Skip to content

Compliance and audits

Testing machine calibration and verification: ASTM E4 and ISO 7500-1 side by side

10 min read Published

A tension or compression testing machine is verified by placing a calibrated force-proving instrument in its load path and comparing the force the machine indicates with the force the instrument measures, at several forces across each range in repeated runs. ISO 7500-1 then gives each range a class, such as 0.5 or 1, while ASTM E4 has no classes and judges the machine's force error against its own allowable limit.

On this page

How is a testing machine verified, step by step?

The run below follows ISO 7500-1:2018 and EURAMET Calibration Guide No. 4. ASTM E4 works on the same principle, with its own runs, forces and limits set out in the paid text.

  1. Inspect the machine first. ISO 7500-1 calls for a general inspection of the machine and its force-application accessories (Annex A), and notes that good practice is to run a calibration before any maintenance or adjustment, so that the as-found condition is on record (clause 5).
  2. Mount a force-proving instrument calibrated to ISO 376, of a class at least as good as the class being verified, so that bending effects are kept to a minimum (6.4.1).
  3. Let the instrument reach a stable temperature. ISO 7500-1 wants the room between 10 °C and 35 °C and the temperature recorded, and the instrument preloaded at least three times to the maximum force (6.4.2, 6.4.3).
  4. Run three series of increasing forces, each with at least five force levels spread from 20% to 100% of the range, and add points if the range is to be used below 20% (6.4.5).
  5. At each level, either set the machine to a nominal reading and note the force the instrument measures, or bring the instrument to a nominal force and note the machine's reading. EURAMET recommends the first way (cg-4, 7.2).
  6. Read the zero about 30 seconds after each series is unloaded. Add a series with decreasing forces only where reversibility is required (6.4.5, 6.4.8).
  7. Work out the indication error and the repeatability at each force, then the class of the range from all the criteria together (6.5, clause 7).
  8. Write the verification report and set the next date (clauses 8 and 9).

ASTM E4 and ISO 7500-1 side by side

Both standards verify the force-measuring system of a static machine: the force sensing, the electronics and every indicator the results are read from, whether a dial, a recorder chart, a digital display or the data on a computer (ASTM E4-24, 1.3 and 1.5). Both are static. ISO 7500-1 notes that its values do not necessarily hold for high-speed or dynamic testing, and ASTM E4 covers static or quasi-static machines.

They differ in the reference they use and in what they report. The table sums up what their published scopes and the public ISO preview show. The limits and the number of runs in ASTM E4 are in the paid text.

ASTM E4-24 and ISO 7500-1:2018 compared
PointASTM E4-24ISO 7500-1:2018
Machines coveredStatic or quasi-static tension or compression machines, with or without force indicatorsTension and compression testing machines, static calibration
Reference usedStandard weights, equal-arm balances with weights, or elastic force measurement standards calibrated to ASTM E74Force-proving instruments calibrated to ISO 376, of a class at least as good as the machine's. Known masses only below the smallest instrument's range, with the relative error of the force they generate within ±0.1%
ResultThe force error is judged against an allowable error, which the scope refers to as ±1.0%A class for each range: 0.5, 1, 2 or 3
Measurement uncertaintyNot subtracted from the allowable error, so the true deviation can exceed ±1.0%Estimated by the method in informative Annex C
What is reportedA certificate and report of calibration and verification for each force-measuring system the owner namesA verification report with general information and the results (clause 8)
Tip: If a customer or a test method names one of the two, verify to that one. A range verified to ISO 7500-1 class 1 has not been verified to ASTM E4 just because the limits look alike, and the reverse holds too.

What does an ISO 7500-1 class tell you?

ISO 7500-1 classes each force range of a machine as 0.5, 1, 2 or 3, with 0.5 the best. The class rests on several measured quantities together: the indication error and the repeatability at each force, the reversibility where it is required, the zero error, the relative resolution of the indicator and the class of the proving instrument (EURAMET cg-4, 7.2). A range has to meet every criterion of a class to hold it.

The resolution of the indicator sets the bottom of a range. The lowest force that can be verified must be a set number of resolution steps, and the better the class, the more steps it needs (6.4.5). With a display that reads in 1 N steps, the lowest force verifiable to class 1 is 200 N, and to class 0.5 it is 400 N.

The limits for each class are in Table 2 of the paid standard. Your test method decides which class you need, so check it before you order the verification, and check that the certificate states the class and the forces covered for each range you use.

How does ASTM E4 handle the same job?

ASTM E4 allows three ways of applying known forces: standard weights, equal-arm balances with standard weights, and elastic force measurement standards such as proving rings and load cells. Each comes with a measurement uncertainty and traceability to the SI (E4-24, 1.2).

The elastic standards are calibrated to ASTM E74, which sets out how force-measuring instruments are calibrated so that a machine verified with them is traceable to the SI (E74, 4.1). E74 calibrations are static, and E74 says their results cannot be assumed valid for dynamic or high-speed force measurement (1.3).

The owner names the force-measuring systems to be verified, and the verification covers the indicators those systems use, down to displays and printouts of the data system (1.3, 1.5). The E4 scope also says plainly that the allowable error is not reduced by the measurement uncertainty, so a machine that passes can deviate from the true force by more than ±1.0% once the uncertainty is added (1.6).

Tip: Ask for the uncertainty at each force with the E4 results. It is the figure you need if your own test method or customer calls for a guard band.

Force-proving instruments: ISO 376 and ASTM E74

The reference in the load path is a force-proving instrument, an elastic device such as a proving ring or a load cell with its indicator. Under ISO 7500-1 it must comply with ISO 376 and be of a class at least as good as the class being verified. When more than one instrument covers a range, the top force on the smaller one must equal the bottom force on the next larger one (6.1).

ISO 376:2011 grades instruments as class 00, 0.5, 1 or 2. For an instrument classified for interpolation that only just meets each class, EURAMET gives the worst-case expanded uncertainty as 0.08%, 0.16%, 0.32% and 0.64%, before any temperature effect (cg-4, Table 6.1). An instrument classified for interpolation can be used at any force in its range, while one classified for specific forces can only be used at the forces it was calibrated at (NPL Guide to the Measurement of Force).

The instrument's share of the uncertainty covers its calibration, its sensitivity to temperature, its drift since calibration and any approximation in the equation that converts its output (cg-4, 7.2). ISO 376 waits 30 seconds before each reading, while ISO 7500-1 allows a slowly increasing force, so an instrument that creeps can read differently in the two (cg-4, 7.1).

What do the results look like? A worked example

EURAMET Calibration Guide No. 4 works through a 10 kN range verified at 19.5 °C, in three increasing series with nine forces from 2 kN to 10 kN. At the top force in the first series the machine displayed 10.01 kN while the proving instrument measured 9.951 kN. The relative indication error is the machine's reading minus the reference force, as a share of the reference force, here +0.59%.

Across the three series the mean error at 10 kN was 0.52%, with an expanded uncertainty of 0.22%. At 2 kN the mean error was 0.44% but the expanded uncertainty was 0.53%, because the indicator's resolution weighs more at a low force.

Selected results from EURAMET Calibration Guide No. 4, Annex B (10 kN range)
Nominal forceMean indication errorExpanded uncertainty
2 kN0.44%0.53%
4 kN0.67%0.38%
6 kN0.44%0.36%
8 kN0.51%0.27%
10 kN0.52%0.22%
Tip: EURAMET does not state a class for this example. Comparing errors like these, and the other criteria, with the class limits in your edition of ISO 7500-1 is the step that gives the class.

What should the verification report contain?

ISO 7500-1 clause 8 splits the report into general information and the results of the verification. ASTM E4 calls for a certificate and report of calibration and verification for each force-measuring system the owner names. The list below is good practice for checking a report, not a copy of either clause.

  • The machine, its serial number and each force-measuring system and range verified, in tension, compression or both.
  • The standard and edition, such as ISO 7500-1:2018 or ASTM E4-24.
  • Each force-proving instrument, with its certificate, its class or loading range and its due date.
  • The temperature during the run.
  • At each force, the machine's reading, the reference force, the error and its uncertainty, with the repeatability.
  • The class of each range under ISO 7500-1, or the range of forces verified under ASTM E4.
  • The as-found results before any adjustment and the as-left results after it.
  • The date, the laboratory and its accreditation.

How often should a testing machine be verified?

ISO 7500-1 has a clause on the intervals between verifications (clause 9). Its content, and any interval in ASTM E4, is in the paid texts, so take the interval from the edition you work to, and from your test method or customer if they set a shorter one.

Verify again after the machine is moved, repaired or adjusted, or after its load cell or force indicator is changed, whatever the date. ILAC-G24:2022 / OIML D 10:2022 lists the factors for setting an interval from your own results, such as drift between verifications and how heavily the machine is used.

Between verifications, a check at one or two forces with a reference load cell shows whether the machine has moved. Plot the results on a control chart so that drift shows before the next verification is due.

What if the machine fails as found?

Record the as-found results before anyone adjusts the machine. ISO 7500-1 notes that good practice calls for a calibration run before any maintenance, for exactly this reason (clause 5). If a range was outside its class or limit, ISO 9001 clause 7.1.5.2 asks you to determine whether earlier results were adversely affected and to act as needed.

Results from a testing machine often decide whether material is accepted. List the tests run on the affected range since the last good verification, and compare the error found with the margin each result had to its specification. A result far from its limit may stand, while one close to it may need a retest or a word with the customer.

How Axiospec helps with testing machines

Axiospec is calibration management software. It does not verify a testing machine or assign its class. The standard, the class or limit and the interval stay with you or your calibration provider. Each machine can be an instrument on the register with its verification history, certificates and next due date, and the interval can be set in days, months or years. A machine with several load cells can carry each as its own instrument.

One verification record holds up to 100 named test points, each with a nominal force, a tolerance as an absolute band, a percentage of nominal or a percentage of span, and the as-found and as-left readings, so each force in each range can be its own point. With a standard such as ISO 9001 or ISO/IEC 17025 selected for the workspace, you pick the force-proving instrument from the register as the reference standard, and its Reverse Traceability card lists the machines it was used on.

An accepted Fail or Damaged result puts the machine in quarantine. With a standard selected, any approved Fail, Damaged or Pass with an as-found reading out of tolerance needs an impact written, and any not yet written show in Needs attention. The free plan covers up to 50 active instruments. A new workspace starts with a 45-day trial of Professional, with room for 1,500 active instruments and no credit card. We load your existing records for you at no charge on every plan.

Common questions

What is the difference between ASTM E4 and ISO 7500-1?
Both verify the force-measuring system of static tension and compression testing machines. ISO 7500-1 uses ISO 376 proving instruments and gives each range a class of 0.5, 1, 2 or 3. ASTM E4 uses weights or E74-calibrated instruments and judges the force error against its allowable limit without subtracting the uncertainty.
What are the ISO 7500-1 classes?
Classes 0.5, 1, 2 and 3, with 0.5 the best. A range is classed from its indication error, repeatability, zero error, resolution, reversibility where required and the class of the proving instrument together, and it has to meet every criterion of its class.
Which force-proving instrument does ISO 7500-1 require?
One calibrated to ISO 376, with a class equal to or better than the class the machine is verified to. Below the lower limit of the smallest force-proving instrument, known masses are used instead, and the force they generate must be within ±0.1%.
How many runs does an ISO 7500-1 verification need?
Three series with increasing force, each with at least five force levels between 20% and 100% of the range, and more points if the range is used below 20%. A series with decreasing force is added only when reversibility is required.
How often should a tensile testing machine be calibrated?
ISO 7500-1 has a clause on the intervals between verifications, so follow the edition of the standard you work to and any shorter interval in your test method or contract. Verify again after the machine is moved, repaired or adjusted.
Does ASTM E4 take measurement uncertainty into account?
ASTM E4 does not reduce the allowable force error by the measurement uncertainty. Its scope notes that a machine verified to E4 can therefore deviate from the true force by more than ±1.0% once the uncertainty is added.

Put it into practice

Got a gage list? Send it over and we load it for you, usually in a couple of business days. Free on every plan. Then log every calibration to a tamper-evident audit trail and produce records on demand.

Axiospec is a documentation and workflow tool. It helps you keep clean, traceable, audit-ready records. Certification depends on your own processes, scope and assessor.