Inside micrometer and bore micrometer: finding the true diameter and calibrating them

11 min read Published

A two-point inside micrometer gives the true diameter only on the line through the centre of the bore, so you rock it along the bore axis for the smallest reading and swing it across the bore for the largest. A three-point bore micrometer centres itself on three contacts and is set and checked on a setting ring.

On this page

Two-point or three-point: which inside micrometer do you have?

Inside micrometers come in two families. A two-point instrument touches the bore at two opposite points, like an outside micrometer turned inside out. A three-point instrument, often called a bore micrometer, has three contacts that a cone pushes outward together, and they centre the head in the bore. CEM, the Spanish national metrology institute, publishes a separate calibration procedure for each: DI-021 for two-contact and DI-022 for three-contact inside micrometers (both 2019).

Three-point micrometers are often called by a maker's name, such as Holtest (Mitutoyo) or Intrimik (TESA). They usually come in sets. Several heads share one reading unit, extensions reach deep bores, and one or more setting rings come in the case.

The common types of inside micrometer, with ranges and scale divisions from CEM procedures DI-021 and DI-022
TypeHow it touches the boreUsual range and scale divisionCalibrated against
Tubular or rod type (two-point, axial)Two contacts on the axis of the micrometer head. Extension rods or tubes screw on for larger bores.Usually 25 mm of travel on the head, read to 0.01 mm or 0.001 mmGauge blocks held between two flat jaws
Caliper type (two-point, lateral)Two jaws set off to the side of the micrometer headUsually 25 mm, read to 0.01 mm or 0.001 mmGauge blocks held between two flat jaws
Three-point bore micrometerThree contacts spaced 120° apart, pushed out by a cone so the head centres itselfUsually 2 mm or 3 mm per head, sometimes up to 25 mm. Read to 0.01, 0.005, 0.002 or 0.001 mmSetting rings, or three-flat internal diameter standards
Tip: A dial bore gauge also measures holes, but it is a comparator set to zero on a known size rather than a micrometer. It has its own guide.

How do you find the true diameter with a two-point inside micrometer?

A two-point inside micrometer reads the diameter only when it lies on a line through the centre of the bore, square to the bore axis. Any other position reads a different length, and the sign of the error is known in advance. Mitutoyo's Quick Guide to Precision Measuring Instruments gives the geometry. Call the bore diameter ℓ and move one end of the micrometer by a distance X. If X is along the bore axis, the micrometer spans √(ℓ² + X²) and reads too large. If X is across the bore, it spans √(ℓ² − X²) and reads too small.

That gives the method. Rock the micrometer along the bore axis and keep the smallest reading. Swing it across the bore and keep the largest. Where the two searches meet, the micrometer lies on a true diameter.

The error grows with the square of the offset. In a 200 mm bore, one end 2 mm out of line costs about 0.010 mm, a whole division on a 0.01 mm micrometer. At 5 mm out of line the error is about 0.06 mm. The table works the formula for two sizes.

Size of the misalignment error, √(ℓ² ± X²) − ℓ. Too large when the offset is along the axis, too small when it is across the bore.
One end out of line byError in a 200 mm boreError in a 500 mm bore
1 mm0.0025 mm0.0010 mm
2 mm0.0100 mm0.0040 mm
5 mm0.0625 mm0.0250 mm

Reading a three-point bore micrometer

A three-point micrometer needs no search for the centre. Turning the thimble drives a cone that pushes the three contacts outward until they touch the wall, and the head centres itself. CEM describes these instruments as self-centring. Turn the thimble by its ratchet, as on an outside micrometer, so the measuring force is the same at each reading, and keep the head square to the bore.

The scale reads like an outside micrometer. The sleeve gives the whole millimetres and the thimble gives the fraction. Mitutoyo's example for a Holtest graduated in 0.005 mm reads 35 mm on the sleeve and 0.015 mm on the thimble, so the bore is 35.015 mm.

Where the contacts touch matters. Mitutoyo notes that a Holtest reads differently when the contacts bear across their full face and when only their tips touch, as near the bottom of a blind hole. Set the zero under the same condition you will measure in.

Two-point and three-point instruments also see the shape of a bore differently, because each design misses certain out-of-round shapes. The dial bore gauge guide shows which ones.

Extension rods and long micrometers

A tubular inside micrometer reaches large bores with extension rods or tubes screwed onto the head. Each joint adds the error of its rod and a place for dirt to sit between faces. As practice, clean the faces and threads before you assemble a combination and use the fewest pieces that reach the size.

Calibration treats the rods on their own. CEM procedure DI-021 calibrates the head over its basic range against gauge blocks, then measures each extension on a horizontal one-coordinate measuring machine and works out the correction and uncertainty for the combination in use. ISO 5059-1:2025 specifies the length measurement error of a two-point inside micrometer both with and without interchangeable extensions.

A long micrometer also bends under its own weight when it lies on two supports. Mitutoyo gives the spacings. Supports about 0.577 of the length apart, the Airy points, keep the ends of the bar level. Supports about 0.559 of the length apart, the Bessel points, keep the change in length smallest. As practice, set and use a long combination in the same position, horizontal or vertical, so the sag is the same both times.

Why the heat of your hand changes the reading

Steel grows by about 11.5 µm per metre for each degree Celsius (NIST, The Gage Block Handbook). A 300 mm inside micrometer that warms by 3 °C in the hand grows by about 0.010 mm, a full division on a 0.01 mm scale. The longer the combination, the larger the effect.

Hold the micrometer by its insulated grip or with gloves, as Mitutoyo advises, and let the micrometer, its setting standard and the part reach the same temperature before you measure. ISO 5059-1 lists a thermally insulating grip among the design characteristics of a two-point inside micrometer. Lengths are specified at 20 °C, the reference temperature in ISO 1:2022.

Setting rings: the reference for a bore micrometer

CEM procedure DI-022 notes that three-point sets usually include one or more setting rings, which show whether the micrometer still works properly and when it needs recalibration. As practice, check the reading on the ring before a run of measurements and again after the micrometer has been knocked.

A setting ring is a reference standard with its own calibration. EURAMET Calibration Guide No. 6 (version 3.0, 2024) describes three levels for cylindrical diameter standards. The basic level, five diameters around the nominal direction, is for rings that set two-contact instruments. The diameter and roundness level adds roundness at three heights with the diameter in each of those planes, and it also covers rings for three-contact instruments, as does the full level. The guide asks the certificate to state clearly where the diameter was measured.

In the US, ASME B89.1.6 covers the measurement of plain internal diameters for use as master rings or ring gages.

How is an inside micrometer calibrated?

A calibration compares the micrometer with traceable reference standards at points across its range and reports the error at each point with its uncertainty. The two families use different references and different sampling, as the table shows.

CEM procedure DI-022 also says when to adjust a three-point micrometer. If the correction at the mid-range point is at least twice the scale division on a micrometer reading to 0.002 mm or finer, or at least one division on a coarser scale up to 0.01 mm, the scale is adjusted where that is possible and the ten readings at that point are repeated.

The specifications behind the limits differ by region. ISO 5059-1:2025, first published in April 2025, covers the design and metrological characteristics of two-point inside micrometers, with default maximum permissible error (MPE) values in an annex. In Germany it partly replaces DIN 863-4:1999. ASME B89.1.13-2013, reaffirmed in 2022, has a section on inside micrometers with its own MPE tables. The German test instructions are VDI/VDE/DGQ 2618 Blatt 10.7 (2010) for two-point contact and VDI/VDE/DGQ/DKD 2618 Blatt 10.8 (2024) for 3-line contact.

Calibration of inside micrometers in CEM procedures DI-021 (two-point) and DI-022 (three-point)
StepTwo-point (DI-021)Three-point (DI-022)
ReferenceGauge blocks held between two flat jaws in a holder. Grade 1 or better for a 0.001 mm scale, grade 2 or better for 0.01 mm.Setting rings or three-flat internal diameter standards
Test pointsAt least six, roughly evenly spaced over the rangeTen readings near mid-range first, then further points that divide the range into roughly equal parts
Repeat readingsTen at each point recommended, changing the micrometer's position between the jaws each timeIf eight or more of the first ten readings agree, one reading at each further point. Otherwise ten at each.
Extensions and headsEach extension measured on a horizontal one-coordinate measuring machineEach head of a set calibrated in turn
Room20 °C ± 5 °C, after at least 30 minutes to settle20 °C ± 5 °C, after at least 30 minutes to settle
Tip: This guide does not reproduce MPE tables. The limit for your micrometer comes from the standard it was made to or from the maker's specification.

Recalibration and checks between calibrations

CEM suggests 12 months as a starting interval for both two-point and three-point inside micrometers and leaves the final period to the user. Both procedures shorten it for heavy use, poor conditions or special safety needs, and call for recalibration whenever the results raise doubt.

Between calibrations, a reading on the setting ring or on gauge blocks in their holder shows drift early. Record each check with its date. A zero that keeps moving is a reason to shorten the interval. ISO/IEC 17025:2017 clause 6.4.10 calls checks like these intermediate checks.

If a micrometer is found out of tolerance, record the as-found readings before any adjustment. ISO 9001 clause 7.1.5.2 asks you to judge whether earlier results were adversely affected and to act as needed.

How to measure a bore with a two-point inside micrometer, step by step

  1. Clean the contacts, the extension faces and the bore. Assemble the head with the fewest extensions that reach the size.
  2. Let the micrometer, its setting standard and the part reach the same temperature. Hold the micrometer by its insulated grip.
  3. Check the reading on gauge blocks in a holder, or on a setting ring, at a size close to the bore.
  4. Put one contact against the wall and open the micrometer until the other contact just touches the opposite side.
  5. Rock the free end along the bore axis and find the smallest reading.
  6. Swing the free end across the bore and find the largest reading. Repeat both searches until the reading no longer changes.
  7. Lock the spindle if the micrometer has a clamp and read it. Add the reading of the head to the stated length of each extension in the combination.
  8. Where the form of the bore matters, measure at more than one depth and in two directions, and record each reading with its depth and direction.

Keeping inside micrometers and setting rings in Axiospec

Axiospec is calibration management software. It does not measure bores or decide whether a micrometer passes. The limits and the interval stay with you or your laboratory. Each inside micrometer, bore micrometer set or setting ring can be its own instrument on the register, with its calibration history, its certificates and a next due date set in days, weeks, months or years.

One calibration record holds up to 100 named test points, each with a nominal value, its own tolerance and the as-found and as-left readings, so each point on the head and each extension combination can be its own point. A setting ring can be marked as a reference standard on the register. With a standard such as ISO 9001 or ISO/IEC 17025 selected for the workspace, you pick it when you log a micrometer calibration, and its Reverse Traceability card lists the calibrations it was used on.

An accepted Fail or Damaged result puts the instrument in quarantine. With a standard selected, any approved Fail, Damaged or Pass with an as-found reading out of tolerance needs an impact written, and those not yet written show in Needs attention. The free plan covers up to 50 active instruments. A new workspace starts with a 45-day trial of Professional, with room for 1,500 active instruments and no credit card.

Common questions

What is the difference between an inside micrometer and a bore micrometer?
An inside micrometer usually means a two-point instrument that touches the bore at two opposite points, such as a tubular micrometer with extension rods. A bore micrometer is a three-point instrument whose three contacts centre it in the bore, known by makers' names such as Holtest or Intrimik.
How do you find the true diameter with an inside micrometer?
Rock the micrometer along the bore axis and keep the smallest reading, then swing it across the bore and keep the largest. A tilt along the axis always reads too large and an offset across the bore always reads too small, so the true diameter is where the two searches meet.
How is a three-point inside micrometer calibrated?
Against setting rings or three-flat internal diameter standards. CEM procedure DI-022 takes ten readings near mid-range, then readings at further points spread over the range, and calibrates each head of a set in turn.
Do extension rods need to be calibrated?
Yes. Each rod adds its own length error, so CEM procedure DI-021 measures each extension on a horizontal one-coordinate measuring machine and combines the results for the combination in use.
What standard covers inside micrometers?
ISO 5059-1:2025 covers two-point inside micrometers, with and without extensions. ASME B89.1.13-2013 has a section on inside micrometers, and in Germany VDI/VDE/DGQ 2618 Blatt 10.7 and Blatt 10.8 are the test instructions for two-point and 3-line contact instruments.
How often should an inside micrometer be calibrated?
CEM suggests 12 months as a starting point for two-point and three-point inside micrometers and leaves the final interval to the user. Shorten it for heavy use or poor conditions, and check the reading on a setting ring or gauge blocks between calibrations.