Force gauge calibration: how to use a push-pull gauge and how to calibrate it

10 min read Published

A force gauge is calibrated by applying known forces, from certified dead weights or a reference load cell, at five or more points up to full scale in compression and again in tension, and comparing each reading with the force applied. How well it then measures depends just as much on how it is used: loaded along its shaft through the right attachment, and read in the right mode.

On this page

What is a force gauge?

A force gauge measures a push (compression) or a pull (tension) along its loading shaft. It is held by hand or mounted on a test stand. Makers include Mark-10, Imada, Shimpo and Chatillon. Mark-10 lists springs, electronic components, fasteners, caps, films and mechanical assemblies among the items its gauges typically test.

A digital gauge senses force with a load cell and shows it on a display that can also hold peak values and send readings to a computer. A mechanical gauge uses a spring and a dial with a needle pointer. Some mechanical models add a slave pointer that stays at the highest force reached (Mecmesin).

A test stand holds the gauge and moves it toward or away from the part, by hand or with a motor. It keeps the gauge square to the part, and a motorized stand repeats the same travel at the same speed, which a hand cannot do.

What does ±0.1% of full scale mean at low readings?

Makers state force gauge accuracy as a percentage of full scale. Mark-10 gives ±0.1% of full scale for its Series 7 gauges and ±0.3% for Series 3 (user's guides). A full-scale figure is a fixed amount of force across the whole range. On a 500 N gauge, ±0.1% of full scale is ±0.5 N whether the reading is 450 N or 25 N, and at 25 N that is 2% of the reading.

So choose the capacity for the forces you measure. A gauge read mostly in the bottom tenth of its range has a large error relative to the reading even when it meets its specification. A smaller gauge does better, as long as the peaks stay within its capacity. Mark-10 rates its Series 3 and Series 7 gauges for a safe overload of 200% of full scale and shows OVER from 110%.

The same ±0.1% of full scale specification on a 500 N gauge, as a share of the reading
ReadingAllowed errorShare of the reading
500 N±0.5 N0.1%
250 N±0.5 N0.2%
100 N±0.5 N0.5%
50 N±0.5 N1%
25 N±0.5 N2%

Peak or real time: which reading mode should you use?

Digital gauges show the live force in real-time mode and the highest force reached in peak mode, with separate peaks for compression and tension. Mark-10's gauges hold the peak until ZERO is pressed or the gauge is switched off and on.

Use peak mode when the result is the highest force in an event, such as a break, a pull-off or the force needed to start a movement. Use real-time mode when you hold a load and read it, such as a spring compressed to a set length. Some gauges detect a break from the fall in force after the peak and then store or send the peak by themselves (Mark-10 Series 7).

A short peak is only caught if the gauge samples fast enough. Mark-10 quotes 14,000 Hz for Series 7 and 2,000 Hz for Series 3, and recommends keeping the current-reading filter at its lowest setting for the best response. A calibration applies steady forces. DKD-R 3-3, for example, covers static loading only, so a certificate does not show how well a gauge catches a fast peak.

How do you get repeatable force readings?

NPL's Guide to the Measurement of Force treats axial alignment as a major factor in any force measurement. A force applied off the axis, with side loads or bending, can give large errors that no calibration covers. The points below are good practice, with the maker's instructions where noted.

  • Load along the shaft. Mark-10 tells users to position grips and fixtures so that the load is axial to the gauge's loading shaft.
  • Fit attachments by hand. Mark-10 says to finger-tighten grips and attachments and not to use jam nuts or tools on the shaft.
  • Pick the attachment for the contact. Gauges come with tips such as flat, chisel, cone, V-groove and hook, plus an extension rod (Mark-10 Series 7). Use the same attachment each time a test is repeated.
  • Zero with the attachment fitted, in the position you will test in. The ZERO key tares the attachment's weight and clears stored peaks.
  • Use a stand when results must be compared over time. Mount the gauge on the stand's plate with its dowel pin, because Mark-10 says the mounting screws must not carry the load.
  • Stay inside the capacity. Mark-10 warns that more than 200% of capacity can damage the internal load cell, whether the gauge is on or off.

Which standards apply to a force gauge?

ISO 376 does not govern hand-held force gauges. Its scope is the calibration and classification of force-proving instruments used for the static verification of uniaxial testing machines, the reference devices a laboratory places in a tensile tester. ASTM E74 is the US counterpart, but its scope is wider. It covers the calibration of force-measuring instruments in general, both elastic instruments and force-multiplying systems. ASTM E4 relies on such instruments being traceable to the SI, and E74 sets out how they are calibrated (E74, 1.1 and 4.1).

A force gauge can still be calibrated with the same methods. In Germany, the PTB guideline DKD-R 3-3 (edition 03/2024) applies to force measuring devices that determine force through the elastic deformation of a body, as long as they can be loaded up and down with at least three force steps each way. NPL's guide notes that BS 8422 was written for force traceability outside materials testing. Otherwise the laboratory follows the maker's procedure.

Ask the laboratory which procedure it follows and whether its accreditation covers it, and name both on your order.

Dead weights or a reference load cell?

Mark-10's calibration instructions use certified dead weights or master load cells, with the gauge mounted vertically on a test stand or fixture strong enough for its full capacity.

A dead weight makes a force through gravity, so the force depends on where you are. A mass m gives a force F = mg(1 − ρa/ρm), where g is the local acceleration due to gravity, ρa the density of the air and ρm the density of the weight, the formula ISO 7500-1 uses for dead weights. For steel weights the air term takes off about 0.015%. A rounded g costs more. If the local value is 9.803 m/s² and a laboratory uses 9.81 m/s², every force it applies is about 0.07% too high. At full scale on a 500 N gauge that is 0.35 N of a ±0.5 N allowance.

Fixtures and hangers count. In Mark-10's procedure the weight fixtures are fitted before the zero is set, so their weight is tared and only the calibration weights load the gauge.

A reference load cell, loaded in series with the gauge in a stand or a frame, reaches forces that a weight stack cannot easily give. Its own calibration and uncertainty then become part of the gauge's result. NPL's guide suggests a calibration uncertainty three times better than the accuracy the gauge has to meet.

How to calibrate a force gauge, step by step

  1. Check the gauge as found. Look for damage to the shaft and housing, check that it zeros, and record readings at a few forces before anything is adjusted.
  2. Mount it vertically on a rigid stand or fixture rated for its full capacity, with its axis in line with the weights or the reference load cell.
  3. Fit the hanger or loading fixture and set zero with it in place.
  4. Exercise the gauge to full scale a few times. DKD-R 3-3 asks for two or three preloads at the maximum force, with at least a minute of preloading.
  5. Apply at least five evenly spaced forces up to full scale in compression, such as 20%, 40%, 60%, 80% and 100%, and record each reading once it is steady. DKD-R 3-3 takes a reading no sooner than 30 seconds after the force starts to change.
  6. Unload, check the return to zero, then repeat the same points in tension. DKD-R 3-3 treats tension and compression as separate calibrations.
  7. Compare each reading with the applied force and with the gauge's specification, as an error in force or as a share of full scale.
  8. If the gauge is adjusted, run the points again and record them as left.

Reading the results: an example

The table shows illustrative as-found results for a 500 N gauge with a specification of ±0.1% of full scale, so ±0.5 N at every point. The error grows in step with the force. That pattern points to a span (gain) error rather than a zero offset, which would shift every point by the same amount. Mark-10's procedure adjusts the gain with a full-scale load before the calibration points are entered.

Record both directions. A gauge can pass in compression and fail in tension, and the record should show which.

Illustrative as-found results in compression for a 500 N gauge, specification ±0.5 N
Applied forceReadingErrorWithin ±0.5 N?
100 N100.2 N+0.2 NYes
200 N200.3 N+0.3 NYes
300 N300.4 N+0.4 NYes
400 N400.6 N+0.6 NNo
500 N500.7 N+0.7 NNo

How often should a force gauge be calibrated?

Mark-10 recommends calibration about once a year, and more often under heavy use or where company policy calls for it. Treat that as a starting point and adjust it from the gauge's own history. NPL's guide advises lengthening the period when several calibrations show no appreciable change in sensitivity and shortening it when they show significant change, and lists conditions of use, frequency of use, harsh handling and the accuracy you need among the factors.

Check the gauge sooner after an overload, a drop or a repair. Between calibrations, a quick check with a known weight shows whether it still reads where it did, before parts are judged with a gauge that has drifted.

What should the calibration certificate show?

Mark-10 offers a certificate with data that lists 10 points in each direction, tension and compression, with as-found data on request. DKD-R 3-3 has the certificate state the unloaded zero signal and the position it was measured in, and say when the laboratory's own indicator was used instead of the customer's. The list below is good practice for checking any force gauge certificate.

  • The gauge's ID and serial number, its capacity and any remote sensor calibrated with it.
  • The procedure followed and the reference used, dead weights or a load cell, with its traceability.
  • Each applied force and reading in compression and in tension, with the error and the uncertainty.
  • The as-found results before any adjustment and the as-left results after it.
  • The temperature, the date and the next due date.

What if a force gauge fails as found?

Record the as-found readings before anyone adjusts the gauge. If they were outside the specification, ISO 9001 clause 7.1.5.2 asks you to determine whether earlier results were adversely affected and to act as needed. For a force gauge that means the parts and lots accepted with it since its last good calibration, and how far the error moves each result relative to its own limit. A pull-off test with a wide margin may be unaffected, while one close to its limit may need a retest.

How Axiospec helps with force gauges

Axiospec is calibration management software. It does not calibrate a force gauge or decide whether it passes. The specification, the decision rule and the interval stay with you or your calibration provider. Each gauge can be an instrument on the register with its calibration history, certificates and next due date, and a remote sensor can be an instrument of its own where it is calibrated separately.

One calibration record holds up to 100 test points. Each has a label, a nominal value, a tolerance as an absolute band, a percentage of nominal or a percentage of span, whether it was read up-scale or down-scale, and the as-found and as-left readings, so the compression and tension points can sit on one record. A specification of ±0.1% of full scale is a percentage of span. With a standard such as ISO 9001 or ISO/IEC 17025 selected for the workspace, the record also names the reference standard used, such as the weight set or the reference load cell.

On an in-house calibration, a final reading outside its tolerance sets the result to Fail, and recording a pass anyway takes a written reason. An accepted Fail or Damaged result puts the gauge in quarantine. With a standard selected, any approved Fail, Damaged or Pass with an as-found reading out of tolerance needs an impact written, and any not yet written show in Needs attention. The free plan covers up to 50 active instruments. A new workspace starts with a 45-day trial of Professional, with room for 1,500 active instruments and no credit card. We load your existing records for you at no charge on every plan.

Common questions

How do you calibrate a force gauge?
Mount it vertically on a rigid stand, zero it with the hanger or fixture fitted, and apply known forces from certified weights or a reference load cell at five or more evenly spaced points up to full scale. Do it in compression and again in tension, and compare each reading with the applied force and the gauge's specification.
Does ISO 376 apply to hand-held force gauges?
No. ISO 376 covers force-proving instruments used to verify uniaxial testing machines. A force gauge can be calibrated with similar methods, for example under the German guideline DKD-R 3-3, which covers force measuring devices that work by elastic deformation, or to the maker's procedure.
How often should a force gauge be calibrated?
Mark-10 recommends about once a year, and more often under heavy use. Adjust the interval from the gauge's own as-found history, and check it sooner after an overload or a drop.
What is the difference between peak and real-time mode on a force gauge?
Real-time mode shows the force as it changes. Peak mode holds the highest force reached in compression or in tension until it is cleared, which suits breaks, pull-offs and other events where the maximum is the result.
What does ±0.1% of full scale mean on a force gauge?
It is a fixed allowance in force across the whole range. On a 500 N gauge it is ±0.5 N at every reading, which is 2% of a 25 N reading, so choose a capacity close to the forces you measure.
Can I check a force gauge with weights myself?
Yes, as a check between calibrations. Zero the gauge with the hanger fitted, hang a known mass and compare the reading with the mass times local gravity. A calibration also needs traceable references, a defined procedure and a stated uncertainty, which a laboratory provides on its certificate.

Put it into practice

Got a gage list? Send it over and we load it for you, usually in a couple of business days. Free on every plan. Then log every calibration to a tamper-evident audit trail and produce records on demand.

Axiospec is a documentation and workflow tool. It helps you keep clean, traceable, audit-ready records. Certification depends on your own processes, scope and assessor.