Accuracy vs precision in measurement: trueness, bias and resolution
Accuracy is how close a measured value is to the true value, and precision is how closely repeated readings agree with each other, so an instrument can be precise and still inaccurate when every reading carries the same offset. The VIM (JCGM 200:2012, entries 2.13 to 2.15) treats that offset as a question of trueness, measured as bias, and relates accuracy to both trueness and precision.
On this page
What do accuracy, trueness and precision mean?
The VIM, the international vocabulary of metrology that the BIPM publishes free as JCGM 200:2012, defines three related ideas. Accuracy (2.13) compares one measured value with the true value of the quantity measured. Trueness (2.14) compares the average of a very large number of repeated values with a reference value. Precision (2.15) compares repeated readings with each other, under conditions you state.
None of the three is a number in the VIM. A measurement is called more accurate when its error is smaller (2.13, note 1). Trueness cannot be given a numerical value either, and it tracks systematic error only, not random error (2.14, notes 1 and 2). Precision is put into numbers through measures of imprecision such as the standard deviation (2.15, note 1).
The VIM also warns against two swaps. Accuracy should not be used for trueness, and precision should not be used for accuracy (2.13, note 2, and 2.15, note 4). ISO 5725-1:2023, the general part of the ISO series on the accuracy of measurement methods and results, puts the same structure in its title: accuracy, made up of trueness and precision.
| Term | VIM entry | What it compares | The error behind it | How it gets a number |
|---|---|---|---|---|
| Accuracy | 2.13 | One measured value with the true value | All of it, systematic and random | Indirectly, as the error found at a calibration point and its uncertainty |
| Trueness | 2.14 | The average of many repeats with a reference value | Systematic error | As bias, the average minus the reference value (2.18) |
| Precision | 2.15 | Repeated readings with each other | Random error | As a standard deviation, with the conditions stated |
| Resolution | 4.14 | The smallest change in the quantity that shows in the reading | The limit of what the instrument can show, such as its display step, noise or friction | As the display step, entered in a budget as about 0.29 times the step (GUM F.2.2.1) |
The target picture, in words
The usual illustration is a target with a group of shots on it. The center of the target stands for the true or reference value, the center of the group for the average reading and the size of the group for the scatter between repeats. The table describes the four patterns without the picture.
| Pattern of shots | Trueness: where the average sits | Precision: how tight the group is | A single reading |
|---|---|---|---|
| Tight group on the center | On the reference value | High | Accurate |
| Tight group to one side | Off by a bias | High | Off by about the bias, every time |
| Wide group around the center | On the reference value | Low | Often well off, although the average is right |
| Wide group to one side | Off by a bias | Low | Off by the bias plus the scatter |
Why can a precise instrument still be wrong?
Precision only says the readings agree with each other. It says nothing about where they sit against the true value, so a gage can repeat to the last digit and still read high. In the bias example in our gage bias guide, a digital micrometer measures a 12.5 mm gauge block ten times. The readings scatter with a standard deviation of 0.92 µm, but their average sits 2.2 µm above the block's certified size.
More readings do not fix that. Data case: n independent readings of one object under repeatability conditions. Their average has a standard deviation of s / √n (GUM 4.2.3), so averaging shrinks the random part of the error of the mean. The bias is in every reading, so it stays in the average at full size.
Only a comparison with a reference shows a bias. The NIST/SEMATECH e-Handbook of Statistical Methods (section 2.1.1.3) lists calibration by a reference laboratory, check standards on control charts, blind test items and interlaboratory comparisons as ways to find one. It also warns that a bias can remain even when instruments are calibrated, and names temperature as probably the factor with the most potential to cause one. Once a systematic error is known it can be corrected (VIM 2.17, note 2), and the uncertainty of the correction stays with the result.
Precision under which conditions?
A precision figure means little until you say what was held constant. The NIST/SEMATECH e-Handbook (section 2.1.1.4) makes the practical point. The short-term repeatability standard deviation covers one run of minutes or hours. Day-to-day effects such as the environment and handling can add more, so a measurement process is not fully described until that longer-term variability is estimated too.
A repeatability figure on its own can therefore understate the spread you see in routine use, and precision figures can only be compared when their conditions match. The VIM names three sets of conditions (2.15, note 2), and each gives its own kind of precision.
| Precision | VIM entries | What is held the same | What changes |
|---|---|---|---|
| Repeatability | 2.20 and 2.21 | The procedure, operators, measuring system, operating conditions and location, over a short time | Only the repeat itself |
| Intermediate precision | 2.22 and 2.23 | The procedure and location, over an extended time | Possibly calibrations, calibrators, operators and measuring systems |
| Reproducibility | 2.24 and 2.25 | Measurements of the same or similar objects | Locations, operators and measuring systems, and possibly the procedure |
Where does resolution fit?
Resolution is the smallest change in the measured quantity that causes a change you can see in the reading (VIM 4.14). For a display it is the smallest difference between readings that can be told apart (4.15), which on a digital display is the step of the last digit. The VIM notes that resolution can also depend on noise or friction, and on the value being measured.
Resolution is not accuracy. A caliper that displays 0.01 mm reports to the nearest 0.01 mm, but that says nothing about whether the reading is within 0.01 mm of the true size. How close it is comes from calibration against a reference.
Resolution does put a floor under the precision you can see. Data case: repeated readings from a digital display with a step δx. If they are all identical, their standard deviation is zero, but the GUM (F.2.2.1) points out that the uncertainty is not, because any input within ±δx/2 gives the same display. It treats the input as spread evenly over that width, with a standard uncertainty of δx / √12, about 0.29 δx.
How is each one checked and expressed?
Each property has its own test and its own number. Accuracy is the one without a test of its own. In practice it gets its number through measurement uncertainty (VIM 2.26), which includes components from systematic effects, such as the uncertainty of a correction or of a reference standard's value, as well as random ones. The VIM also notes that an estimated systematic effect is sometimes left uncorrected and covered by an uncertainty component instead (2.26, note 1).
The table lists the usual checks.
| Property | How you check it | What you record | Where it is used |
|---|---|---|---|
| Trueness | Measure a reference of known value several times and compare the average with it | The bias, with a t test of whether it differs from zero | Bias and linearity studies, calibration |
| Precision | Repeat readings under stated conditions | A standard deviation and the conditions | Repeatability checks, Gage R&R, Type 1 gage studies |
| Accuracy at a point | Compare a reading with a calibrated reference | The error of indication and the measurement uncertainty of the calibration | Calibration certificates |
| Resolution | Read the display step, or find the smallest change that moves the reading | The step, and its uncertainty term | Uncertainty budgets |
How do you read an accuracy spec?
On a data sheet, accuracy usually means a limit: the largest error the maker allows, which the VIM calls a maximum permissible error (4.26), or an accuracy class that sets such limits (4.25). It is not the error of your instrument. Calibration finds that. A limit given as a percentage only means something once you know what it is a percentage of.
- An absolute limit, such as ±0.02 mm. The same band applies across the range.
- Percent of reading. The limit scales with the value shown, so it is tight at the bottom of the range.
- Percent of full scale. A fixed band equal to a share of the upper range limit, so it is large compared with small readings.
- Percent of span. A share of the whole range between its two limits. The VIM's example (4.5) is a range of −10 V to +10 V with a span of 20 V, so ±1% of span is ±0.2 V where ±1% of the upper limit would be ±0.1 V.
- Percent of reading plus counts or digits. Common on digital multimeters: one part scales with the reading and the other is a fixed number of steps of the last digit.
- An accuracy class, as on pressure gauges, where the class number sets the limit as a share of the range.
| Reading | ±0.5% of full scale | That limit as a share of the reading | ±0.5% of reading |
|---|---|---|---|
| 1 bar | ±0.05 bar | 5% | ±0.005 bar |
| 5 bar | ±0.05 bar | 1% | ±0.025 bar |
| 10 bar | ±0.05 bar | 0.5% | ±0.05 bar |
- Multimeter specs in percent of reading plus counts
- Pressure gauge accuracy classes
- Hygrometer calibration
How this connects to Gage R&R, bias studies and uncertainty
Measurement systems analysis splits along the same line. Bias, linearity and stability studies look at where readings sit against a reference, which is trueness. Gage R&R looks at how readings scatter between repeats and between appraisers, which is precision. An uncertainty budget brings both into one stated uncertainty for a given measurement.
| Study | Trueness or precision | What it measures |
|---|---|---|
| Bias study | Trueness | The bias at one reference value, tested against zero |
| Linearity study | Trueness | How the bias changes across the operating range |
| Stability study | Trueness over time | How the bias on one reference part moves as time passes |
| Gage R&R | Precision | Repeatability (EV) and reproducibility (AV), against the part variation or the tolerance |
| Type 1 gage study | Both | The spread (Cg), and the spread together with the bias (Cgk), on one reference part, against the tolerance |
| Uncertainty budget | Both | Each significant source, combined into a standard and an expanded uncertainty |
Common mix-ups
- Calling the number of digits precision. That is resolution, and a fine display can sit on top of a scattered or biased measurement.
- Reading a data sheet's accuracy as the error of your instrument. It is a limit. The error at each point comes from the calibration.
- Fixing a bias by averaging. Averaging reduces the random part of the error of the mean and leaves the bias where it was.
- Quoting a precision without its conditions. A repeatability standard deviation and a reproducibility standard deviation describe different things.
- Treating a percent of full scale spec and a percent of reading spec as the same. Work both out at the readings you actually use.
Where Axiospec fits
Axiospec is calibration management software. A calibration record keeps the as-found and as-left readings at up to 100 test points, each with its own tolerance as an absolute band, a percentage of nominal or a percentage of span, so the limit sits next to the error found. The bias and linearity, Gage R&R, Type A uncertainty and resolution calculators on this site are standalone tools that need no account.
Common questions
- What is the difference between accuracy and precision?
- Accuracy is how close a measured value is to the true value. Precision is how closely repeated readings agree with each other under stated conditions. A gage with a constant offset can be very precise and still inaccurate.
- What is trueness in measurement?
- Trueness is how close the average of many repeated measurements is to a reference value (VIM 2.14). It reflects systematic error only, and it is put into numbers as bias, the average minus the reference value.
- Can a measurement be precise but not accurate?
- Yes. If every reading carries the same bias, the readings agree closely with each other and all miss the true value by about the same amount. Repeating the measurement does not reveal that, and only a comparison with a reference of known value does.
- Is resolution the same as accuracy?
- No. Resolution is the smallest change an instrument can show, such as the last digit of a display (VIM 4.14 and 4.15). An instrument can display 0.001 mm and still be off by more than that, which only a calibration shows.
- What does ±1% of full scale mean?
- It means the allowed error is 1% of the instrument's upper range limit at every reading. On a 0 to 10 bar gauge that is ±0.1 bar everywhere, which is 10% of a 1 bar reading. A percent of reading spec instead shrinks with the value shown.
- How is precision expressed?
- As a standard deviation, a variance or a coefficient of variation of repeated readings, with the conditions stated (VIM 2.15). Repeatability, intermediate precision and reproducibility are precision under different conditions, so each needs its own figure.
Put it into practice
Got a gage list? Send it over and we load it for you, usually in a couple of business days. Free on every plan. Then log every calibration to a tamper-evident audit trail and produce records on demand.
Axiospec is a documentation and workflow tool. It helps you keep clean, traceable, audit-ready records. Certification depends on your own processes, scope and assessor.
Take the audit-prep checklist with you
First the checklist, with each item phrased the way an auditor asks it. Then four short emails with practical tips. You can unsubscribe in one click.
First we email you a link to confirm your address. Nothing else comes until you confirm, and you can unsubscribe from any email after that. See our Privacy Policy.