Calibrations and due dates

Log a calibration

Record a result in a few fields. Add readings and a certificate when you have them.

Updated

On this page

Open the form

Open the instrument and press Log a calibration. In the Calendar, the Log button next to an instrument opens the same form. On your Instruments list, Log pass opens it with Pass and today already filled in.

Fill it in

A record needs a result, a date and who did the work. Everything else is optional unless your standard asks for it.

  1. Pick the result: Pass, Fail or Other. Other holds Pass with adjustment, Limited calibration and Damaged.
  2. Set the Date. It cannot be in the future.
  3. Under Calibrated by, pick Me (in house) for work done in your shop. For an outside lab, pick it from your providers or choose Another lab or provider.
  4. Type the readings if you have them. With no standard selected, press Add readings first. Each row takes the nominal value with the as found and as left readings, plus the tolerance.
  5. Drop the certificate in the Certificate box and type its number.
  6. Check Next due. It fills in from the interval.
  7. Press Save. When approval applies, the button reads Submit for approval.

Tip: Logging several in a row? Save and log another keeps the form open.

Readings and tolerance

As found is the reading before any adjustment. As left is the reading after it. A reading outside tolerance sets the result to Fail. If you keep Pass, the form asks for a Reason for passing outside tolerance.

With a standard selected, the form shows Condition as received. Temperature and humidity are there too. Adjustments or repairs shows for Pass with adjustment and Limited calibration.

Signing

Some workspaces sign each calibration. There, saving opens Review this calibration. Enter your password and press Confirm and sign. If your account has no password yet, you set one in the same step.

After a failure

A failed result puts the instrument in quarantine. Open the record from Calibration history and press Record out-of-tolerance impact. Write what the instrument measured since its last good calibration and what was done about it. A Pass whose as found readings were out of tolerance asks for this too.

When a quarantined instrument passes again, tick Return to service on the form to release it.

Pass many instruments at once

On your Instruments list, tick the rows and press Log pass in the bar at the bottom. Confirm the count. Some instruments are skipped and listed in a summary afterwards:

  • Instruments with no earlier calibration on record.
  • Instruments that are quarantined or out at a lab.
  • Instruments whose standard needs readings.

Who can log

Technicians, managers and admins log calibrations. Auditors read records but cannot add them. Only a manager or admin can change Next due on the form.