Repeatability vs reproducibility: what changes between them, and how a Gage R&R measures each

6 min read Published

Repeatability is the spread of repeated readings when nothing changes: the same procedure, operator, measuring system, conditions and place, over a short time. Reproducibility is the spread when the operators, measuring systems and locations change, so it shows whether different people, gages or laboratories get the same result.

On this page

What stays the same, and what changes?

The VIM (JCGM 200:2012) defines both as measurement precision under a set of conditions, and the conditions are the whole difference. The table sets them side by side.

Repeatability and reproducibility compared
AspectRepeatabilityReproducibility
VIM entries2.20 (the condition) and 2.212.24 (the condition) and 2.25
Held the sameProcedure, operators, measuring system, operating conditions and locationThe same or similar objects are measured
Allowed to changeNothing but the repeat, over a short periodLocations, operators and measuring systems, and possibly the procedure
Question it answersDoes the gage give the same reading twice?Do different people, gages or laboratories agree?
In an AIAG Gage R&REquipment variation (EV)Appraiser variation (AV)
In ISO 5725Repeatability standard deviation, within one laboratoryReproducibility standard deviation, between laboratories

Where does intermediate precision fit?

Between the two sits intermediate precision (VIM 2.22 and 2.23). The procedure and the location stay the same and the measurements run over an extended period, while other conditions may change. The VIM gives new calibrations, calibrators, operators and measuring systems as examples. For the statistics, the VIM points to ISO 5725-3.

This matters for studies inside one plant. A Gage R&R there usually changes only the appraiser, with the same gage, place and method. That is neither the repeatability condition, because the operator changes, nor the full reproducibility condition, because the location and measuring system do not. The VIM's answer is to state the conditions changed and unchanged (2.22, note 2, and 2.24, note 2), and a study report should do the same.

How do EV and AV in a Gage R&R map to them?

In the AIAG Measurement Systems Analysis (MSA) reference manual, a Gage R&R study has several appraisers measure the same parts more than once with one gage. Equipment variation (EV) is the repeatability, the spread when one appraiser measures the same part again. Appraiser variation (AV) is the reproducibility, the spread between the appraisers' averages. The two combine as GRR = √(EV² + AV²).

One step is easy to miss. Each appraiser's average still carries some repeatability, so the average and range method subtracts that share before it reports AV and sets AV to zero when the subtraction goes negative. The Gage R&R calculator on this site does the same.

AV is therefore only the part the appraisers add. A reproducibility figure in the VIM's sense describes the spread of single results under reproducibility conditions, so it includes the repeatability. The NIST/SEMATECH e-Handbook (section 2.4.3.1) gives the ASTM reproducibility standard deviation that way, as the root sum of squares of the day-to-day and short-term parts. A figure of that kind lines up with GRR, not with AV.

Worked example: three inspectors, one micrometer, one part

Three inspectors each measure the same pin five times with the same digital micrometer, which reads to 0.001 mm. A full Gage R&R uses about ten parts and the AIAG average and range method. One part is enough to show the arithmetic with variances, working in µm (1 µm = 0.001 mm).

  1. Repeatability. Pool the three variances. The repeatability variance is 0.50 µm², so the repeatability standard deviation is 0.71 µm.
  2. Spread of the means. The three means, 12.003, 12.006 and 12.004 mm, have a variance of 2.33 µm².
  3. Take out the repeatability. Each mean averages n = 5 readings, so it still carries 0.50 / 5 = 0.10 µm² of repeatability variance. That leaves 2.33 − 0.10 = 2.23 µm² between inspectors, a standard deviation of 1.49 µm. The NIST/SEMATECH e-Handbook (section 2.4.4.2) uses the same subtraction for day-to-day effects.
  4. Combine. The measurement system's standard deviation is √(0.50 + 2.23) = √2.73 = 1.65 µm, the figure AIAG calls GRR.
Five readings of one pin by each inspector
InspectorReadings (mm)Mean (mm)Variance (µm²)
A12.002, 12.003, 12.004, 12.003, 12.00312.0030.50
B12.005, 12.006, 12.007, 12.006, 12.00612.0060.50
C12.003, 12.004, 12.005, 12.004, 12.00412.0040.50
Tip: The inspector part is about twice the repeatability. The micrometer's repeatability standard deviation is under 1 µm, yet inspector B reads 3 µm above inspector A on the same pin, so look at the method before the instrument: how each person uses the ratchet, how long they hold the micrometer and how they seat the pin. Against a tolerance of ±0.025 mm, 6 × 1.65 µm = 9.9 µm, about 20 percent of the 50 µm tolerance width. With three inspectors the inspector part rests on two degrees of freedom, so treat this as a first look.

Which reproducibility does your document mean?

The word changes meaning from one document to the next, so check which conditions changed before you compare two figures.

  • VIM: precision under conditions that include different locations, operators and measuring systems (2.24 and 2.25).
  • ISO 5725-2: the reproducibility standard deviation of a standard measurement method, estimated from an interlaboratory study in which each laboratory measures the same sample several times under repeatability conditions.
  • AIAG MSA: appraiser variation (AV), the spread between appraisers in one study, with the repeatability taken out.
  • NIST/SEMATECH e-Handbook, gauge studies: level 2 of a nested design is labelled reproducibility and means day-to-day variation (section 2.4.4).
Tip: Two more mix-ups. Good repeatability says nothing about bias, because a gage can repeat a wrong value all day. Repeatability is not resolution either: a display can show 0.001 mm while repeated readings scatter over several digits.

Where Axiospec fits

In Axiospec, a workspace with the Gage R&R studies module on runs these studies from the instrument page, so the result stays with that instrument and its calibrations. The free Gage R&R calculator on this site does the AIAG average and range arithmetic in the browser with no account.

Common questions

What is the difference between repeatability and reproducibility?
Repeatability is the spread of repeated readings with nothing changed: the same procedure, operator, measuring system, conditions and place over a short time (VIM 2.20 and 2.21). Reproducibility is the spread when the locations, operators and measuring systems change (VIM 2.24 and 2.25). A study that changes only some of them should say which ones changed and which did not (2.24, note 2).
Is EV repeatability and AV reproducibility?
Yes, in an AIAG Gage R&R. Equipment variation (EV) is the repeatability and appraiser variation (AV) is the variation between appraisers. GRR combines them as the square root of EV² + AV².
Can reproducibility be smaller than repeatability?
In a Gage R&R it can. AV is only the part the appraisers add once the repeatability is taken out, so it can be smaller than EV or even zero. A reproducibility standard deviation that includes the repeatability, such as the ASTM figure, cannot be smaller than the repeatability.
What is intermediate precision?
Precision with the same procedure and location over an extended period, when other conditions such as calibrations, calibrators, operators or measuring systems may change (VIM 2.22 and 2.23). It sits between repeatability and reproducibility.
How do you improve poor repeatability or poor reproducibility?
Poor repeatability usually points at the gage, its fixturing or the part itself, so check the instrument's condition and how the part is held. Poor reproducibility usually points at the method, so write down how the measurement is taken and train the appraisers on it.

Put it into practice