Type 1 gage study: Cg and Cgk formulas, limits and a worked example

10 min read Published Updated

A Type 1 gage study checks a gage on its own: one operator measures one calibrated reference part, usually 50 times and at least 25, and the spread and the bias of the readings are compared with the tolerance. The results are Cg = 0.2T / 6s and Cgk = (0.1T − |x̄ − xm|) / 3s, and both usually have to reach 1.33 before the gage goes on to a Gage R&R study.

On this page

What is a Type 1 gage study?

A Type 1 gage study looks at the measuring equipment alone, without the effects of different parts or different operators. One person measures one reference part many times. The part's value is known from calibration. The readings show two things. Their spread is the gage's repeatability, and the distance of their average from the reference value is its bias.

The method comes from automotive practice. In German it is Verfahren 1, procedure 1 of a measurement system analysis. Bosch Booklet 10, Capability of Measurement and Test Processes (edition November 2019), publishes it in full. The booklet notes that procedure 1 is not part of the AIAG MSA manual but extends the manual's bias analysis, and that many automotive manufacturers require it in their own guidelines.

Run it before a Gage R&R. A gage that scatters too much or reads off on a reference part will not pass a study with real parts and several operators, so there is no point spending their time until the gage passes on its own. Bosch Booklet 10 makes procedure 1 a prerequisite for its Gage R&R procedures.

The Cg and Cgk formulas

Data case: n readings of one reference part by one operator, taken under repeatability conditions, with the part removed and put back before each reading. From them you need the mean x̄ and the sample standard deviation s, calculated with n − 1. The tolerance is T = USL − LSL, and xm is the reference value of the part.

Cg, the potential capability index, compares 20 percent of the tolerance with a spread of six standard deviations: Cg = 0.2T / 6s.

Cgk, the critical capability index, also counts the bias. It takes half of the 20 percent band, 0.1T, subtracts the bias and compares what is left with three standard deviations: Cgk = (0.1T − |x̄ − xm|) / 3s.

These are the formulas of Bosch Booklet 10, section 4.1. Minitab writes them with two settings: K, the percentage of the tolerance (20 by default), and the number of standard deviations in the study variation (6 by default, which covers 99.73 percent of a normal distribution). Changing either changes the result, so record the settings with the study.

Because Cgk subtracts the bias, it can never be larger than Cg. The two are equal only when the bias is zero, and Cgk turns negative when the bias is larger than 10 percent of the tolerance.

Tip: The 20 percent is a convention. Some customers set a different K or a different multiplier, so check their measurement system requirements before you report a pass.

What Cg and Cgk value is acceptable?

Bosch Booklet 10 requires Cg ≥ 1.33 and Cgk ≥ 1.33. It adds that the current minimum values are set by Bosch's directive CDQ 0301, so the figure can change. Minitab also calls 1.33 a common benchmark for a capable gage.

The figure has a plain meaning. With K = 20, the share of the tolerance taken up by the gage's 6s spread is 20 percent divided by Cg. At Cg = 1.33 that is 15 percent, which is the %Var (repeatability) many reports show next to Cg. Bosch's booklet also allows for higher requirements, such as 1.67 or 2.00. Use the figure in your customer's requirement or control plan.

Cg and the share of the tolerance taken by the gage's 6s spread, with K = 20
Cg6s as a share of the tolerance
1.0020%
1.3315%
1.6712%
2.0010%

How many readings, and what reference part?

Bosch Booklet 10 asks for 50 readings, and at least 25. With fewer readings the standard deviation is less certain, so a result close to 1.33 can pass or fail by chance.

The reference part has to be stable and give a clear reading under repeatability conditions. It should carry the same characteristic as the production parts, and it can be made from one. It must be labeled, calibrated and kept under your gage control. Where possible its value should sit near the middle of the tolerance.

Its calibration uncertainty should be small next to the tolerance. Bosch's rule of thumb is Ucal < 0.01T in the ideal case and Ucal < 0.1T at least.

Check the resolution before you start. Bosch Booklet 10 asks for a resolution below 5 percent of the tolerance, and Minitab uses the same 5 percent guideline. Where no finer gage is available, Bosch accepts up to 10 percent. Beyond that it treats a capability study as not reasonable.

Worked example: a bench gauge on a setting master

A shaft diameter is toleranced at 10.00 ± 0.03 mm, so T = 0.060 mm. The bench gauge reads to 0.0005 mm, which is 0.8 percent of the tolerance. The setting master is calibrated at xm = 10.0010 mm with Ucal = 0.0004 mm, under 1 percent of the tolerance. One operator measures it 50 times and takes it out and puts it back each time. The 50 readings have x̄ = 10.0018 mm and s = 0.0012 mm.

  • Cg = (0.2 × 0.060) / (6 × 0.0012) = 0.012 / 0.0072 = 1.67
  • Bias = 10.0018 − 10.0010 = 0.0008 mm
  • Cgk = (0.1 × 0.060 − 0.0008) / (3 × 0.0012) = 0.0052 / 0.0036 = 1.44
  • Both indices are above 1.33, so the gauge passes.
Tip: Now suppose the same gauge had been set wrongly and read 10.0035 mm on average. The bias becomes 0.0025 mm and Cgk = (0.006 − 0.0025) / 0.0036 = 0.97, while Cg stays at 1.67. The gauge is precise but off. Resetting it against the master is the fix, not a new gauge.

Is the bias significant?

A Type 1 study can also tell you whether the bias is real or just scatter. The usual check is a one-sample t-test: t = (x̄ − xm) / (s / √n), compared with the t distribution on n − 1 degrees of freedom. Bosch Booklet 10 (appendix C) describes it at a 95 percent confidence level, and Minitab reports it as a p-value.

In the example, t = 0.0008 / (0.0012 / √50) = 4.7. The two-sided critical value for 49 degrees of freedom at 95 percent is 2.01, so the bias is significant even though Cgk passes. Bosch notes, citing the AIAG MSA manual, that a significant bias should generally be corrected by adjusting the gage. Where that is not possible, each result can be corrected by the known bias.

The test compares the bias with the scatter, not with the tolerance. Bosch points out that a very precise gage can show a significant bias that is technically negligible, while a noisy one can hide a large bias. Read the test next to Cgk, not instead of it.

What to do when a Type 1 study fails

Read Cg and Cgk together. The pattern tells you where to look.

Bosch Booklet 10 gives a checklist that runs from fixtures and measuring forces to cleanliness, temperature and operator handling. It asks for a root cause analysis before a measurement process is approved with conditions or rejected. Fix the cause and repeat the study. If the gage cannot reach 1.33 on this tolerance, the options are a better gage, a different measuring method or a review of the tolerance with the design owner.

Do not delete readings to make a study pass. Bosch's workflow for procedure 1 says outright that no measured value is to be discarded.

Reading a failed Type 1 study
ResultWhat it meansWhere to look
Cg passes, Cgk failsThe gage repeats well but reads off the reference value.The zero or setting, the setting master, temperature and the way the part is located.
Cg failsThe gage scatters too much for this tolerance.Resolution, clamping and locating, wear or play in the gage, vibration and dirt.
Both pass, but the bias is significantThe gage is capable but has a real offset.Adjust the gage or correct the results, and record which you did.

How it relates to Gage R&R, linearity and stability

A Type 1 study is the first step of a measurement system analysis. The other studies build on it.

  • Gage R&R. Once the gage passes on its own, a Gage R&R adds real parts and, where people affect the reading, several operators. Bosch Booklet 10 calls these procedure 2 (with operator influence) and procedure 3 (without). On the same standard deviation, %GRR = 20% / Cg, so Cg = 1.33 matches 15 percent. Bosch notes that the two classifications do not agree in every case: a gage at Cg = 1.33 is capable under procedure 1 but only conditionally capable at %GRR = 15 percent under procedures 2 and 3.
  • Bias. The bias found here is the same quantity an AIAG bias study estimates from repeated readings of one reference part, and the t-test is the same.
  • Linearity. A Type 1 study checks one point of the range. To cover the range, Bosch suggests repeating procedure 1 on several masters spread across it, at least 5 where possible, or the AIAG linearity study when a customer asks for it.
  • Stability. A Type 1 study is a snapshot. Bosch procedure 5 follows the gage over time by measuring a reference part at least three times per interval and plotting the results on a control chart. Bosch advises against taking the chart's standard deviation from the Type 1 study, because that is a short-term figure.

When a Type 1 study does not fit

Cg and Cgk need a tolerance width. When a characteristic has one specification limit and a natural limit, such as a surface roughness that cannot go below zero, Bosch Booklet 10 uses T* = USL − 0, and Minitab does the same. With one limit and no natural limit there is no tolerance at all. Bosch then sets an acceptance limit for each later reading instead, such as USL − (x̄ − xm) − 4s for an upper limit.

The statistics assume roughly normal readings. Bosch also says procedures 1 to 4 apply only in part, or not at all, to some quantities such as hardness and torque, and to parts that are not uniform.

For measurement tasks that change often, as in development and test departments, Bosch recommends determining the measurement uncertainty rather than capability indices. ISO 22514-7 also judges a measurement process from its uncertainty, with acceptance criteria stated as capability figures or ratios (QMS and QMP).

How to run a Type 1 gage study, step by step

  1. Confirm that the gage is calibrated and that its resolution is below 5 percent of the tolerance.
  2. Choose a stable reference part with the same characteristic as the production parts, calibrated with an uncertainty well below the tolerance, and with a value near the middle of the tolerance.
  3. Set the gage up as it will be used in production, with the same fixture, settings and measuring point.
  4. Measure the reference part 50 times, and at least 25, removing and replacing it before each reading. Record every reading and discard none.
  5. Calculate x̄, s and the bias, then Cg and Cgk, and run the t-test on the bias.
  6. Compare Cg and Cgk with your requirement, usually 1.33. If either fails, find the cause, fix it and repeat the study.
  7. When the gage passes, go on to a Gage R&R with real parts and the usual operators.
  8. File the study with the gage's records: the date, the operator, the reference part and its certificate, the readings, the K and multiplier settings and the result.

Where Axiospec fits

Axiospec is calibration management software and does not calculate Cg or Cgk. The Type 1 gage study template on this site does, in Excel, with no account needed.

With the Gage R&R studies module switched on, an instrument's page can hold Gage R&R, bias, linearity and attribute agreement studies. The bias study there takes repeated readings of a reference and runs the same two-sided t-test at 95 percent. You can attach the Type 1 workbook to the instrument as a document, next to its calibrations, certificates and due date. The setting master behind the study can be tracked as an instrument too, with its own calibration and due date.

The free plan covers up to 50 active instruments. A new workspace starts with a 45-day trial of Professional, with room for 1,500 active instruments and no credit card.

Common questions

What is a Type 1 gage study?
It is a study of the gage alone, run before a Gage R&R. One operator measures one calibrated reference part about 50 times, and at least 25, and the spread and bias of the readings are compared with the tolerance as Cg and Cgk. In German it is known as Verfahren 1.
How do you calculate Cg and Cgk?
Cg = 0.2T / 6s and Cgk = (0.1T − |x̄ − xm|) / 3s, where T is the tolerance, s the standard deviation of the readings, x̄ their mean and xm the reference value. The 0.2 is 20 percent of the tolerance, and 6s covers 99.73 percent of normally distributed readings. Cgk is never larger than Cg.
What is an acceptable Cg and Cgk?
1.33 or more for both is the usual requirement, and it is the limit in Bosch Booklet 10. At Cg = 1.33 the gage's 6s spread uses 15 percent of the tolerance. Higher requirements such as 1.67 are also used.
What does it mean when Cg passes and Cgk fails?
The gage repeats well but its average reading is off the reference value. Check the zero or setting, the setting master and the way the part is located. Adjusting or resetting the gage is the usual fix, not a new gage.
How many measurements does a Type 1 study need?
Bosch Booklet 10 asks for 50 readings of the reference part, and at least 25. Take each reading the way it will be taken in production, removing and replacing the part each time.
What is the difference between a Type 1 study and a Gage R&R?
A Type 1 study looks at the gage alone on one reference part. A Gage R&R adds several production parts and, where people affect the result, several operators, and it splits the variation into repeatability and reproducibility. Run the Type 1 study first.