Refractometer: how it works, how to use it and how to calibrate it
A refractometer measures how strongly a liquid bends light, its refractive index, and converts it to a concentration such as Brix, the percentage of sucrose by mass. To calibrate one, set it to zero with distilled water (refractive index 1.33299 at 20 °C), then check it with a sucrose solution or a certified standard of known value at a known temperature.
On this page
What is a refractometer and how does it work?
Light slows down and changes direction when it passes from air into a liquid. The refractive index says by how much. OIML R 108 defines it as the speed of light in standard air divided by its speed in the medium, a number with no unit (3.1). Distilled water has a refractive index of 1.33299 at 20 °C, and the more sucrose is dissolved in it, the higher the value (OIML R 108, Table 1).
Most refractometers find the refractive index from the critical angle of total internal reflection. The sample lies on a prism with a higher refractive index, and the instrument locates the boundary between light and dark that this angle creates. OIML R 142:2008 (revised in 2025 as R 142-1 and R 142-2) describes the method in 3.2.2, and the R 142 clause numbers in this guide are those of the 2008 edition. In a handheld optical refractometer you see that boundary as a shadow line across a scale in the eyepiece. A digital refractometer lights the prism with an LED, often close to the sodium D line at about 590 nm, and finds the boundary with a light sensor (OIML R 142, note to 4.2).
Refractive index changes with the wavelength of the light and with temperature, so it is stated with both. The usual reference is nD20: the sodium D line at 589.3 nm and a temperature of 20 °C (OIML R 142, 4.2).
| Scale | What it reports | Typical use |
|---|---|---|
| Refractive index (nD) | The refractive index itself, usually referred to 20 °C | Laboratories, oils, solvents and incoming checks |
| Brix | The mass fraction of sucrose in a solution of sucrose in water, in percent, read through the ICUMSA table | Juice, soft drinks, fruit, jam, syrup and sugar |
| Coolant concentration | A Brix reading multiplied by the coolant maker's refractometer factor | Water-mixed metalworking fluids in machine shops |
| Product scales, such as salinity | A scale the maker built for one kind of liquid | Only the liquid the scale was made for |
What does Brix mean?
Brix is the mass fraction of sucrose in a solution of sucrose in water, in percent: grams of sucrose per 100 grams of solution. OIML R 108 defines this mass fraction and writes it as % mas (3.2). OIML R 142 writes the same scale as % mas (Brix) in clause 6.5. A 10 % Brix solution holds 10 g of sucrose in 100 g of solution.
The link between Brix and refractive index is a table adopted by ICUMSA, the International Commission for Uniform Methods of Sugar Analysis, for 589.3 nm and 20 °C. OIML R 108 reproduces it (clause 4 and Table 1), and ICUMSA's current refractometry standard is SPS-3 (2000).
Juice, fruit and most real samples hold more than sucrose. ICUMSA says the refractometer then gives the dry substance only approximately, and asks for such results to carry the word refractometric. ISO 2173:2003, the refractometric method for soluble solids in fruit and vegetable products, also calls the result approximate when other substances are dissolved, and by convention reports it as the soluble solids content.
| Sucrose, % by mass (Brix) | Refractive index nD20 |
|---|---|
| 0 | 1.33299 |
| 5 | 1.34026 |
| 10 | 1.34782 |
| 15 | 1.35568 |
| 20 | 1.36384 |
| 30 | 1.38115 |
| 40 | 1.39986 |
| 50 | 1.42009 |
| 60 | 1.44193 |
How do you use a refractometer correctly?
A good reading depends as much on the sample as on the instrument. The points below come from makers' guidance, such as Mettler Toledo's Good Refractometry Practice brochure, and from the verification method in OIML R 142.
- Start with a clean, dry prism. Clean it with distilled water or a solvent that dissolves the last sample, then wipe it dry with a soft tissue. OIML R 142 allows no residue or stains on the prism before a reading (7.4.3).
- Use enough sample to cover the measuring surface. For most benchtop refractometers Mettler Toledo gives about 0.5 mL as the minimum, and about 1 mL for samples such as ethanol that spread thin. With only a few drops, stray light can reach the sensor and spoil the result.
- Avoid bubbles and air pockets. Press or stir a thick sample against the prism so it makes full contact.
- Let the sample reach the prism temperature before you read. Mettler Toledo suggests waiting 20 to 60 seconds for sticky or viscous samples such as syrup.
- Stir suspensions before you sample. For a sample that cannot be made uniform, such as ketchup, repeat the reading and use the mean.
- On a handheld refractometer, close the cover plate, hold it toward the light and focus the eyepiece until the scale and the boundary are sharp. Read where the boundary crosses the scale.
Why temperature matters, and what ATC does
Refractive index falls as temperature rises. A certified dodecane standard in Mettler Toledo's example has a refractive index of 1.42382 at 15 °C, 1.42164 at 20 °C and 1.41955 at 25 °C. For a 10 % sucrose solution, the correction table in OIML R 108 shows about 0.07 % Brix for each degree near 20 °C. The table gives the corrections at other temperatures.
Automatic temperature compensation (ATC) measures the temperature and corrects the reading to 20 °C. On Brix scales, makers such as Mettler Toledo and A.KRÜSS base it on the ICUMSA corrections. Mettler Toledo's RM benchtop refractometers, for example, can compensate Brix readings taken between 10 °C and 40 °C to 20 °C as per ICUMSA. Those corrections were worked out for sucrose in water: OIML R 108 has the thermometer of a hand refractometer show the correction values, and the refractometer or its instructions state that they refer to solutions of sucrose in water (11.6). For other liquids, a compensated reading can still carry a temperature error.
OIML R 142 states the principle for automated refractometers. Away from 20 °C the specification holds only with the temperature correction applied, and the correction is stated in the refractometer's specification (5.7). Mettler Toledo adds that an adjustment should be made at the temperature you usually measure at.
| Temperature of the reading | Correction to the Brix reading |
|---|---|
| 15 °C | Subtract 0.32 |
| 18 °C | Subtract 0.13 |
| 20 °C | None |
| 22 °C | Add 0.14 |
| 25 °C | Add 0.36 |
| 30 °C | Add 0.77 |
How to zero a refractometer with distilled water
Zeroing sets the start of the scale on a liquid whose value is known. Distilled water at 20 °C has a refractive index of 1.33299 at the sodium D line, which is 0 % on a Brix scale. OIML R 108 allows the water reading to differ from those values by no more than 0.2 of a scale interval (6.2), and OIML R 142 checks the low end of the range with double distilled water (7.4.2).
Put distilled water on the clean prism, let it reach the temperature, and read it before you change anything. That reading is the as-found zero. Adjust only if it is outside your limit. On a handheld refractometer with a calibration screw, turn the screw until the boundary sits on zero. On a digital one, use its zero or adjustment function as the maker describes. Then dry the prism.
Mettler Toledo points out that frequent adjustment does not guarantee accurate results. An adjustment changes the instrument's settings, and a wrong one, for example on a prism with residue, shifts later results by the same amount across the whole range. It recommends testing with water and adjusting only when the test fails in a repeatable way after a thorough cleaning. After an adjustment, measure water again to confirm it.
How to check a refractometer with reference solutions
Water checks one point, at the bottom of the scale. To see whether the rest of the scale is right, measure references of known value across the range you use. OIML R 142 verifies refractometers with glass plates or prisms of known index and with liquids: certified refractometric liquids and sucrose solutions (7.4).
A certified liquid should come with a certificate that states its refractive index relative to air at 20 °C, its temperature correction, the batch and the expiry date (OIML R 142, note to 7.4). R 142 expects the certified uncertainty to be no more than 0.00002 for water and 0.00003 for other refractometric liquids (Annex 3).
For a Brix refractometer, fresh sucrose solutions checked against the ICUMSA table also serve (OIML R 142, 7.4.4). They do not keep. In R 142's own procedure the test solutions are certified with an interference refractometer and used within two weeks (Annex 7). Mettler Toledo notes that Brix standards are often used for routine tests.
Pure sucrose is itself a reference material. NIST SRM 17g is crystalline sucrose certified for purity, a sucrose mass fraction of 99.941 % with a 95 % coverage interval of 99.761 % to 100.000 %. NIST issues it for calibrating polarimeters and its certificate gives no refractive index, so a solution made from it takes its Brix value from the weighing and the ICUMSA table.
R 142 also sets the reading pattern. Measure each reference five times, removing and reapplying the sample each time, with at least 60 seconds between readings for cleaning and drying (7.4.3). The error at that point is the mean reading minus the reference value (7.5.3).
| Liquid | Nominal nD |
|---|---|
| Double distilled water | 1.3330 |
| n-heptane | 1.3877 |
| Methylcyclohexane | 1.4238 |
| Toluene | 1.4967 |
| α-bromonaphthalene | 1.6580 |
What limits apply to a refractometer?
OIML R 108 is the legal metrology recommendation for refractometers that measure the sugar content of fruit juices. It sets the maximum permissible error on verification at half a scale interval for analog scales and one scale interval for digital scales (6.1). On hand refractometers, a Brix scale has an interval of 0.1 %, 0.2 % or 0.5 % when analog and 0.1 % when digital (10.2.3). A handheld refractometer with a 0.2 % Brix scale interval therefore has a limit of ±0.1 % Brix.
OIML R 142:2008, for automated refractometers, left the maximum permissible error to the operating manual (5.2). Its 2025 revision still has the manual state it, and sets it at ±1 scale interval at type approval, initial verification and verification after repair, increased by half a scale interval for instruments in service (R 142-1:2025, 5.1.1 and 5.1.2). Both editions ask for the error at each point to be within it, and for the expanded uncertainty of the check, at k = 2, to be no more than a third of that limit (7.5.9 and 7.5.10 in 2008, 6.5.9 and 6.5.10 in R 142-2:2025). A digital refractometer under R 142 resolves 0.0001 in refractive index and 0.1 % Brix, or better (6.5).
Outside legal metrology, your limit comes from the maker's specification or from what the measurement has to decide. For the routine water test, Mettler Toledo suggests a tolerance of twice the resolution plus the operator's repeatability. With a resolution of 0.0001 and a repeatability of 0.00005, that rounds up to ±0.0003.
Coolant concentration in machine shops
Water-mixed metalworking fluids are checked with a refractometer because it is quick. Quaker Houghton, a coolant maker, describes the method. The refractometer measures the total actives of the product, the reading is in Brix, and the product's own refractometer factor turns it into a concentration.
The factor matters. A reading of 5 means 5.0 % only if the product is 100 % active. If the concentrate contains 20 % water, a 10 % dilution reads 8.0 at best. For a product whose maker gives a factor of 1.25, a reading of 8.0 means 10.0 %. Quaker Houghton says each supplier should provide a chart of readings over the normal concentration range.
- Remove tramp oil from the sample first. Quaker Houghton notes that it throws the reading off.
- Zero the refractometer on distilled water before the check.
- Choose a scale that fits. Quaker Houghton names 0 to 10 and 0 to 32 as the common ranges, and a wider scale has fewer graduations, which costs accuracy.
- Calibrate the refractometer itself on its Brix scale with water and a sucrose standard. The refractometer factor belongs to the coolant, not to the instrument.
What changes for food, beverage and sugar?
- Juice, fruit and vegetable products. ISO 2173:2003 is the refractometric method for soluble solids in these products, particularly suited to thick products, products with suspended matter and products rich in sugar. ISO confirmed it again in 2022.
- Sugar. ICUMSA SPS-3 (2000) covers refractometry and its tables for sucrose, glucose, fructose and invert sugar. For a pure sucrose solution the refractive index gives the exact dry substance. For technical solutions it gives an approximation that depends on purity.
- Legal control. Where a country applies OIML R 108, its control can include verifying new, repaired or readjusted refractometers and checking those in service periodically, in whole or in part as its national regulations set (clause 12). OIML R 124 covers refractometers for the sugar content of grape must.
How often should a refractometer be checked and calibrated?
No standard fixes one interval for every refractometer. Mettler Toledo suggests routine tests at short intervals, days or weeks, and notes that a water test is often done every day. It suggests a separate test with certified, traceable standards at longer intervals, such as months or a year.
Where OIML R 108 applies, a new, repaired or readjusted refractometer is verified, and periodic checks follow national rules (12.2 and 12.3). ILAC-G24:2022 / OIML D 10:2022 lists the factors for choosing an interval, such as how much the instrument is used (5.1), and describes how to review it from calibration history (6).
- How to set calibration intervals
- Intermediate checks between calibrations
- Calibration interval calculator
What should the calibration record show?
OIML R 142:2008 Annex 8 gave a model verification report. It named the instrument and the room conditions. It then recorded the temperature readings, the mean reading, the error and its uncertainty, the conformity decision, who did the work and the date. The details of each reference followed the note to R 142 clause 7.4, and Annex 7 for sucrose solutions. Whether ATC was on, the as-found and as-left readings and the next due date are good practice for your own records.
- The refractometer's ID, model, serial number, scale and range, and whether ATC was on.
- The room temperature, pressure and humidity during the work.
- Each reference used: its name, its certified value at 20 °C, the lot, the certificate and the expiry date. For a sucrose solution, how and when it was made.
- The prism temperature at each reading.
- The readings at each point, their mean, the error and the limit.
- The water reading before and after any adjustment, and the as-found and as-left results at the other points.
- The uncertainty, the decision on conformity, the date, who did the work and the next due date.
What if the as-found reading is out of tolerance?
Record the water and standard readings before anyone adjusts the refractometer. If an as-found error is outside its limit, ISO 9001 clause 7.1.5.2 asks you to determine whether earlier results were adversely affected and to act as needed. For a refractometer, look at what its readings decided since its last good check, such as juice or syrup batches released on Brix, coolant mixes adjusted on its readings, or goods accepted on refractive index.
How to calibrate a refractometer, step by step
- Identify the refractometer, its scale and range, whether it has ATC and the limit that applies, from the maker's specification or your own requirement.
- Inspect the prism for scratches and residue, clean it with distilled water or a solvent that dissolves the last sample, and dry it with a soft tissue.
- Let the refractometer, the water and the references reach the same temperature, ideally 20 °C, and note it.
- Measure distilled water before adjusting anything and record the as-found zero.
- Measure each reference across the range you use five times, cleaning the prism and reapplying the sample between readings. Record these as-found readings.
- Work out the error at each point as the mean reading minus the reference value, corrected to 20 °C where needed, and compare it with the limit.
- Adjust only if an error is outside the limit and repeatable after cleaning, and set the zero on water as the maker describes.
- Measure the water and the references again and record the as-left results.
- If an as-found result was out of tolerance, assess what the refractometer measured since its last good check. Then set the next due date.
Where Axiospec fits
Axiospec is calibration management software. It keeps each refractometer, handheld or benchtop, with its calibration history, certificates and next due date, and the interval can be set in days, months or years. One calibration record holds up to 100 test points, each with its nominal value, its own tolerance, a name and the as-found and as-left readings, so water and each standard can be a point of their own. With a standard such as ISO 9001 or ISO/IEC 17025 selected for the workspace, each calibration also keeps the room temperature and humidity and names the reference standard used, as free text or as a link to a master instrument, so the lot of a Brix standard can sit on the record.
On an in-house calibration, a final reading outside its tolerance sets the result to Fail, and recording a pass anyway takes a written reason. An accepted Fail or Damaged puts the instrument in Quarantined. With a standard selected, any approved Fail, Damaged or Pass with as-found out of tolerance needs an impact written, and unrecorded ones show in Needs attention. Axiospec does not run the calibration or pick the points, the limit or the interval. Those stay with you or your lab.
The free plan covers up to 50 active instruments. A new workspace starts with a 45-day trial of Professional, with room for 1,500 active instruments and no credit card. We load your existing records for you at no charge on every plan.
Common questions
- How do you calibrate a refractometer?
- Clean the prism, put distilled water on it at a known temperature, ideally 20 °C, and check that it reads zero (0.0 % Brix, or nD 1.33299). Then measure a sucrose solution or a certified standard of known value near the readings you use, and compare the result with its value at that temperature. Adjust only if the error is outside your limit, then measure again.
- What should a refractometer read with distilled water?
- 0.0 on a Brix scale and 1.33299 on a refractive index scale, for distilled water at 20 °C and the sodium D line. OIML R 108 allows the water reading to differ from these values by no more than 0.2 of a scale interval.
- What does ATC mean on a refractometer?
- Automatic temperature compensation. The refractometer measures the temperature and corrects the reading to 20 °C. On a Brix scale the corrections are those for sucrose solutions, so readings of other liquids away from 20 °C can still be off.
- What is Brix?
- Brix is the mass fraction of sucrose in a solution of sucrose in water, in percent. A refractometer reads it from the refractive index through the ICUMSA table, so for juice and other mixtures the result is the sucrose equivalent of the dissolved solids.
- How do you check coolant concentration with a refractometer?
- Zero the refractometer on distilled water, remove any tramp oil from the sample, read the Brix value and multiply it by the refractometer factor that the coolant maker gives for that product. The reading equals the concentration only for a product that is 100 % active.
- How often should a refractometer be calibrated?
- No standard sets one interval. Makers such as Mettler Toledo suggest a water test at short intervals, often daily, and a test with certified standards every few months or once a year. Check again after any repair, drop or adjustment.
Put it into practice
Got a gage list? Send it over and we load it for you, usually in a couple of business days. Free on every plan. Then log every calibration to a tamper-evident audit trail and produce records on demand.
Axiospec is a documentation and workflow tool. It helps you keep clean, traceable, audit-ready records. Certification depends on your own processes, scope and assessor.
Take the audit-prep checklist with you
First the checklist, with each item phrased the way an auditor asks it. Then four short emails with practical tips. You can unsubscribe in one click.
First we email you a link to confirm your address. Nothing else comes until you confirm, and you can unsubscribe from any email after that. See our Privacy Policy.