Infrared thermometer calibration: blackbody, emissivity and spot size
An infrared thermometer is calibrated by aiming it at a blackbody source of known temperature at a set distance, with its emissivity setting matched to the source, and comparing its reading with the source at several temperatures. ASTM E2847 is the test method for wideband instruments below 1000 °C, and because the thermometer reads thermal radiation and never touches the object, the source size, emissivity and reflections matter as much as the temperature.
On this page
What does an infrared thermometer actually measure?
An infrared thermometer, also called a radiation thermometer or pyrometer, works like a camera. A lens and a band-pass filter focus thermal radiation onto a detector, and the instrument converts the signal into a temperature (NPL Good Practice Guide No. 125, Chapter III). Because it reads radiation and not the temperature of a sensor in contact with the object, it must be calibrated against a known source of thermal radiation.
The waveband depends on the temperature it is made for. Instruments for room temperature commonly work between 8 µm and 14 µm, others between 3 µm and 5 µm, and instruments for high temperatures, above about 700 °C, between 0.65 µm and 1.0 µm (NPL GPG 125).
Which standards cover infrared thermometer calibration?
ASTM E2847 does not cover instruments above 1000 °C or narrowband instruments, and it does not treat the design of infrared thermometers. It may also be applied to other radiation thermometers and to thermal imagers.
| Document | What it covers |
|---|---|
| ASTM E2847-21(2026) | Test method for calibration and accuracy verification of wideband infrared thermometers below 1000 °C, primarily handheld instruments. It includes a structure for calculating uncertainty and the required content of the report. |
| ASTM E2758 | Guide to selecting and using wideband, low temperature infrared thermometers. |
| ASTM E1256 | Test methods for radiation thermometers of the single waveband type. |
| IEC TS 62492-1:2008 | Defines the technical data a radiation thermometer's data sheet gives, such as field of view, distance ratio and size-of-source effect. |
| BIPM Guide to the Realization of the ITS-90, Radiation Thermometry | How the temperature scale is realized by radiation thermometry above the silver point, 961.78 °C. |
Blackbody cavity or flat-plate source?
NPL Good Practice Guide No. 125 calls the blackbody the reference condition: a calibration uses a source of known temperature with an emissivity of, or close to, 1, and calibrations at other emissivities can then be calculated. A well-designed cavity comes close. NPL's variable-temperature sources are generally at least 0.999, and NPL notes that a well-built cavity can exceed 0.9999. The temperature comes from a calibrated contact sensor, such as a platinum resistance thermometer or a thermocouple, which gives traceability to the ITS-90. Often the cavity and sensor are calibrated together against a reference blackbody.
Cavities can be large and slow, so blackened plates, usually grooved, are common. A plate is heated from behind with a sensor near its surface. Because the surface is exposed, its emissivity is likely to be substantially below 1, and there are temperature differences across the plate and between the sensor and the surface. NPL's guide says the plate itself must be calibrated against a reference blackbody.
| Source | Emissivity | What to watch |
|---|---|---|
| Blackbody cavity | Generally at least 0.999 for a good variable-temperature cavity (NPL) | The aperture has to be larger than the thermometer's spot. NPL's industrial sources have maximum apertures of 15 mm to 75 mm, depending on the source. |
| Flat plate | Below 1. Use the value on the plate's certificate | Reflected room radiation and temperature differences across the plate |
How should the emissivity be set during calibration?
Set the thermometer to match the source where the setting can be changed. IEC TS 62492-1:2008 clause 4.1.1 gives data sheet values at an emissivity setting of 1 unless stated otherwise, and for an instrument with a fixed setting other than 1 it requires that value to be stated with the specification.
A real surface also reflects radiation from its surroundings. NPL Good Practice Guide No. 125 writes the radiance reaching the thermometer as L = εL(Ts) + (1 − ε)L(Tb), where ε is the emissivity, Ts the surface temperature and Tb the background temperature. With a cavity and a setting of 1, the second term is close to zero. With a flat plate, or a thermometer fixed at a setting that differs from the source, use that relation to work out what the thermometer should read from the source emissivity and the background radiance. ASTM E2847 gives the test setup and a structure for the uncertainty calculation.
A certificate says nothing about the emissivity of your own part. NPL warns that errors of tens of degrees are common when the emissivity of a surface is wrongly estimated, so service use needs its own control: a known value for that material, or a high-emissivity coating where one is allowed (NPL GPG 125).
Distance, spot size and the size-of-source effect
The distance ratio, often written D:S, is the measuring distance divided by the diameter of the field of view (IEC TS 62492-1:2008 clause 3.1.6). A thermometer rated 12:1 at 300 mm sees a spot 25 mm across. The specification should state from which part of the instrument the distance is measured (clause 4.1.1.4.2), and the calibration should use that point.
The field of view has no sharp edge. IEC TS 62492-1:2008 clause 4.1.1.5 therefore states it as the diameter that contains a stated fraction of the signal, at least 90% and typically 90%, 95% or 99%. Some radiation from outside that circle still reaches the detector, so the reading changes with the size of the radiating area. That is the size-of-source effect (clause 3.1.7). The BIPM radiation thermometry guide attributes it to scatter and reflections in the optics and notes that dust on the objective can change it.
Two practical rules follow. Use a source significantly larger than the stated spot, as NPL's guide advises for any surface measured. And test for the effect: change the source aperture or the distance and see whether the reading moves. IEC TS 62492-1:2008 clause 4.1.1.7 compares the nominal field of view with twice its area.
What ambient conditions matter?
- Warm-up. Let the thermometer stabilize in the room first. IEC TS 62492-1:2008 treats its specifications as valid after the warm-up time and under stable ambient conditions, and defines an added uncertainty when the instrument's own temperature departs from them (clause 3.1.10).
- Reflections. Lights, heaters and even people add to the signal at low temperatures (NPL GPG 125). A flat plate reflects part of what surrounds it, so keep its surroundings steady.
- Line of sight. Dust, water vapor and carbon dioxide can absorb or scatter radiation (NPL GPG 125). Keep the path short and clear.
- Lens. Dust on the objective can change the size-of-source effect (BIPM radiation thermometry guide). Clean it the way the maker says to.
- Record the room temperature and humidity, since a humidity effect may be stated too (IEC TS 62492-1:2008 clause 3.1.11).
What is the difference between an accuracy check and a calibration?
ASTM E2847 is titled for both calibration and accuracy verification. A calibration compares the thermometer with a traceable source under stated conditions and reports the error and its uncertainty at each point. A check confirms that the thermometer still behaves, at one or a few points, and it does not replace a calibration or reset the due date.
A check only tells you something when it is repeated under the same conditions: the same stable source, the same distance and the same emissivity setting, compared with earlier results. ISO/IEC 17025:2017 clause 6.4.10 requires a procedure for intermediate checks where they are needed to keep confidence in the equipment.
Pyrometers, narrowband instruments and very high temperatures
ASTM E2847 stops at 1000 °C and excludes narrowband instruments. ASTM E1256 gives the test methods for single waveband radiation thermometers. Instruments for high temperatures use short wavelengths, where the temperature error may be 10 to 20 times smaller than the emissivity error. An emissivity error of 1% then gives less than 0.1% of the temperature in kelvin (NPL GPG 125).
Above the freezing point of silver, 961.78 °C, the ITS-90 is defined through radiation thermometry, with the silver, gold or copper fixed point as the reference (BIPM Guide to the Realization of the ITS-90, Radiation Thermometry). Pyrometers in that range trace to fixed-point or high-temperature blackbodies through a calibration laboratory.
Two-color or ratio pyrometers help with small targets that do not fill the field of view and in dirty atmospheres, but they do not remove the emissivity problem: the emissivity cancels only if it is the same at both wavelengths (NPL GPG 125).
How often should an infrared thermometer be calibrated?
No standard sets the interval, and NIST does not recommend a set recalibration interval. ILAC-G24:2022 / OIML D 10:2022 lists the factors to weigh, including the tendency to drift and the severity of use. Start from the maker's recommendation, then adjust it from your as-found history and the results of your checks. Shorten it after a drop, damage to the lens or use near the limits of the range, and record the reason for each change.
What should the certificate and record show?
ASTM E2847 specifies the elements of a calibration report (section 5.3), and ISO/IEC 17025:2017 clause 7.8 sets the general content of a report for accredited work. Keep at least the following.
- The thermometer and its emissivity setting during the calibration.
- The source type, cavity or flat plate, with its size, its emissivity and the reference used for its temperature.
- The measuring distance and the point on the thermometer it was measured from.
- The temperatures calibrated, with the source value, the reading, the error and the uncertainty at each point.
- The room temperature and humidity.
- The readings before and after any adjustment, the date, who did the work and the next due date.
What if an infrared thermometer is out of tolerance?
Record the as-found readings before anyone adjusts it. If a reading is outside its limit, the thermometer may have been out of tolerance for an unknown part of the time since its last good calibration or check. ISO 9001 clause 7.1.5.2 expects you to look back at what it measured and deal with anything that may be wrong. For an infrared thermometer, look at what its readings decided, such as an oven or furnace temperature, a process limit or a safety check.
How to calibrate an infrared thermometer, step by step
- Identify the thermometer, its range, its emissivity setting, its distance ratio and the limit that applies.
- Choose calibration points across the range you actually use, including its ends.
- Choose a source larger than the thermometer's field of view at the calibration distance, and check its certificate and due date.
- Let the source and the thermometer stabilize, note the room temperature and clean the lens the way the maker says to.
- Set the emissivity to match the source, or record the fixed value so you can correct for it.
- Align the thermometer on the source at the stated distance, measured from the point the maker defines.
- Record the as-found readings at each point before any adjustment.
- Check the size-of-source effect at one point by changing the aperture or the distance and watching the reading.
- Work out the error at each point, apply any emissivity correction and compare with the limit.
- If it fails, adjust or repair it, repeat the points as left, assess what it measured and set the next due date.
Where Axiospec fits
Axiospec is calibration management software. It keeps each infrared thermometer and pyrometer with its calibration history, certificates and next due date. Each calibration names the reference standard used, as free text or as a link to a master instrument, such as your blackbody source. One calibration record holds up to 100 test points, each with its nominal temperature, its own tolerance and the as-found and as-left readings, and the record also keeps the room temperature and humidity.
On an in-house calibration, a final reading outside its tolerance sets the result to Fail, and recording a pass anyway takes a written reason. An accepted Fail or Damaged puts the thermometer in Quarantined. Any approved Fail, Damaged or Pass with as-found out of tolerance needs an impact written, and unrecorded ones show in Needs attention. Axiospec does not run the comparison or pick the points, the limit or the interval. Those stay with you or your lab.
The free plan covers up to 50 active instruments. A new workspace starts with a 45-day trial of Professional, with room for 1,500 active instruments and no credit card. We load your existing records for you at no charge on every plan.
Common questions
- How do you calibrate an infrared thermometer?
- Aim it at a blackbody source of known temperature at the distance the thermometer specifies, with the source larger than the spot. Set the emissivity to match the source, read the thermometer at several temperatures across its range and compare with the source. Record the error and the uncertainty at each point.
- What emissivity setting should be used for calibration?
- Match the setting to the source: 1 or close to it for a cavity blackbody, and the value on the plate's certificate for a flat plate. If the thermometer has a fixed setting that differs, calculate what it should read from the source emissivity and the background radiance.
- What does the distance to spot ratio mean?
- It is the measuring distance divided by the diameter of the field of view, so a 12:1 thermometer at 300 mm sees a spot 25 mm across. The field of view has no sharp edge, so a target only slightly larger than the stated spot can give a reading that depends on the target's size. That is the size-of-source effect.
- What is the difference between an accuracy check and a calibration?
- A calibration compares the thermometer with a traceable source under stated conditions and reports the error and its uncertainty at each point. A check confirms that it still behaves against a stable source, and it does not replace a calibration or reset the due date.
- Which standard covers infrared thermometer calibration?
- ASTM E2847-21(2026) is the test method for wideband infrared thermometers below 1000 °C, mainly handheld instruments. Single waveband radiation thermometers are covered by ASTM E1256, and IEC TS 62492-1:2008 defines the technical data for radiation thermometers.
- How often should an infrared thermometer be calibrated?
- No standard sets an interval, and NIST does not recommend a set one. Base it on the maker's recommendation, the as-found history and how the thermometer is used, and shorten it after a drop or damage to the lens.
Put it into practice
Got a gage list? Send it over and we load it for you, usually in a couple of business days. Free on every plan. Then log every calibration to a tamper-evident audit trail and produce records on demand.
Axiospec is a documentation and workflow tool. It helps you keep clean, traceable, audit-ready records. Certification depends on your own processes, scope and assessor.
Take the audit-prep checklist with you
First the checklist, with each item phrased the way an auditor asks it. Then four short emails with practical tips. You can unsubscribe in one click.
First we email you a link to confirm your address. Nothing else comes until you confirm, and you can unsubscribe from any email after that. See our Privacy Policy.