Surface roughness tester calibration: Ra, Rz, reference specimens and filter settings
A surface roughness tester is calibrated with reference specimens of certified value: a depth specimen with grooves of known depth sets the vertical scale, a roughness specimen with certified Ra and Rz values checks the whole instrument, and a tip condition specimen shows whether the stylus is worn. Between calibrations, a short check on the tester's own roughness specimen shows whether it still reads within your limit.
On this page
What do Ra, Rz and Rq measure?
Ra is the arithmetic mean of the absolute heights of the roughness profile, measured from its mean line. Picture the valleys folded up above the line: Ra is the average height of the result (NPL Good Practice Guide No. 37). A single odd peak or scratch moves it very little. Surfaces with very different shapes can share the same Ra, so it says nothing about whether a surface is peaky or has deep valleys. Under ISO 21920-2, most parameters, Ra among them, are worked out once over the whole evaluation length instead of being averaged over several sampling lengths (Polytec and Digital Surf guides to ISO 21920).
Rz is the maximum height of the profile: the height of the highest peak plus the depth of the deepest valley within a section, averaged over the sections measured, five unless the specification says otherwise. It stays a section-length parameter under ISO 21920, with the same definition as before. Rz reacts far more than Ra to one scratch or burr. Watch for an older meaning: an earlier ISO edition used the symbol Rz for the ten-point height, and some instruments still offer that parameter (NPL). Check which Rz the drawing means and which one the tester reports.
Rq is the root mean square of the profile heights. Squaring gives extra weight to the larger heights, so on the same profile Rq is never smaller than Ra. NPL notes that Rq is less affected by the phase effects of electrical filters and stays common in optics, while Ra has largely replaced it on machining drawings.
There is no fixed factor between Ra and Rz. The ratio depends on the shape of the surface, so a table that converts one into the other gives an estimate, not a measurement.
| Parameter | What it is | Length it is worked out over under ISO 21920-2 |
|---|---|---|
| Ra | Mean of the absolute heights from the mean line | The evaluation length |
| Rq | Root mean square of the heights | The evaluation length |
| Rz | Highest peak plus deepest valley in a section, averaged over the sections | Each section, then averaged |
Which standards apply to roughness testers now?
The profile roughness standards were rewritten in 2021. PTB, Germany's national metrology institute, summarizes the change: ISO 21920-1 replaces ISO 1302 for drawing indications, ISO 21920-2 replaces ISO 4287 and ISO 13565-2 and -3 for terms and parameters, and ISO 21920-3 replaces ISO 4288 for the default measuring conditions. Polytec's guide to the new series notes that drawings made to the old standards keep their validity and that most results change little.
- In the US, ASME B46.1, Surface Texture (Surface Roughness, Waviness, and Lay), defines the parameters. Its current edition is from 2019, and NIST calculates the parameters in its roughness calibrations as ASME B46.1 describes them.
- Many testers let you choose the standard. Mitutoyo lists ISO 4287, ISO 21920:2021, ASME B46.1 and several JIS editions for its SJ-220, so the setting on the tester is part of the result.
- DKD-R 4-2, the German Calibration Service guideline for roughness measurement published by PTB, predates ISO 21920. Its sheet 2 (2018) still cites ISO 3274, ISO 4287 and ISO 4288. Its procedures stay useful, so map its references to the new numbers in your own procedure.
| Subject | Current document | It replaced |
|---|---|---|
| Indication on drawings | ISO 21920-1:2021 | ISO 1302 |
| Parameters such as Ra, Rz and Rq | ISO 21920-2:2021 | ISO 4287, ISO 13565-2 and ISO 13565-3 |
| Default measuring conditions and acceptance rules | ISO 21920-3:2021 | ISO 4288 |
| Design and characteristics of stylus instruments | ISO 25178-601:2025 | ISO 3274:1996, withdrawn in February 2025 |
| Calibration of stylus instruments | ISO 12179:2026 | ISO 12179:2021 |
| Reference specimens (material measures) | ISO 5436-1:2000, now being revised as ISO 25178-70 | ISO 5436:1985 |
| Gaussian profile filter | ISO 16610-21:2025 | ISO 16610-21:2011 |
How do you check a roughness tester before use?
On the specimen that came with it. Mitutoyo, for example, lists a roughness specimen and a calibration stage among the standard accessories of its SJ-220, and a reference specimen of Ra 3 µm with the SJ-310. Measure the specimen at the conditions on its certificate and compare the reading with the certified value.
Switch the instrument on early. NPL suggests at least an hour before measuring, or the maker's stated warm-up time, so that it settles. Clean the specimen, and look at the stylus for damage before you start.
Keep the readings. NPL advises keeping a history of the specimen reading against its certified value and treating a change of 10 % from that history as a sign that the stylus may be worn or damaged.
Many testers adjust their gain to the specimen: you enter the certificate value and the instrument corrects itself, as NPL describes. Write down the reading before you adjust. That as-found reading is the one that tells you whether parts measured since the last check were affected.
A check on one specimen is not a calibration. It covers one roughness at one cutoff. A calibration also covers the vertical scale at several depths, the instrument's noise and the stylus.
| Question | Check before use | Intermediate check | Calibration |
|---|---|---|---|
| What is measured | The tester's own roughness specimen | Rz and RSm on a periodic specimen, ISO 5436-1 type C1 (DKD-R 4-2) | An optical flat, depth specimens and roughness specimens (DKD-R 4-2) |
| What it shows | That the tester still reads that specimen correctly | Drift in the vertical and horizontal scales between calibrations | The deviation at each specimen, with its uncertainty |
| Who does it | The operator | The operator or the metrology lab | A calibration lab, at the place of use where possible |
| Due date | Unchanged | Unchanged | Set again from the calibration date |
What does a full calibration involve?
ISO 5436-1 defines five types of specimen, also called material measures. NPL suggests that users of stylus instruments keep at least four of them, types A, B, C and D, each calibrated on a more accurate traceable system. ISO 5436-2 adds software measurement standards for checking the instrument's calculations.
- DKD-R 4-2 sheet 2 is a model procedure for accredited labs calibrating the vertical system of a stylus tester. Five traces on an optical flat give the instrument's own noise, Rz0, and the straightness of its guide, Wt0. The depth specimen is measured at the depths that matter in later use, five times in the same place and five times in traces close by. The roughness specimen is traced 12 times across its measuring area, and the mean of each parameter is compared with its certificate (sections 5.1 to 5.3).
- The deviations are reported in micrometres and as a percentage of the value. Calibration with depth specimens gives the traceability to the national length standard, and the roughness specimen confirms the whole chain from stylus to result (sections 1 and 5).
- The guideline works in two steps. If, after adjustment, the depth specimen still deviates by more than the expanded uncertainty plus a small constant, 0.01 µm in accredited labs' experience, the instrument cannot be calibrated. Otherwise the roughness specimen follows, and the tester is fit for roughness measurement when those results agree within their expanded uncertainty plus about 1 % of the certified values (section 7).
- Calibrate the tester where it is used, so the conditions it works in are part of the calibration, and keep direct sunlight and other temperature gradients away (section 4). When components are combined differently, such as another pick-up on the same base unit and drive unit, each combination is calibrated on its own (section 3).
| Type | What it is | What it checks |
|---|---|---|
| A, depth | Wide grooves of known depth, flat or rounded at the bottom | The vertical scale |
| B, tip condition | Narrow grooves, groove patterns or a fine edge, sensitive to the size of the tip | The condition of the stylus tip |
| C, spacing | Repeating grooves, such as a sine wave or a triangle profile | The vertical scale, and the horizontal scale under some conditions |
| D, roughness | An irregular profile, like a ground surface, that repeats along its length | The whole instrument, as an overall check, at the cutoff on its certificate |
| E, profile coordinate | A sphere or a precision prism | The coordinate system, for form measurement |
How do you know the stylus tip is still good?
The tip is the part that touches the work. NPL describes the usual stylus as a cone of 60° or 90° with a spherical tip of 1 µm, 2 µm, 5 µm or 10 µm radius. Mitutoyo offers its SJ-220, for example, with a 2 µm, 60° tip at 0.75 mN or a 5 µm, 90° tip at 4 mN. A larger tip cannot reach the bottom of narrow valleys, so height parameters read lower than the true surface, and a worn or chipped tip leads to serious errors (NPL).
A 2 µm tip cannot be judged by eye. NPL notes that seeing it may take a scanning electron microscope. Use a tip condition specimen instead. On a type B2 specimen, one groove pattern gives an Ra that depends on the tip size and the other an Ra that does not, so comparing the two shows the tip's condition. A type B3 specimen is a fine edge, such as a razor blade, that traces the tip's own shape, but it is easily damaged.
NIST works the same way for its own stylus. It calibrates the radius by tracing a standard wire and by the razor blade trace method, and states 1.52 µm ± 0.15 µm (k = 2) for the stylus it uses.
Which cutoff and filter settings cause errors?
The cutoff wavelength λc separates roughness from waviness. Detail shorter than λc counts as roughness and longer waves are filtered out, and the cutoff is the wavelength the filter passes at 50 % (NPL). A short-wave filter, λs, removes the finest detail. NIST's customary conditions for roughness are λc = 0.8 mm, λs = 2.5 µm and a 4 mm evaluation length, and NPL calls 0.8 mm the most commonly used value.
- A cutoff that is too short filters out part of the roughness and reads low. One that is too long lets waviness in and reads high. ISO 4288 chose the default cutoff from the roughness of the part, while ISO 21920-3 takes the default for Ra, Rq and Rz from the limit written on the drawing (Polytec).
- Set λs to match λc. NIST pairs 2.5 µm with 0.8 mm, and changing one without the other changes the band of wavelengths you measure.
- Measure the full evaluation length the specification calls for. A short trace is less typical of the surface, which is why NPL recommends assessing several consecutive sampling lengths.
- Trace across the lay, at right angles to the machining marks, unless the drawing says otherwise (NPL).
- The skid or the instrument's own guide is the reference the stylus is measured against, and any flaw in it adds to the reading (NPL).
- Check which parameter definitions the tester uses. A tester set to an older standard or a JIS edition can report a different Rz from one set to ISO 21920 on the same trace.
- Use a type D specimen only at the cutoff on its certificate (NPL). Measured at another cutoff, a correct tester can look wrong.
- The default acceptance rule for parts changed. Under ISO 4288 the 16 % rule applied unless the drawing said max, and under ISO 21920 the max rule applies unless the drawing says otherwise, so a single value over the limit fails the part (NPL, Polytec).
How often should a roughness tester be calibrated?
DKD-R 4-2 sheet 2 says recalibration intervals from 1 month to 2 years are usual, depending on the measuring task. It names the stability of the environment, changes of instrument components, how often the tester is used and the uncertainty required as the factors (section 9).
For the time in between, it suggests measuring Rz and RSm on a periodic type C1 specimen at shorter intervals and documenting the deviations. ILAC-G24:2022 / OIML D 10:2022 gives general guidance on choosing an interval (5.1) and on reviewing it from results (6). The specimens need calibrating too, and DKD-R 4-2 sheet 1 covers their calibration.
What should the calibration certificate show?
DKD-R 4-2 sheet 2 lists what a tester's certificate should contain (section 8). The last item is good practice for your own records. A certificate from an accredited lab also follows ISO/IEC 17025:2017 clause 7.8.
- Where the calibration was done, and each part of the instrument combination with its maker, type and serial number.
- The specimens used, with their identification and the date of their last calibration.
- The method, or the standard or guideline it follows.
- The conditions: temperature, measuring range, traverse speed, traversing length, cutoff, filter type and tip radius.
- The noise value Rz0 measured on the optical flat.
- The results with their measurement uncertainty, and the deviations from the certified values of the depth and roughness specimens, in micrometres and as a percentage.
- Which parameter definitions were used, ISO 21920-2 or the older ISO 4287, the readings before any adjustment and after it, the date, who did the work and the next due date.
What if the tester was out of tolerance as found?
Record the readings before anyone adjusts the tester. If an as-found result is out of tolerance, ISO 9001 clause 7.1.5.2 asks you to determine whether earlier results were adversely affected and to act as needed. For a roughness tester, look at the parts accepted on its readings since the last good check. A tester that read low may have passed surfaces that were too rough, which matters most where the finish governs friction, wear or lubrication.
How to calibrate a surface roughness tester, step by step
- Identify the tester, its drive unit, pick-up and stylus, the parameters and ranges you use and the limit that applies.
- Switch it on in good time, check that it works as the maker describes and inspect the stylus and the specimens.
- Check that the depth, roughness and tip condition specimens and the optical flat have current certificates.
- Align the specimen with the traverse and set the measuring conditions: cutoff, short-wave filter, evaluation length, speed and range.
- Trace the optical flat for the noise, then the depth specimen at the depths you use, repeating each trace.
- Trace the roughness specimen at the cutoff on its certificate, spread the traces over its area and compare the mean values with the certificate.
- Check the tip on a tip condition specimen.
- Record these as-found results before any adjustment. If you adjust, repeat the traces and record the as-left results.
- If a result was out of tolerance, assess the parts measured since the last good check. Then set the next due date.
Where Axiospec fits
Axiospec is calibration management software. It does not trace a specimen or judge a tester, and the limits and the decision rule stay with you or your lab. Each roughness tester and each specimen can be an instrument on the register, with its calibration history, certificates and next due date, and the interval can be set in days, months or years.
One calibration record holds up to 100 named test points, each with a nominal value, its own tolerance and the as-found and as-left readings, so the depth, the specimen's Ra and Rz and the tip check can each be a point of their own. Flag the specimens as reference standards on the register. With a standard such as ISO 9001 or ISO/IEC 17025 selected for the workspace, you pick the specimen you used when you log the tester's calibration, and the specimen's Reverse Traceability section lists the instruments it was used to calibrate.
An accepted Fail or Damaged result puts the tester in quarantine. With a standard selected, any approved Fail, Damaged, or Pass with an as-found reading out of tolerance needs an impact written, and unrecorded ones show in Needs attention. The free plan covers up to 50 active instruments. A new workspace starts with a 45-day trial of Professional, with room for 1,500 active instruments and no credit card.
Common questions
- How do you calibrate a surface roughness tester?
- Measure certified reference specimens with it: a depth specimen at the depths you use for the vertical scale, a roughness specimen at the cutoff on its certificate for the whole instrument, and a tip condition specimen for the stylus. Compare the readings with the certified values and record them before any adjustment.
- What is the difference between Ra and Rz?
- Ra is the average absolute height of the roughness profile from its mean line. Rz is the height from the highest peak to the deepest valley in each section, averaged over the sections, so it reacts much more to single scratches and burrs. There is no fixed factor that converts one into the other.
- How often should a roughness tester be calibrated?
- DKD-R 4-2 says intervals from 1 month to 2 years are usual, depending on the task, the environment, how often the tester is used and the uncertainty needed. Check it on its reference specimen before use and set the interval from its as-found history.
- What replaced ISO 4287 and ISO 4288?
- The ISO 21920 series of 2021. ISO 21920-2 replaced ISO 4287 for terms and parameters, ISO 21920-3 replaced ISO 4288 for the default measuring conditions, and ISO 21920-1 replaced ISO 1302 for drawing indications.
- What cutoff should I use to measure Ra?
- Use the cutoff on the drawing or specification when it gives one. NPL calls 0.8 mm the most commonly used value, and NIST measures roughness with a 0.8 mm cutoff over a 4 mm evaluation length. Under ISO 21920-3 the default follows from the limit written on the drawing.
- What is a roughness reference specimen?
- It is a small block with a certified surface profile, such as a sine wave or an irregular ground profile, and certified values such as Ra and Rz. It shows whether a roughness tester reads correctly, and it needs its own calibration.
Put it into practice
Got a gage list? Send it over and we load it for you, usually in a couple of business days. Free on every plan. Then log every calibration to a tamper-evident audit trail and produce records on demand.
Axiospec is a documentation and workflow tool. It helps you keep clean, traceable, audit-ready records. Certification depends on your own processes, scope and assessor.
Take the audit-prep checklist with you
First the checklist, with each item phrased the way an auditor asks it. Then four short emails with practical tips. You can unsubscribe in one click.
First we email you a link to confirm your address. Nothing else comes until you confirm, and you can unsubscribe from any email after that. See our Privacy Policy.