Compliance and audits
AMS2750 pyrometry: SAT and TUS for heat treating furnaces
A system accuracy test (SAT) checks that a furnace's own temperature measuring system reads true, by comparing it with a calibrated test sensor placed beside the furnace sensor. A temperature uniformity survey (TUS) checks that the whole qualified work zone stays within its class tolerance of setpoint, and SAE AMS2750 (current revision H, July 2024) sets out both for heat treating equipment.
On this page
SAT vs TUS: what is the difference?
The two tests answer different questions, so a pyrometry program needs both. A SAT looks at one measuring chain: the sensor, its wiring and the instrument that reads it. A TUS looks at the space where the parts sit and asks whether all of it reaches the temperature the controller shows.
A furnace can pass one and fail the other. Its control system can read true at the sensor while one corner of the load runs cold. Or the zone can be even while a drifting sensor makes the controller hold the whole zone at the wrong temperature.
| Question | System accuracy test (SAT) | Temperature uniformity survey (TUS) |
|---|---|---|
| What it checks | That a control or recording system reads true at its sensor | That the qualified work zone stays within tolerance of setpoint |
| Where the test sensor goes | Beside the measuring tip of the furnace sensor | At positions spread through the work zone |
| What is compared | The furnace reading with the corrected test reading | Each corrected test reading with the setpoint |
| Typical causes of a failure | Sensor drift, an instrument error or a wiring fault | Hot or cold spots from heating, airflow, insulation or door seals |
| What it cannot show | Whether the rest of the zone is at temperature | Whether the furnace sensors read true between surveys |
What does AMS2750 cover?
AMS2750 is an SAE Aerospace Material Specification titled Pyrometry. Its scope is the pyrometric requirements for equipment that thermally processes metallic materials, heat treating included. That takes in temperature sensors, instrumentation and the thermal processing equipment itself, along with correction factors and instrument offsets, system accuracy tests and temperature uniformity surveys.
The current revision is AMS2750H, revised in July 2024. SAE lists AMS2750G from June 2022 and AMS2750F from June 2020 before it, and the first issue dates from 1980. Requirements change between revisions, so check which one your customer's purchase order or material specification names.
AMS2750 is a paid SAE document. This guide explains its ideas in plain words and does not reproduce its tables, so take tolerances, intervals and sensor limits from the specification itself. It applies when a customer or a specification calls for it, as Nadcap heat treating audits do.
Thermocouples and other sensors
Most furnace pyrometry runs on thermocouples. A thermocouple is a pair of wires of different alloys joined at the measuring end. A temperature difference along the wires produces a small voltage, and the instrument converts it to temperature. Types are named by letter. Base metal types such as K and N are common in furnaces. Noble metal types such as R and S use platinum and platinum-rhodium wire and often serve as reference sensors.
Thermocouples drift. Heat changes the wire, so the same junction gives a different voltage after time at temperature. ASTM E601 notes that base metal thermoelements become thermoelectrically inhomogeneous once exposed above 200 °C. That is why AMS2750 limits how a sensor may be reused, depending on its type and the temperatures it has seen, and why the SAT exists for the sensors that stay in the furnace.
Sensors are named by their job. Control sensors feed the controller. Recording sensors feed the record of each cycle. Load sensors sit on or in the parts, and test sensors are brought in for a SAT or a TUS. A sensor is expendable when any part of its element is exposed to the furnace environment, and nonexpendable when a sheath or protection tube covers it.
Resistance thermometers (RTDs) are used too, mostly at lower temperatures. CQI-9 brought them into its thermocouple section in its 4th edition.
Instruments, correction factors and offsets
Each sensor is read by an instrument: the furnace controller, a recorder, or a portable field test instrument used for SAT and TUS work. Instruments are calibrated on their own, usually by feeding them a known thermocouple signal from a calibrator. Sensors are calibrated by comparison with a reference sensor.
Those calibrations give two kinds of number that pyrometry keeps apart. A correction factor is the amount, with its sign, that a calibration says to add to the reading of a sensor or instrument to get the true temperature. It is the error with the sign reversed. An instrument offset is a deliberate shift entered into a controller or recorder, for example to compensate for a known error. AMS2750 has rules for both, and SAT and TUS results are calculated with them. A sign error in either one can turn a pass into a fail, or the reverse.
Furnace class and instrumentation type
AMS2750 describes each furnace with two labels. The furnace class says how tightly the work zone must hold temperature, as a plus or minus tolerance around setpoint. The class a furnace needs follows from the heat treat specifications it runs, and a TUS is what proves it.
The instrumentation type is a letter, A to E, with a D+ variant in recent revisions. It says which sensors and recorders each control zone carries: a control sensor, recording sensors, load sensors and sensors at the hot and cold spots found in the last survey. Type A carries the most.
Together the two labels set how often SAT and TUS are due. A furnace with more instrumentation can usually be tested less often, and a run of passing surveys can lengthen the TUS interval. The specification gives the intervals, and this guide leaves them there.
How a SAT is run
A SAT is done on site, with the furnace at a normal operating temperature and the furnace sensor left in place. The technician puts a calibrated test sensor beside the measuring tip of the furnace sensor, connects it to a calibrated field test instrument and waits for both readings to settle.
Both are then read at the same time. The test reading is corrected with the correction factors of the test sensor and the test instrument. The difference between the furnace reading and the corrected test reading is the SAT result, and it has to fall within the limit the specification sets. The specification says which of the furnace's systems need a SAT, and each one is tested this way.
When the difference is too large, find the cause before you test again. A drifted sensor is replaced, an instrument that reads wrong is recalibrated, and a wiring fault is repaired. Then run the SAT again and keep both results.
How a TUS is run
A TUS proves the qualified work zone, the space inside the furnace where parts may be placed. Test sensors go at positions spread through that zone. Their number and positions follow from the zone's size and shape under the specification's rules, and they are read on a calibrated field test instrument or recorder.
The survey runs at temperatures chosen to cover the range the furnace is qualified for. Readings are taken through the rise to setpoint and then on through a soak period after the zone has stabilized. The specification sets when recording must start and the minimum time it continues after stabilization. The report shows the highest and lowest corrected readings, where they were and any overshoot above setpoint. The furnace passes when, after stabilization, all corrected readings stay within its class tolerance of setpoint.
The hottest and coldest positions matter after the survey too. Some instrumentation types place recording sensors at those spots, so the survey decides where they go.
Some furnaces cannot be surveyed in the usual way because of their design. For those cases CQI-9's 4th edition expanded its guidance on alternatives such as property surveys, which judge the zone by testing parts or test pieces processed in it.
What should SAT and TUS records show?
AMS2750 lists the contents it requires in each report, and CQI-9's 4th edition updated its own pyrometry reporting requirements. The lists differ in detail, so check them against the revision you work to. A record an auditor can follow usually holds the following.
- The furnace, the zone and the sensor or system tested.
- The date, the person who ran the test and the setpoint or operating temperature.
- The test sensor and field test instrument used, with their calibration dates and correction factors.
- For a SAT, the furnace reading, any offset, the corrected test reading and the difference.
- For a TUS, the sensor positions, the readings over time, the stabilization time, the highest and lowest readings and any overshoot.
- The result against the limit, and the date the next test is due.
How does AMS2750 relate to Nadcap and CQI-9?
Nadcap is the accreditation program for aerospace critical processes, run by the Performance Review Institute (PRI). Its heat treating audits use the criteria AC7102, and pyrometry has its own checklist, AC7102/8. Suppliers see the criteria through PRI's EAN platform, formerly eAuditNet. PRI lists failure to implement and maintain pyrometry in line with AMS2750 and customer specifications among the common nonconformances in heat treating audits.
CQI-9 is AIAG's Special Process: Heat Treat System Assessment, in its 4th edition since June 2020. It is the automotive counterpart: an assessment of a heat treat system that suppliers carry out themselves, with its own pyrometry section on thermocouples, instruments, SAT and TUS. Its requirements are written separately from AMS2750 and differ in detail, so a furnace that serves both markets has to meet both. Automotive customers that require CQI-9 say so in their customer-specific requirements.
| Question | AMS2750 | Nadcap heat treating | CQI-9 |
|---|---|---|---|
| Published by | SAE International | Performance Review Institute (PRI) | AIAG |
| What it is | A specification of pyrometry requirements | An accreditation program with audit criteria, AC7102/8 for pyrometry | A heat treat system assessment with a pyrometry section |
| Where it is used | Aerospace heat treating and the specifications that call for it | Aerospace, defense and space suppliers | Automotive heat treating |
| How it is checked | By the customer or an auditor, against the revision called for | By a Nadcap audit | By the supplier's own assessment and customer review |
What if a SAT or TUS fails?
A failed test means some readings could not be trusted, or some part of the zone was outside its tolerance, for a period that began after the last passing test. Correct the cause first, repeat the test and keep both results.
Then look back. List the loads processed since the last passing test and decide whether the error could have taken any of them outside the temperature tolerance of its specification. For the measuring system, ISO 9001 clause 7.1.5.2 asks the same question of any measuring equipment found unfit for its purpose: whether the validity of earlier results was affected, and what action to take.
Your customers' specifications can add their own steps, such as telling them. Follow those where they apply.
How Axiospec helps with pyrometry equipment
Axiospec is calibration management software. It does not run a SAT or a TUS or calculate their results. The tolerances, the intervals and the decision whether a furnace passes stay with you and your pyrometry provider. What it keeps is the equipment behind the tests. Field test instruments, test thermocouples, controllers and recorders can each be an instrument on the register with its calibration history, certificates and next due date.
Reference thermocouples can sit on the register as reference standards with their own certificates and due dates. With a standard such as ISO 9001 or ISO/IEC 17025 selected for the workspace, you pick the reference from the register when you log a calibration, and its Reverse Traceability card then lists the instruments it was used on. One record holds up to 100 named test points, each with a nominal value, a tolerance and the as-found and as-left readings, so each calibration temperature of a field test instrument can be its own point.
An accepted Fail or Damaged result puts the instrument in quarantine. With a standard selected, any approved Fail, Damaged or Pass with an as-found reading out of tolerance needs an impact written, and any not yet written show in Needs attention. That is where to note which furnace tests and loads the instrument touched since its last good calibration. The free plan covers up to 50 active instruments. A new workspace starts with a 45-day trial of Professional, with room for 1,500 active instruments and no credit card.
Common questions
- What is the difference between a SAT and a TUS?
- A SAT checks that a furnace's control or recording system reads true, by comparing it with a calibrated test sensor placed beside the furnace sensor. A TUS checks that the whole qualified work zone stays within its class tolerance of setpoint, using test sensors spread through the zone.
- What is the current revision of AMS2750?
- AMS2750H, revised by SAE in July 2024. The revision before it was AMS2750G from June 2022. Check which revision your customer's purchase order or specification names.
- What is a furnace class in AMS2750?
- The furnace class states how tightly the qualified work zone must hold temperature, as a tolerance around setpoint. The heat treat specifications a furnace runs decide the class it needs, and a passing TUS proves it.
- What is an instrumentation type?
- It is a letter from A to E, with a D+ variant in recent revisions, that says which control, recording and load sensors a furnace zone carries. Together with the furnace class it sets how often SAT and TUS are due.
- Is AMS2750 the same as Nadcap?
- No. AMS2750 is an SAE specification. Nadcap is the accreditation program run by PRI, and its heat treating audits check pyrometry against AMS2750 and customer specifications through the pyrometry criteria AC7102/8.
- Does CQI-9 follow AMS2750?
- CQI-9 has its own pyrometry section, written for automotive heat treating. It covers the same ideas, from thermocouples and instruments to SAT and TUS, but its requirements differ in detail. A furnace that serves both markets has to meet both.
Put it into practice
Got a gage list? Send it over and we load it for you, usually in a couple of business days. Free on every plan. Then log every calibration to a tamper-evident audit trail and produce records on demand.
Axiospec is a documentation and workflow tool. It helps you keep clean, traceable, audit-ready records. Certification depends on your own processes, scope and assessor.
Take the audit-prep checklist with you
First the checklist, with each item phrased the way an auditor asks it. Then four short emails with practical tips. You can unsubscribe in one click.
First we email you a link to confirm your address. Nothing else comes until you confirm, and you can unsubscribe from any email after that. See our Privacy Policy.