Accuracy vs precision in measurement: trueness, bias and resolution

10 min read Published Updated

Accuracy is how close a measured value is to the true value, and precision is how closely repeated readings agree with each other, so an instrument can be precise and still inaccurate when every reading carries the same offset. The VIM (JCGM 200:2012, entries 2.13 to 2.15) treats that offset as a question of trueness, measured as bias, and relates accuracy to both trueness and precision.

On this page

What do accuracy, trueness and precision mean?

The VIM, the international vocabulary of metrology that the BIPM publishes free as JCGM 200:2012, defines three related ideas. Accuracy (2.13) compares one measured value with the true value of the quantity measured. Trueness (2.14) compares the average of a very large number of repeated values with a reference value. Precision (2.15) compares repeated readings with each other, under conditions you state.

None of the three is a number in the VIM. A measurement is called more accurate when its error is smaller (2.13, note 1). Trueness cannot be given a numerical value either, and it tracks systematic error only, not random error (2.14, notes 1 and 2). Precision is put into numbers through measures of imprecision such as the standard deviation (2.15, note 1).

The VIM also warns against two swaps. Accuracy should not be used for trueness, and precision should not be used for accuracy (2.13, note 2, and 2.15, note 4). ISO 5725-1:2023, the general part of the ISO series on the accuracy of measurement methods and results, puts the same structure in its title: accuracy, made up of trueness and precision.

Accuracy, trueness, precision and resolution in VIM terms
TermVIM entryWhat it comparesThe error behind itHow it gets a number
Accuracy2.13One measured value with the true valueAll of it, systematic and randomIndirectly, as the error found at a calibration point and its uncertainty
Trueness2.14The average of many repeats with a reference valueSystematic errorAs bias, the average minus the reference value (2.18)
Precision2.15Repeated readings with each otherRandom errorAs a standard deviation, with the conditions stated
Resolution4.14The smallest change in the quantity that shows in the readingThe limit of what the instrument can show, such as its display step, noise or frictionAs the display step, entered in a budget as about 0.29 times the step (GUM F.2.2.1)

The target picture, in words

The usual illustration is a target with a group of shots on it. The center of the target stands for the true or reference value, the center of the group for the average reading and the size of the group for the scatter between repeats. The table describes the four patterns without the picture.

The four target patterns in VIM terms
Pattern of shotsTrueness: where the average sitsPrecision: how tight the group isA single reading
Tight group on the centerOn the reference valueHighAccurate
Tight group to one sideOff by a biasHighOff by about the bias, every time
Wide group around the centerOn the reference valueLowOften well off, although the average is right
Wide group to one sideOff by a biasLowOff by the bias plus the scatter
Tip: The third row is where shop talk and the VIM part ways. Shop talk calls it accurate but not precise. In VIM terms the average is true, but a single reading is not accurate, because accuracy belongs to each measured value and the scatter is part of its error.

Why can a precise instrument still be wrong?

Precision only says the readings agree with each other. It says nothing about where they sit against the true value, so a gage can repeat to the last digit and still read high. In the bias example in our gage bias guide, a digital micrometer measures a 12.5 mm gauge block ten times. The readings scatter with a standard deviation of 0.92 µm, but their average sits 2.2 µm above the block's certified size.

More readings do not fix that. Data case: n independent readings of one object under repeatability conditions. Their average has a standard deviation of s / √n (GUM 4.2.3), so averaging shrinks the random part of the error of the mean. The bias is in every reading, so it stays in the average at full size.

Only a comparison with a reference shows a bias. The NIST/SEMATECH e-Handbook of Statistical Methods (section 2.1.1.3) lists calibration by a reference laboratory, check standards on control charts, blind test items and interlaboratory comparisons as ways to find one. It also warns that a bias can remain even when instruments are calibrated, and names temperature as probably the factor with the most potential to cause one. Once a systematic error is known it can be corrected (VIM 2.17, note 2), and the uncertainty of the correction stays with the result.

Precision under which conditions?

A precision figure means little until you say what was held constant. The NIST/SEMATECH e-Handbook (section 2.1.1.4) makes the practical point. The short-term repeatability standard deviation covers one run of minutes or hours. Day-to-day effects such as the environment and handling can add more, so a measurement process is not fully described until that longer-term variability is estimated too.

A repeatability figure on its own can therefore understate the spread you see in routine use, and precision figures can only be compared when their conditions match. The VIM names three sets of conditions (2.15, note 2), and each gives its own kind of precision.

The three precision conditions in the VIM
PrecisionVIM entriesWhat is held the sameWhat changes
Repeatability2.20 and 2.21The procedure, operators, measuring system, operating conditions and location, over a short timeOnly the repeat itself
Intermediate precision2.22 and 2.23The procedure and location, over an extended timePossibly calibrations, calibrators, operators and measuring systems
Reproducibility2.24 and 2.25Measurements of the same or similar objectsLocations, operators and measuring systems, and possibly the procedure
Tip: In a Gage R&R study, repeatability is called equipment variation (EV) and the spread between appraisers is reproducibility, called appraiser variation (AV).

Where does resolution fit?

Resolution is the smallest change in the measured quantity that causes a change you can see in the reading (VIM 4.14). For a display it is the smallest difference between readings that can be told apart (4.15), which on a digital display is the step of the last digit. The VIM notes that resolution can also depend on noise or friction, and on the value being measured.

Resolution is not accuracy. A caliper that displays 0.01 mm reports to the nearest 0.01 mm, but that says nothing about whether the reading is within 0.01 mm of the true size. How close it is comes from calibration against a reference.

Resolution does put a floor under the precision you can see. Data case: repeated readings from a digital display with a step δx. If they are all identical, their standard deviation is zero, but the GUM (F.2.2.1) points out that the uncertainty is not, because any input within ±δx/2 gives the same display. It treats the input as spread evenly over that width, with a standard uncertainty of δx / √12, about 0.29 δx.

How is each one checked and expressed?

Each property has its own test and its own number. Accuracy is the one without a test of its own. In practice it gets its number through measurement uncertainty (VIM 2.26), which includes components from systematic effects, such as the uncertainty of a correction or of a reference standard's value, as well as random ones. The VIM also notes that an estimated systematic effect is sometimes left uncorrected and covered by an uncertainty component instead (2.26, note 1).

The table lists the usual checks.

Checking and expressing trueness, precision, accuracy and resolution
PropertyHow you check itWhat you recordWhere it is used
TruenessMeasure a reference of known value several times and compare the average with itThe bias, with a t test of whether it differs from zeroBias and linearity studies, calibration
PrecisionRepeat readings under stated conditionsA standard deviation and the conditionsRepeatability checks, Gage R&R, Type 1 gage studies
Accuracy at a pointCompare a reading with a calibrated referenceThe error of indication and the measurement uncertainty of the calibrationCalibration certificates
ResolutionRead the display step, or find the smallest change that moves the readingThe step, and its uncertainty termUncertainty budgets
Tip: NIST Technical Note 1297 (Appendix D.1) recommends putting numbers on uncertainties and standard deviations, not on the words accuracy and precision. Writing that the standard deviation under repeatability conditions is 2 µΩ is clear. Writing that the precision is 2 µΩ is not.

How do you read an accuracy spec?

On a data sheet, accuracy usually means a limit: the largest error the maker allows, which the VIM calls a maximum permissible error (4.26), or an accuracy class that sets such limits (4.25). It is not the error of your instrument. Calibration finds that. A limit given as a percentage only means something once you know what it is a percentage of.

  • An absolute limit, such as ±0.02 mm. The same band applies across the range.
  • Percent of reading. The limit scales with the value shown, so it is tight at the bottom of the range.
  • Percent of full scale. A fixed band equal to a share of the upper range limit, so it is large compared with small readings.
  • Percent of span. A share of the whole range between its two limits. The VIM's example (4.5) is a range of −10 V to +10 V with a span of 20 V, so ±1% of span is ±0.2 V where ±1% of the upper limit would be ±0.1 V.
  • Percent of reading plus counts or digits. Common on digital multimeters: one part scales with the reading and the other is a fixed number of steps of the last digit.
  • An accuracy class, as on pressure gauges, where the class number sets the limit as a share of the range.
The same ±0.5% on a 0 to 10 bar instrument, as a share of full scale and as a share of reading
Reading±0.5% of full scaleThat limit as a share of the reading±0.5% of reading
1 bar±0.05 bar5%±0.005 bar
5 bar±0.05 bar1%±0.025 bar
10 bar±0.05 bar0.5%±0.05 bar
Tip: Watch the units in humidity specs. A hygrometer specified to ±2 %rh may be off by 2 points of relative humidity, which is 5% of a 40 %rh reading. NPL's beginner's guide to humidity measurement gives the same warning: a change of 10% of reading is not a change of 10 %rh.

How this connects to Gage R&R, bias studies and uncertainty

Measurement systems analysis splits along the same line. Bias, linearity and stability studies look at where readings sit against a reference, which is trueness. Gage R&R looks at how readings scatter between repeats and between appraisers, which is precision. An uncertainty budget brings both into one stated uncertainty for a given measurement.

Where each study sits
StudyTrueness or precisionWhat it measures
Bias studyTruenessThe bias at one reference value, tested against zero
Linearity studyTruenessHow the bias changes across the operating range
Stability studyTrueness over timeHow the bias on one reference part moves as time passes
Gage R&RPrecisionRepeatability (EV) and reproducibility (AV), against the part variation or the tolerance
Type 1 gage studyBothThe spread (Cg), and the spread together with the bias (Cgk), on one reference part, against the tolerance
Uncertainty budgetBothEach significant source, combined into a standard and an expanded uncertainty
Tip: Type A and Type B are methods of evaluation, not labels for random and systematic. The GUM says so in clause 3.3.3, and NIST Technical Note 1297 repeats it (D.1.3). The uncertainty of a correction for a known bias can come from either type of evaluation, and so can the uncertainty from a random effect.

Common mix-ups

  • Calling the number of digits precision. That is resolution, and a fine display can sit on top of a scattered or biased measurement.
  • Reading a data sheet's accuracy as the error of your instrument. It is a limit. The error at each point comes from the calibration.
  • Fixing a bias by averaging. Averaging reduces the random part of the error of the mean and leaves the bias where it was.
  • Quoting a precision without its conditions. A repeatability standard deviation and a reproducibility standard deviation describe different things.
  • Treating a percent of full scale spec and a percent of reading spec as the same. Work both out at the readings you actually use.

Where Axiospec fits

Axiospec is calibration management software. A calibration record keeps the as-found and as-left readings at up to 100 test points, each with its own tolerance as an absolute band, a percentage of nominal or a percentage of span, so the limit sits next to the error found. The bias and linearity, Gage R&R, Type A uncertainty and resolution calculators on this site are standalone tools that need no account.

Common questions

What is the difference between accuracy and precision?
Accuracy is how close a measured value is to the true value. Precision is how closely repeated readings agree with each other under stated conditions. A gage with a constant offset can be very precise and still inaccurate.
What is trueness in measurement?
Trueness is how close the average of many repeated measurements is to a reference value (VIM 2.14). It reflects systematic error only, and it is put into numbers as bias, the average minus the reference value.
Can a measurement be precise but not accurate?
Yes. If every reading carries the same bias, the readings agree closely with each other and all miss the true value by about the same amount. Repeating the measurement does not reveal that, and only a comparison with a reference of known value does.
Is resolution the same as accuracy?
No. Resolution is the smallest change an instrument can show, such as the last digit of a display (VIM 4.14 and 4.15). An instrument can display 0.001 mm and still be off by more than that, which only a calibration shows.
What does ±1% of full scale mean?
It means the allowed error is 1% of the instrument's upper range limit at every reading. On a 0 to 10 bar gauge that is ±0.1 bar everywhere, which is 10% of a 1 bar reading. A percent of reading spec instead shrinks with the value shown.
How is precision expressed?
As a standard deviation, a variance or a coefficient of variation of repeated readings, with the conditions stated (VIM 2.15). Repeatability, intermediate precision and reproducibility are precision under different conditions, so each needs its own figure.

Put it into practice