressure or sound travel time that it changes. To calibrate one, compare its reading at several steady speeds in a wind tunnel with a reference such as a laser Doppler anemometer or a calibrated Pitot-static tube.
On this page
What is an anemometer, and which type do you have?
An anemometer is an instrument that measures the speed of moving air. The name covers several principles, and each behaves differently near the limits of its range. NIST's air speed calibration service, for example, takes Pitot tubes and hot-wire, thermal, ultrasonic, cup and propeller anemometers.
| Type | How it measures | Typical use | Watch for |
|---|---|---|---|
| Vane (rotating vane) | A small propeller turns at a rate that follows the air speed along its axis | Grilles, diffusers, ducts and site work | A minimum speed before it turns, and air arriving at an angle |
| Cup | Cups on arms around a vertical shaft spin at a rate that follows the wind speed | Weather stations and wind energy masts | Readings that depart from a straight line at low speed, and overspeeding in gusty wind |
| Thermal (hot-wire or hot-film) | The air cools a small heated sensor, and the current or voltage needed to keep it at a constant temperature gives the speed | Ducts, outlets and low air speeds | It reads standard velocity, and the sensor is fragile and sensitive to dirt and moisture |
| Pitot-static tube | Measures the difference between total and static pressure. The speed follows from that pressure and the air density | Duct traverses and reference measurements in wind tunnels | A very small pressure at low speed, so a large uncertainty below about 1 m/s |
| Ultrasonic (sonic) | Times ultrasonic pulses over fixed paths | Weather and wind energy | No moving parts to wear |
How do you use an anemometer correctly?
- Follow the maker's instructions for the head or probe. WMO-No. 8 gives that as the first rule for hand-held anemometers (5.3.1).
- Face it into the flow. A propeller responds to the air speed along its axis (WMO-No. 8, 5.4), and a thermal probe often has a mark that has to face upstream. TSI's VelociCalc manual, for example, asks for the sensor window to be fully exposed with the orientation dimple facing upstream.
- Keep yourself and obstructions out of the air you measure. Outdoors, WMO-No. 8 asks for a point well exposed to the wind, or at least 10 times the height of an obstruction away from it (5.3.1).
- Average over time. Air moves in gusts and eddies, so read an average rather than one value. Synoptic weather reports use 10-minute averages (WMO-No. 8, 5.1.2), and the Guide calls a 1-minute average a long gust (5.1.1).
- Cover the area. Flow across a duct or a grille is not uniform, so take readings at several points across the section. ISO 3966:2025 sets out this velocity-area method for ducts, in which the flow is the velocity integrated over the section, using Pitot-static tubes.
Standard or actual velocity: why air density matters
A thermal sensor is sensitive to air density as well as to speed, so thermal anemometers show velocity referred to a set of standard conditions. TSI defines its standard conditions as 21.1 °C and 101.4 kPa at 0 % relative humidity and notes that other makers may use different values (application note TSI-109). That reading is called standard or mass velocity. Actual velocity is how fast the air really moves.
The two differ whenever the air is not at standard conditions. In TSI's table, a reading of 5 m/s standard is 5.35 m/s actual at 26 °C, 70 %rh and 740 mmHg, a difference of 7.0 %.
A Pitot-static tube works the other way round. It measures the dynamic pressure, and the speed follows from Bernoulli's equation, V = C √(2ΔP/ρ), so the air density ρ has to be worked out from the temperature, pressure and humidity (NIST SP 250-79, 6.1). Rotating anemometers depend much less on density. WMO-No. 8 notes that in a well-designed cup or propeller anemometer the linearity of the calibration does not depend on air density (5.4).
What else affects an anemometer's reading?
- Low speed. A rotating anemometer needs a minimum speed, its starting threshold, before it turns. Near that threshold a cup anemometer's calibration can depart substantially from a straight line, depending on its design (WMO-No. 8, 5.4). A Pitot-static tube gives a very small pressure at low speed, and NIST notes a large speed uncertainty below about 1 m/s in air (SP 250-79, 6.1).
- Gusts. Cup and propeller anemometers speed up faster than they slow down, so in gusty wind they read high on average. WMO-No. 8 reports total overspeeding of as much as 10 % for some cup designs in turbulent wind over very rough terrain (5.4).
- Flow angle. Air arriving at an angle changes the reading of most designs. ISO 17713-1:2007 sets wind tunnel methods for this off-axis response of cup and propeller anemometers, along with their starting threshold, distance constant and transfer function.
- Wear and dirt. Bearing friction from dust, corrosion and wear in the signal generator change a rotating anemometer over time (WMO-No. 8, 5.10). Hot-wire sensors are fragile, and their calibration changes fast in dirty or wet surroundings (5.6).
- Mounting. In wind energy calibrations the cup anemometer sits on a tube of the same diameter as in service, within 1 mm (MEASNET Anemometer Calibration Procedure, F.3).
How is an anemometer calibrated?
In a wind tunnel. The anemometer sits in a steady, uniform air stream, and its reading is compared with a reference at several speeds across its range. WMO-No. 8 says a fully reliable calibration of cup, propeller and vane anemometers is possible only in a wind tunnel (5.10).
NIST's service shows how a national metrology institute does it. Its primary standard is a fiber-optic laser Doppler anemometer (LDA), calibrated with known velocities from a rotating disk, in a tunnel with a test section of 1.5 m by 2.1 m. In NIST's 2007 description of the service (SP 250-79), a Pitot-static tube and a thermal anemometer served as check standards. Today the service covers 0.15 m/s to 75 m/s (NIST service page). A normal calibration runs two cycles separated by a shutdown, with five measurements at each speed in each cycle, so each result is the average of ten measurements, at seven to thirteen speeds. The air temperature, relative humidity and pressure in the tunnel are measured and reported.
NIST states its uncertainty as U(v) = 0.44 % + 0.16/v² %, with v in m/s and k = 2. That is about 0.6 % at 1 m/s and about 4.4 % at 0.2 m/s, so the lowest speeds are the hardest to calibrate.
PTB, Germany's national metrology institute, works with calibrated LDAs too. Its researchers, with Deutsche WindGuard, report that a calibrated LDA used as the reference in a wind tunnel can measure wind speeds from 4 m/s to 16 m/s with an uncertainty below 0.2 % (Müller and others, 2010).
| Reference | How it works | Where it is used |
|---|---|---|
| Laser Doppler anemometer (LDA) | Measures the speed of fine particles carried by the air through a pattern of light fringes, with nothing in the flow | Primary standard at NIST, and in use at PTB |
| Pitot-static tube with a calibrated differential pressure gauge | Speed from the dynamic pressure and the air density | Calibration wind tunnels, including wind energy laboratories, and duct traverses |
| Calibrated transfer anemometer | A calibrated anemometer read beside the one under test | A working standard in a laboratory, and field comparisons |
Cup anemometers for wind energy: the MEASNET procedure
Most anemometers in wind energy work are cup anemometers. PTB and Deutsche WindGuard described them in 2010 as calibrated by accredited laboratories in wind tunnels, to a procedure first agreed within MEASNET, a network of wind energy institutes. That procedure became Annex F of IEC 61400-12-1 in its 2005 and 2017 editions, and MEASNET's Version 3 of December 2020 applies Annex F of the 2017 edition. In 2022, IEC moved the wind measurement parts of IEC 61400-12-1 into separate standards, and IEC 61400-50-1:2022 now covers wind measuring instruments on meteorological masts, nacelles and spinners.
The points below come from MEASNET Version 3.
- The calibration range is 4 m/s to 16 m/s. It may be extended, but the linear regression on the certificate is calculated for 4 m/s to 16 m/s only.
- The anemometer runs in for at least 5 minutes at about 10 m/s before the calibration starts.
- The blockage ratio of the anemometer in the tunnel may not exceed 0.05, and flow stability is tested at 4, 8, 12 and 16 m/s.
- The reference Pitot-static tube may be misaligned by no more than 0.5°. An LDA is accepted if its uncertainty is no larger.
- A quality control anemometer is calibrated on each day of calibrations to confirm the flow quality.
Why small errors matter for wind data
The result of a cup anemometer calibration is a straight line, a slope and an offset that turn the anemometer's output into wind speed. In 2010, PTB and Deutsche WindGuard put the smallest uncertainties then stated for these calibrations at about 0.1 m/s, which works out to 2.5 % at 4 m/s and about 0.6 % at 16 m/s.
The same paper explains why this matters. The energy and power of a wind turbine go with the cube of the wind speed. In its example, a small turbine gives 300 kW at 9 m/s, and its power curve rises there by 115 kW for each m/s. An uncertainty of 1 % in wind speed then means about 3.5 % in power, which the paper gives as about 4 % of the expected yield.
HVAC airflow: vane and thermal meters
In buildings, anemometers are used to test, adjust and balance air systems. ASHRAE Standard 111-2024 gives uniform procedures for measuring, testing, adjusting, balancing, evaluating and reporting the performance of building heating, ventilating and air-conditioning systems.
Read the accuracy statement with the speed in mind. A thermal meter specified at ±3 % of reading or ±0.015 m/s, whichever is greater, from 0.15 m/s to 30 m/s (TSI VelociCalc 9545) has a limit of ±0.15 m/s at 5 m/s. At 0.3 m/s the fixed ±0.015 m/s takes over, which is 5 % of the reading.
Choose calibration speeds where the meter is used. A meter that reads grilles at 1 m/s to 3 m/s needs calibration points there, not only at higher speeds.
Field checks or laboratory calibration?
Between calibrations, simple checks catch damage and drift. WMO-No. 8 advises regular inspections because anemometers deteriorate in the field (5.10).
- Inspect for physical damage: bent vanes or cups, a cracked thermal sensor or blocked Pitot ports.
- Check the zero. WMO-No. 8 checks a rotating anemometer's zero by holding the cups or propeller still, and asks for zero and range checks on the electronics (5.10).
- Turn the rotor by hand and feel for rough or stiff bearings, a cause of deterioration that WMO-No. 8 lists.
- Compare it with a calibrated anemometer in the same steady flow. WMO-No. 8 suggests this field comparison when wind tunnel tests are not possible (5.4).
| Question | Field check | Laboratory calibration |
|---|---|---|
| Reference | A calibrated anemometer in the same flow, or a zero check | A wind tunnel with an LDA or a calibrated Pitot-static tube |
| Speeds | One or two near the working speed | Five or more across the range |
| Result | Still within your check limit, or not | The error at each speed, or a fitted line, with its uncertainty, on a certificate |
| Due date | Unchanged | Set again from the calibration date |
How often should an anemometer be calibrated?
No standard fixes one interval for every anemometer. Makers such as TSI recommend annual recalibration of their hand-held meters. ILAC-G24:2022 / OIML D 10:2022 lists the factors for choosing an interval, among them the tendency to wear and drift, the extent and severity of use and the environment (5.1). Its clause 6 describes how to review the interval from your own results.
Calibrate sooner after a drop, a bent vane, a damaged sensor or use in dust, rain or corrosive air. Lengthen the interval only on a record of as-found results inside the limit.
What should the certificate and record show?
ISO/IEC 17025:2017 clause 7.8 sets the general content of a calibration certificate. NIST's reports add the air conditions in the tunnel, and the MEASNET procedure asks for the mounting and orientation of the anemometer to be documented. The list below combines these with good practice for your own records.
- The anemometer's ID, the serial number of its head or probe and its range.
- The reference used, with its ID, certificate and due date.
- The air temperature, pressure and relative humidity during the calibration.
- Whether the readings are standard or actual velocity, and the standard conditions used.
- For a directional instrument, the mounting and the orientation.
- At each speed, the reference speed, the reading and the error, or for a cup anemometer the slope and offset of the fitted line.
- The as-found and as-left results, the uncertainty, the date, who did the work and the next due date.
What if the as-found reading is out of tolerance?
Record the readings before anyone adjusts the anemometer. If an as-found error is outside its limit, ISO 9001 clause 7.1.5.2 asks you to determine whether earlier results were adversely affected and to act as needed. For an anemometer, look at what its readings decided since its last good calibration, such as air balance reports, extraction and ventilation checks, wind limits for cranes and lifting, or wind data for a turbine site.
How to calibrate an anemometer, step by step
- Identify the anemometer, its head or probe, the range it is used over and the limit that applies, from the maker's specification or your own requirement.
- Inspect it. Check that the vane or cups turn freely, the blades are straight, the thermal sensor is clean and whole and the Pitot ports are clear.
- Choose at least five speeds across the range you use, including the low end, where the errors are largest.
- Check that the reference has a current certificate and an uncertainty that suits the limit.
- Mount the anemometer in the tunnel's test section in the orientation used in service, and record the air temperature, pressure and humidity.
- At each speed, let the flow settle, then record the reference speed and the anemometer reading over the same averaging time. Repeat the readings.
- Work out the error at each speed, or fit the line for a cup anemometer, and compare the results with the limit. Record these as-found results before any adjustment.
- If the anemometer needs adjusting, adjust it as the maker describes and repeat the speeds as left.
- If an as-found result was out of tolerance, assess what the anemometer measured since its last good calibration. Then set the next due date.
Where Axiospec fits
Axiospec is calibration management software. It keeps each anemometer, whether vane, thermal or cup, with its calibration history, certificates and next due date, and the interval can be set in days, months or years. One calibration record holds up to 100 test points. Each point has its nominal value, its tolerance as an absolute band, a percentage of nominal or a percentage of span, and the as-found and as-left readings, so each calibration speed can be its own point. With a standard such as ISO 9001 or ISO/IEC 17025 selected for the workspace, each calibration also keeps the room temperature and humidity and names the reference standard used, as free text or as a link to a master instrument.
On an in-house calibration, a final reading outside its tolerance sets the result to Fail, and recording a pass anyway takes a written reason. An accepted Fail or Damaged puts the instrument in Quarantined. With a standard selected, any approved Fail, Damaged or Pass with as-found out of tolerance needs an impact written, and unrecorded ones show in Needs attention. Axiospec does not run the calibration or pick the points, the limit or the interval. Those stay with you or your lab.
The free plan covers up to 50 active instruments. A new workspace starts with a 45-day trial of Professional, with room for 1,500 active instruments and no credit card. We load your existing records for you at no charge on every plan.
Common questions
- How is an anemometer calibrated?
- In a wind tunnel. The anemometer and a reference, such as a laser Doppler anemometer or a calibrated Pitot-static tube, read the same steady air stream at several speeds across the range. The error at each speed, or a fitted line for a cup anemometer, goes on the certificate with its uncertainty.
- How often should an anemometer be calibrated?
- No standard sets one interval. Makers such as TSI recommend annual recalibration of hand-held meters. Calibrate sooner after a drop or damage, and adjust the interval from each instrument's as-found history.
- What is the difference between a vane and a hot-wire anemometer?
- A vane anemometer has a small propeller that the air turns, and it needs a minimum speed to start. A hot-wire or thermal anemometer senses how much the air cools a heated sensor. It suits low speeds, but it reads standard velocity and the sensor is fragile.
- Why do two anemometers give different readings in the same air?
- Often because one reads standard velocity and the other actual velocity. In TSI's example, 5 m/s standard is 5.35 m/s actual at 26 °C, 70 %rh and 740 mmHg. Position, flow angle and averaging time also change the reading.
- Can you check an anemometer without a wind tunnel?
- You can check it but not fully calibrate it. Inspect it, check the zero by holding the cups or vane still, and compare it with a calibrated anemometer in the same steady flow. WMO-No. 8 says a fully reliable calibration of a rotating anemometer needs a wind tunnel.
- What uncertainty can an anemometer calibration reach?
- It depends on the speed and the reference. NIST states 0.44 % + 0.16/v² % at k = 2, which is about 0.6 % at 1 m/s and more at lower speeds. PTB researchers report below 0.2 % from 4 m/s to 16 m/s with a calibrated laser Doppler anemometer.
Put it into practice
Got a gage list? Send it over and we load it for you, usually in a couple of business days. Free on every plan. Then log every calibration to a tamper-evident audit trail and produce records on demand.
Axiospec is a documentation and workflow tool. It helps you keep clean, traceable, audit-ready records. Certification depends on your own processes, scope and assessor.
Take the audit-prep checklist with you
First the checklist, with each item phrased the way an auditor asks it. Then four short emails with practical tips. You can unsubscribe in one click.
First we email you a link to confirm your address. Nothing else comes until you confirm, and you can unsubscribe from any email after that. See our Privacy Policy.