A knife-edge straightedge or square is checked by the light gap: set against a surface in front of an even light, any gap shows as a line of light, white from about 2 µm and coloured below that. A square's own error is measured against a better reference, or by reversal, which turns the square through 180° so its error changes sign and can be separated from the reference's.

On this page

How does the light-gap method work?

The light-gap method, Lichtspaltverfahren in German, uses a sharp edge as the reference. Stand a knife-edge straightedge or square on the surface to be checked, square to it, with an even light behind the joint, and look along the line of contact. Wherever the surface falls away from the edge, light comes through.

The colour of the light gives a rough width. INSIZE's instructions for its knife-edge straightedges give the guide in the table, and they suggest a soft fluorescent light 60 mm to 90 mm from the gap. The values are approximate and depend on the light and on the eye.

For a number rather than a colour, compare the gap with a standard light gap. INSIZE describes one made from gauge blocks set on an optical flat: two blocks of equal height at the ends carry the straightedge, and blocks 1 µm, 2 µm, 3 µm and 4 µm shorter between them leave gaps of known width under it. The comparison only holds when both gaps see the same light at the same distance, and INSIZE limits the test surface to a roughness of Ra 0.2 µm, because a rougher surface scatters the light and wears the edge.

Approximate colour of a light gap under a knife edge, from INSIZE's instructions for knife-edge straightedges
Gap widthWhat you see
About 0.5 µmLight starts to pass
About 0.8 µmBlue light
1.25 µm to 1.75 µmRed light
Over about 2 µm to 2.5 µmWhite light
Tip: Check in more than one direction. INSIZE has you place the straightedge along the length, across and on the diagonals, and lift it to each new position rather than drag it, since dragging wears the edge.

Which straightedge or square fits the job?

Knife-edge tools are made for checking by eye against light. Flat tools have broad working faces that register against the work, a surface plate or an indicator. Granite squares and cylindrical squares are references that stand on a surface plate. CEM notes that small squares, with arms up to about 200 mm or 300 mm, are used on their own, for example to see straightness against the light or to set parts at 90°, while larger steel or granite squares serve as references together with a surface plate.

Common straightedges and squares and the standards they are made to
ToolWhat it isStandard
Knife-edge straightedge (Haarlineal)A steel straightedge with a narrow lapped edge, for the light-gap checkDIN 874-2:2003
Flat straightedgeA steel or granite bar of rectangular section with flat working facesDIN 874-1:2003 (steel); BS 5204-2:1977 (steel or granite)
Cast iron straightedgeA bow-shaped or I-section casting for long lengths, up to 8 m bow-shaped and 5 m I-section in BS 5204-1BS 5204-1:1975, grades A and B
Flat square and square with a stockA steel square with a plain blade, or with a thicker stock that registers against the work (Anschlagwinkel)DIN 875-1:2005 types A, B and D; BS 939:2007 try squares
Knife-edge square (Haarwinkel)A steel square with knife edges on the blade, for the light-gap checkDIN 875-1:2005 types C and E
Granite squareA square of natural hard stone used on a surface plate as a referenceDIN 875-2:2008
Cylindrical squareA cylinder with its end faces square to its axis, stood on a surface plate as a referenceBS 939:2007, grade AA
Block squareA solid or open block with faces at 90°BS 939:2007

Straightness or flatness: what can a straightedge tell you?

Straightness belongs to a line and flatness to a surface. A straightedge checks one line at a time. Laid along the length, across and on the diagonals, it shows where each line on a surface is high or low. It does not give one flatness value for the whole surface.

A flatness value needs a survey of many lines tied together into one surface, as in the eight-line Moody method for surface plates. That is the subject of the surface plate guide. The plate is also the reference that squares and flat straightedges are checked on, so its own calibration comes first.

Rest a long straightedge on the support points its maker marks. They are placed to keep the sag under its own weight small. Moving the supports changes the shape of the working edge.

Checking a square: compare it, or reverse it

A square has two errors that matter: the angle between its blade and stock, and the straightness of each working edge. In practice the quick check compares the square with a better reference, such as a granite square or a cylindrical square. Stand both on a surface plate, bring the blade against the reference and look for light along the joint.

Without a better reference, reversal separates the errors. Measure the square against a fixed reference, such as an indicator on a column, then turn the square through 180° so that its own error points the other way, and measure again against the same reference. The reference's error is the same in both readings and the square's changes sign, so half the difference between the two readings is the square's error and half the sum is the reference's. National metrology institutes use the same principle. In a EUROMET comparison of squareness measurements, the Slovak Institute of Metrology measured a granite square at 0° and at 180° to separate the square from its measuring column.

CEM procedure DI-009 describes four laboratory methods for squares: a profile projector for small squares with thin edges, a coordinate measuring machine, a column squareness tester with a surface plate for large or heavy squares, and a roundness instrument for small squares with flat rather than knife edges, which it rates as highly accurate. With the column, an indicator is zeroed at the bottom of the blade and read at the top. The squareness deviation is the difference between the two readings divided by the length between them. If a square has both an inside and an outside right angle, DI-009 takes the larger of the two deviations.

A worked example. The indicator reads 0 µm at the bottom of the blade and +4 µm at the top, 150 mm higher. The deviation is 4 µm in 150 mm, about 27 µrad or 5.5 seconds of arc. One maker specifies its 150 × 100 mm grade 00 knife-edge square at 0.004 mm squareness (Hoffmann Group), so a square of that kind would sit right at its limit before the uncertainty of the check is counted.

The standards: DIN 875, DIN 874, BS 939 and BS 5204

Squares. DIN 875-1:2005 sets dimensions, grades and tests for steel squares of types A to E, with tolerances for squareness and for the flatness or straightness of the test faces. It classes squares as 00, 0, 1 and 2, with 00 the finest. DIN Media now lists DIN 875-1 as withdrawn and names no successor, but makers still sell squares to its grades. DIN 875-2:2008 for granite squares is current. CEM's procedure DI-009 notes that DIN 875-1 has no ISO or Spanish UNE equivalent, treats its class table as a guide, and asks a certificate to state the squareness deviation with its expanded uncertainty. If the certificate also gives a class, the deviation plus its uncertainty must be below the class limit.

In the UK, BS 939:2007 covers try squares in grades AA, A and B from 50 mm to 1,000 mm, cylindrical squares in grade AA from 75 mm to 750 mm, and solid and open block squares.

Straightedges. DIN 874-1:2003 covers flat steel straightedges and DIN 874-2:2003 knife-edge straightedges. In the UK, BS 5204-1:1975 covers cast iron straightedges and BS 5204-2:1977 steel or granite straightedges of rectangular section.

Test instructions. In Germany the VDI/VDE/DGQ 2618 series gives step-by-step test instructions for monitoring test equipment. The sheets for squares and straightedges also carry the DKD name: VDI/VDE/DGQ/DKD 2618 Blatt 7.1 (2019) for steel squares, Blatt 5.1 (2022) for flat straightedges and Blatt 5.2 (2024) for knife-edge straightedges, which includes an example uncertainty budget for a straightness test on a 300 mm straightedge.

Tip: This guide does not reproduce the tolerance formulas or tables of these standards. Take the limit for your square or straightedge from the standard it was made to, or from the maker's certificate.

Recalibration periods by grade

CEM procedure DI-009 suggests recalibration periods for squares by grade, shown in the table. It leaves the final period to the user and shortens it for heavy use, poor conditions such as dirt or outdoor work, and special safety needs. It calls for recalibration after any repair and whenever the results raise doubt.

As practice, treat a straightedge the same way. A knife edge that has been dropped or dragged across a rough surface should be checked against a reference before it is trusted again.

Recalibration periods for squares suggested in CEM procedure DI-009
GradeSuggested period
0048 to 60 months
036 to 60 months
1 and 224 to 60 months

Handling, temperature and care

  • Lift, do not slide. Lift a straightedge or square to each new position, because dragging the edge across a surface wears it (INSIZE).
  • Check smooth surfaces only. A rough surface wears the edge and scatters the light, which is why INSIZE sets the Ra 0.2 µm limit.
  • Hold the insulating shield. INSIZE asks you to hold its straightedges by the shield so the heat of the hand does not affect the result.
  • Let large squares settle. The granite square in the EUROMET comparison had an expansion coefficient of 5.6 × 10⁻⁶ per °C, about half the 11.5 × 10⁻⁶ per °C of gauge block steel given by NIST. Granite still has to reach the temperature of the room before it is used as a reference.
  • Keep steel from rust. Wipe the edges after use, apply a rust preventive oil and keep each tool in its case (INSIZE).

How to check a straightedge or square with the light gap, step by step

  1. Clean the working edge and the surface with a clean soft cloth. The surface should be smooth, no rougher than Ra 0.2 µm.
  2. Set an even light behind the joint, such as a soft fluorescent lamp 60 mm to 90 mm from the gap.
  3. Stand the edge on the surface, square to it, and look along the joint at eye level.
  4. Judge any gap by its colour, or compare it with a standard light gap made with gauge blocks on an optical flat.
  5. Lift the edge and repeat along the length, across and on both diagonals.
  6. For a square, stand it and a reference square or cylindrical square on a surface plate, bring the blade against the reference and look for light along the blade.
  7. Record where any gap appeared and its estimated width. Keep a square with an unexplained gap out of use until it has been calibrated.

Tracking squares and straightedges in Axiospec

Axiospec is calibration management software. It does not measure squareness or judge a light gap. The grade, the limits and the interval stay with you or your laboratory. Each square, straightedge and cylindrical square can be its own instrument on the register, with its certificates, its calibration history and a next due date set in days, weeks, months or years, so a long period for a grade 00 reference square sits beside shorter ones for shop squares.

One calibration record holds up to 100 named test points, each with a nominal value, its own tolerance and the as-found and as-left readings, so the inside and outside angles of a square and the straightness of each edge can each be their own point. A reference square or cylindrical square can be marked as a reference standard on the register. With a standard such as ISO 9001 or ISO/IEC 17025 selected for the workspace, you pick it when you log a calibration, and its Reverse Traceability card lists the calibrations it was used on.

An accepted Fail or Damaged result puts the instrument in quarantine. With a standard selected, any approved Fail, Damaged or Pass with an as-found reading out of tolerance needs an impact written, and those not yet written show in Needs attention. The free plan covers up to 50 active instruments. A new workspace starts with a 45-day trial of Professional, with room for 1,500 active instruments and no credit card.

Common questions

How do you check if a square is accurate?
Compare it with a granite or cylindrical square on a surface plate and look for light along the blade. Or measure it against a fixed reference, turn it through 180° and measure again: half the difference between the two readings is the square's own error.
What is the light gap method?
A knife-edge straightedge or square is set on a surface in front of an even light, and any gap shows as light along the edge. The colour gives a rough width, with gaps over about 2 µm looking white in INSIZE's guidance, and a standard light gap made with gauge blocks gives a comparison.
What does DIN 875 grade 00 mean?
It is the finest of the four classes, 00, 0, 1 and 2, that DIN 875-1:2005 sets for steel squares. DIN Media now lists DIN 875-1 as withdrawn, but squares are still sold to its grades.
Can a straightedge measure flatness?
Not on its own. A straightedge checks straightness along one line at a time. Lines along, across and on the diagonals show high and low spots, but a flatness value needs a survey that ties many lines into one surface.
What is the difference between a knife-edge square and a flat square?
A knife-edge square (Haarwinkel) has narrow lapped edges for checking by the light gap. A flat square has broad working faces for work where the square has to register against the part, an indicator or a surface plate.
How often should a square be calibrated?
CEM procedure DI-009 suggests 48 to 60 months for grade 00, 36 to 60 months for grade 0 and 24 to 60 months for grades 1 and 2. It shortens the period for heavy use or poor conditions and calls for recalibration after a repair or whenever results raise doubt.

Put it into practice