Test weight calibration: OIML R 111 and ASTM E617 classes, handling and care
Test weight calibration compares each weight with a reference weight of a better class on a mass comparator and reports its conventional mass with an uncertainty, which for an OIML R 111 weight may be no more than a third of its class limit. The class, E1 to M3 under OIML R 111 or 000 to 7 under ASTM E617, also governs how the weight is made and how it must be handled.
On this page
Which OIML class does a weight need?
OIML R 111-1:2004 is the international recommendation for weights from 1 mg to 5,000 kg, amended in 2025. It has nine accuracy classes. E1 is the most accurate, followed by E2, F1, F2, M1, M1-2, M2, M2-3 and M3. Each class has a maximum permissible error, or MPE, for each nominal value, and the MPE applies to the weight's conventional mass (clause 5.1.1).
The classes form a chain. R 111 intends class E1 weights to link national mass standards to class E2 and below. Each class from E2 to M2 then calibrates the class below it. E2 weights also serve special accuracy (class I) balances, F1 and F2 weights serve high accuracy (class II) instruments, and the M classes serve medium and ordinary accuracy scales (clause 1.3.1). M1-2 and M2-3 cover only large weights, from 50 kg to 5,000 kg, for class III scales.
Each step down allows about three times the error of the class above. The table shows the limits at three nominal values.
| Class | 100 g weight | 1 kg weight | 20 kg weight |
|---|---|---|---|
| E1 | 0.05 mg | 0.5 mg | 10 mg |
| E2 | 0.16 mg | 1.6 mg | 30 mg |
| F1 | 0.5 mg | 5.0 mg | 100 mg |
| F2 | 1.6 mg | 16 mg | 300 mg |
| M1 | 5.0 mg | 50 mg | 1,000 mg |
| M2 | 16 mg | 160 mg | 3,000 mg |
| M3 | 50 mg | 500 mg | 10,000 mg |
ASTM E617 classes for laboratory weights
The US specification for laboratory weights is ASTM E617, Standard Specification for Laboratory Weights and Precision Mass Standards. The 2023 edition has classes 000, 00, 0, 1, 2, 3, 4, 5, 6 and 7. Like R 111, it sets maximum permissible errors (formerly called tolerances) and limits on magnetism, density and surface roughness for each class. Its tables are in the paid standard, so take the limit for a nominal value from your copy or from the weight's certificate.
The ASTM abstract makes two points that matter when you buy or recalibrate weights. A maker who claims E617 compliance must be able to show evidence for the material, density, magnetism, surface finish, mass values and uncertainties of new weights. A later claim of compliance needs a recalibration that meets the standard's own requirements and environmental limits, done by a laboratory that meets ISO/IEC 17025:2017 (clause 1.5).
For legal metrology field work, NIST Handbook 105-1 (2019) points officials to ASTM E617 and OIML R 111 when they choose field standard weights. It also says that the old NIST Class F field standards, with tolerances of about 0.01% of nominal, should not be bought for weights and measures use after January 1, 2020.
What is conventional mass?
A weight certificate states conventional mass, not mass. OIML R 111 and OIML D 28 define it as the mass of a reference weight with a density of 8,000 kg/m³ that the weight would balance in air with a density of 1.2 kg/m³ at 20 °C. Conventional mass leaves the air buoyancy correction out, which works as long as the air and the weight stay close to those reference values.
The difference from mass is small but not always negligible. OIML D 28 puts the relative difference between conventional mass and mass at about −3 × 10⁻⁴ for aluminium and +1 × 10⁻⁴ for platinum. Stainless steel is not exactly 8,000 kg/m³ either. For a 1 kg weight with a density of 7,950 kg/m³, the conventional mass is about 0.94 mg less than the mass. That is nearly twice the 0.5 mg MPE of a 1 kg E1 weight.
Density limits keep the buoyancy effect inside the class. R 111 sets them so that a 10% change in air density from 1.2 kg/m³ causes an error of no more than a quarter of the MPE (clause 10.1). Air is thinner at altitude, so R 111 asks for the density and its uncertainty to be given for class E weights used above 330 m and for class F1 weights used above 800 m (10.2.2).
Magnetism, density and surface: the limits by class
A weight only stays stable if its material, surface and magnetic properties suit its class. R 111 sets limits for each, and its Table 8 shows which are tested at type approval, at initial verification and at later verification.
Magnetism. A magnetized weight is attracted to ferromagnetic parts in and around the balance, so the reading shifts with no change in mass. R 111 sets its limits so that the magnetic effect stays below a tenth of the MPE (9.3). NIST adds that weights can pick up magnetization in a strong field during manufacture or use, and that the resulting errors can go undetected (NIST IR 6969, Good Measurement Practice 10).
Surface and construction. Class E and F weights must be non-porous and glossy. In case of doubt, R 111 gives roughness limits that run from Rz 0.5 µm and Ra 0.1 µm for E1 to Rz 5 µm and Ra 1 µm for F2, doubled for weights over 50 kg (11.1.2, Table 6). Class E weights up to 50 kg must be solid and made in one piece, with no cavity open to the air (7.1.1).
Material. Class E weights from 1 g need hardness and wear resistance at least like austenitic stainless steel, and class F weights from 1 g at least the hardness of drawn brass. Class M weights may be brass or grey cast iron, or a material at least as good, depending on their size and shape (clause 8).
| Class | Maximum polarization µ0M | Density of the weight |
|---|---|---|
| E1 | 2.5 µT | 7,934 to 8,067 kg/m³ |
| E2 | 8 µT | 7,810 to 8,210 kg/m³ |
| F1 | 25 µT | 7,390 to 8,730 kg/m³ |
| F2 | 80 µT | 6,400 to 10,700 kg/m³ |
| M1 | 250 µT | At least 4,400 kg/m³ |
| M2 | 800 µT | At least 2,300 kg/m³ |
| M3 | 2,500 µT | No limit set |
How should test weights be handled, cleaned and stored?
Handling changes mass. Oil and moisture from skin add to it at once and can corrode the surface later, so NIST says small weights should never be touched by hand. Use forceps, clean gloves or lint-free cloth, and keep those clean as well (NIST IR 6969, Good Measurement Practice 10). NIST does allow large weights of lower classes to be handled with bare hands.
- Lift weights rather than slide them. R 111 wants the handling tools for class E and F weights to leave the surface unscratched and unchanged (14.2.2).
- Wear gloves for the weights, not for the room. NIST warns that gloves can pass heat and moisture to a weight, and that gloves used on pens, balances or keyboards carry contamination with them.
- Clean only when you must. Cleaning can change the mass, so R 111 has you remove dust before calibration and wash with alcohol, distilled water or another solvent only for dirt that will not come off. Weights with cavities should not be immersed. If you follow a weight's stability, weigh it before cleaning (B.4.1).
- Wait after a solvent. R 111 gives stabilization times after cleaning with alcohol of 7 to 10 days for E1, 3 to 6 days for E2, 1 to 2 days for F1 and 1 hour for F2 to M3 (Table B.1).
- Let weights reach room temperature. R 111 Table B.2 sets mandatory minimum times by class, size and temperature difference. Some are longer than 24 hours, such as 45 hours for a class E1 weight of 10 kg to 50 kg that starts 20 °C away from room temperature. As a practical guide, R 111 also recommends 24 hours (B.4.3.1). NIST reports errors of up to 0.2 mg on 20 g weights with a 4 °C temperature difference, and warns that frost or condensation can form unseen on weights brought in from conditions more than 10 °C different.
- Store weights in their case. R 111 requires cases with a separate cavity for each class E and F weight, to protect them from shock and vibration (14.2.1). Keep the relative humidity between 40% and 60%, the range R 111 and NIST recommend, and keep weights away from strong magnetic fields.
- Mark weights so they can be told apart. R 111 calls user marking good practice because it links each weight to its certificate, and it limits the number and height of the marks (13.6, Table 7). Class E weights should only be marked when they must be told apart from another class E weight (13.2).
How is a weight calibrated by comparison?
Weights are calibrated by substitution. The test weight and a reference weight of known conventional mass are placed in turn on the same mass comparator or balance, and only the difference between them is measured. R 111 Annex C describes direct comparison against one reference and a subdivision method that calibrates a whole set from one or a few references.
The order of the weighings is the weighing cycle. With A for the reference and B for the test weight, the ABBA and ABA cycles cancel linear drift and are the usual choice for class E and F weights. The AB1…BnA cycle compares several test weights with one reference and is often used for class M (C.4).
The reference should generally be of a higher class than the weight being calibrated (C.2.3). On an electronic balance, NIST keeps the difference between the two weights below 0.05% of the balance capacity and uses a sensitivity weight of 10 to 20 times that difference, but no more than 1% of capacity (NIST IR 6969, Good Measurement Practice 10).
| Class | ABBA cycles | ABA cycles |
|---|---|---|
| E1 | 3 | 5 |
| E2 | 2 | 3 |
| F1 | 1 | 2 |
| F2 | 1 | 1 |
| M1, M2, M3 | 1 | 1 |
Does the result pass? The one-third rule and the acceptance band
R 111 sets two conditions on each weight. The expanded uncertainty U of the conventional mass, with a coverage factor k = 2, must be no more than a third of the MPE (clause 5.2). The conventional mass must then lie within the MPE less that uncertainty on either side of the nominal value (5.3.1). The uncertainty comes out of the limit, so a less certain calibration leaves a narrower band.
A worked example for a 1 kg F1 weight. The MPE is ±5.0 mg, so U may be at most 1.67 mg. If the laboratory reports U = 1.0 mg, the conventional mass must be within ±4.0 mg of 1 kg. A result of 1 kg + 3.2 mg passes. A result of 1 kg + 4.4 mg fails F1, although it is inside ±5.0 mg.
Class E1 and E2 weights always come with a certificate, and R 111 expects their users to apply each weight's actual deviation from its nominal value rather than treat it as exactly nominal (5.3.2).
How a weight is calibrated by comparison, step by step
- Inspect the weight for damage, corrosion and dirt, and record its identification and condition as found.
- Remove dust with a soft brush or lint-free cloth. Use a solvent only if needed, then wait the stabilization time for its class.
- Leave the test weight and the reference beside or inside the comparator until both are at room temperature. Allow at least the time in R 111 Table B.2 for the class, size and temperature difference, and 24 hours as a practical guide.
- Choose a reference weight of a higher class with a current certificate, and check that its uncertainty leaves room for the one-third rule.
- Exercise the comparator a few times with a load near the nominal value before taking readings.
- Run the weighing cycles, ABBA or ABA, at least as many as R 111 Table C.3 asks for the class, keeping the time between weighings the same.
- Calculate the mean difference and add the reference's conventional mass. Include the air buoyancy correction unless it is negligible by the R 111 test: no more than a third of the expanded uncertainty (C.5.1.2). Even then, its uncertainty may still have to be included (C.6.3.1). For class E weights, determine the air density (C.6.3.5).
- Work out the expanded uncertainty and compare the result with the MPE less that uncertainty. Record the as-found value before any adjustment and the as-left value after it.
How often should test weights be recalibrated?
R 111 sets no interval. It expects weights that were calibrated or verified to be recalibrated or verified again, so that you can show they kept their properties, and weights found defective to be readjusted or taken out of use (15.3.1).
NIST's interval guidance for weights and measures laboratories gives starting points. Working standards start at 12 months in a laboratory working at what NIST calls Echelon I or II, and at 24 months for Echelon III calibrations. Cast iron working standards start at 6 months until data show they are stable. No interval may exceed 10 years without exceptional analysis, and the laboratory adjusts each interval from its own calibration history and control charts (NIST IR 6969, Good Measurement Practice 11, 2019).
Between calibrations, a check standard or a control chart of repeated comparisons shows drift early. Recalibrate sooner after a weight is dropped, scratched, cleaned with a solvent or exposed to a strong magnetic field. ILAC-G24:2022 / OIML D 10:2022 lists the factors to weigh when you set or change an interval.
What should a weight certificate state?
R 111 sets a minimum for any calibration or verification certificate: the conventional mass of each weight, whether it was adjusted before calibration, the expanded uncertainty U and the coverage factor k (15.2.1). For class E1 weights it adds the density or volume of each weight and whether that value was measured or estimated (15.2.2.1). Certificates must come from authorized or accredited laboratories, with traceability to national standards (15.2).
When a certificate comes back, check these points before the set goes back into use. The first two come from R 111. The rest are good practice.
- A conventional mass and an uncertainty for each weight, not only a pass for the set.
- The density and how it was found, where the class or the altitude calls for it.
- Each weight identified by its nominal value and its mark or place in the case.
- The values as found for any weight the laboratory adjusted, so you can judge what the set measured before.
- The class the weights were checked against, with the document and edition, such as OIML R 111-1:2004 or ASTM E617-23.
- The laboratory's accreditation and scope for mass, and the calibration date.
What if a reference weight is found out of tolerance?
A weight set is the reference for everything it checks, so a set found outside its class reaches further than one instrument. Record the as-found values first. ISO 9001 clause 7.1.5.2 asks you to determine whether earlier results were adversely affected and to act as needed. List the scales, balances and weights the set was used on since its last good calibration, and compare the error found with their limits.
R 111 provides for a weight that no longer meets its class to be readjusted or taken out of use (15.3.1). Using it at its certified value, or at a lower class, is a decision for your own procedure, and the weight would still have to meet that class's other requirements, such as density and magnetism.
How Axiospec keeps weight sets and their certificates
Axiospec is calibration management software. It does not calibrate weights or decide which class they meet. The class, the limits and the interval stay with you or your laboratory. A weight set can be one instrument on the register, or each weight can be its own, with its certificates, its calibration history and a next due date set in days, months or years.
With a standard such as ISO 9001 or ISO/IEC 17025 selected for the workspace, you pick the set from the register as the reference standard when you log a scale or balance calibration. The set's Reverse Traceability card then lists the instruments it was used on, which is the list you need if the set is found out of tolerance. One calibration record holds up to 100 named test points, each with a nominal value, a tolerance as an absolute band such as the MPE in mg, and the as-found and as-left readings, so each weight in a set can be its own point.
An accepted Fail or Damaged result puts the instrument in quarantine. With a standard selected, any approved Fail, Damaged or Pass with an as-found reading out of tolerance needs an impact written, and any not yet written show in Needs attention. The free plan covers up to 50 active instruments. A new workspace starts with a 45-day trial of Professional, with room for 1,500 active instruments and no credit card. We load your existing records for you at no charge on every plan.
Common questions
- What are the OIML weight classes?
- OIML R 111-1:2004 has nine classes: E1, E2, F1, F2, M1, M1-2, M2, M2-3 and M3, from the most to the least accurate. Each sets a maximum permissible error for each nominal value, for example ±0.5 mg for a 1 kg E1 weight and ±50 mg for a 1 kg M1 weight.
- What is the difference between OIML R 111 and ASTM E617 weight classes?
- OIML R 111 is the international recommendation, with classes E1 to M3. ASTM E617 is the US specification for laboratory weights, with classes 000, 00, 0 and 1 to 7. They are separate documents with their own limits, so judge a weight against the class it is certified to.
- What is conventional mass?
- It is the mass of a reference weight of density 8,000 kg/m³ that a weight would balance in air of density 1.2 kg/m³ at 20 °C. Weight certificates and class limits use conventional mass rather than mass itself.
- How accurate must a weight calibration be?
- Under OIML R 111, the expanded uncertainty (k = 2) of a weight's conventional mass must be no more than a third of its maximum permissible error. The conventional mass must then lie within the maximum permissible error less that uncertainty.
- How often should calibration weights be recalibrated?
- OIML R 111 sets no fixed interval. NIST's guidance for weights and measures laboratories starts working standards at 12 or 24 months depending on the level of work, and cast iron ones at 6 months, then adjusts each interval from calibration history and control charts.
- Can you touch calibration weights with bare hands?
- Not small or precision weights. NIST says small weights should never be touched by hand and should be handled with forceps, clean gloves or lint-free cloth, because skin oils change the mass and can corrode the surface. NIST does allow large weights of lower classes to be handled with bare hands.
Put it into practice
Got a gage list? Send it over and we load it for you, usually in a couple of business days. Free on every plan. Then log every calibration to a tamper-evident audit trail and produce records on demand.
Axiospec is a documentation and workflow tool. It helps you keep clean, traceable, audit-ready records. Certification depends on your own processes, scope and assessor.
Take the audit-prep checklist with you
First the checklist, with each item phrased the way an auditor asks it. Then four short emails with practical tips. You can unsubscribe in one click.
First we email you a link to confirm your address. Nothing else comes until you confirm, and you can unsubscribe from any email after that. See our Privacy Policy.